Technical Note: Insertion of digital lesions in the projection domain for dual-source, dual-energy CT

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

PURPOSE: To compare algorithms performing material decomposition and classification in dual-energy CT, it is desirable to know the ground truth of the lesion to be analyzed in real patient data. In this work, we developed and validated a framework to insert digital lesions of arbitrary chemical composition into patient projection data acquired on a dual-source, dual-energy CT system.

METHODS: A model that takes into account beam-hardening effects was developed to predict the CT number of objects with known chemical composition. The model utilizes information about the x-ray energy spectra, the patient/phantom attenuation, and the x-ray detector energy response. The beam-hardening model was validated on samples of iodine (I) and calcium (Ca) for a second-generation dual-source, dual-energy CT scanner for all tube potentials available and a wide range of patient sizes. The seven most prevalent mineral components of renal stones were modeled and digital stones were created with CT numbers computed for each patient/phantom size and x-ray energy spectra using the developed beam-hardening model. Each digital stone was inserted in the dual-energy projection data of a water phantom scanned on a dual-source scanner and reconstructed with the routine algorithms in use in our practice. The geometry of the forward projection for dual-energy data was validated by comparing CT number accuracy and high-contrast resolution of simulated dual-energy CT data of the ACR phantom with experimentally acquired data.

RESULTS: The beam-hardening model and forward projection method accurately predicted the CT number of I and Ca over a wide range of tube potentials and phantom sizes. The images reconstructed after the insertion of digital kidney stones were consistent with the images reconstructed from the scanner, and the CT number ratios for different kidney stone types were consistent with data in the literature. A sample application of the proposed tool was also demonstrated.

CONCLUSION: A framework was developed and validated for the creation of digital objects of known mineral composition, and for inserting the digital objects into projection data from a commercial dual-source, dual-energy CT scanner. Among other applications, it will allow a systematic investigation of the impact of scan and reconstruction parameters on kidney stone dual-energy properties under rigorously controlled conditions.

Original languageEnglish (US)
Pages (from-to)1655-1660
Number of pages6
JournalMedical Physics
Volume44
Issue number5
DOIs
StatePublished - May 1 2017

Fingerprint

Kidney Calculi
X-Rays
Minerals
Calcium
Iodine
Kidney
Water

Keywords

  • computed tomography (CT)
  • dual-energy CT
  • lesion insertion
  • material decomposition
  • projection domain

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Technical Note : Insertion of digital lesions in the projection domain for dual-source, dual-energy CT. / Ferrero, Andrea; Chen, Baiyu; Li, Zhoubo; Yu, Lifeng; McCollough, Cynthia H.

In: Medical Physics, Vol. 44, No. 5, 01.05.2017, p. 1655-1660.

Research output: Contribution to journalArticle

@article{6e691b9f1cdf444ba3e754cbbf8c04cb,
title = "Technical Note: Insertion of digital lesions in the projection domain for dual-source, dual-energy CT",
abstract = "PURPOSE: To compare algorithms performing material decomposition and classification in dual-energy CT, it is desirable to know the ground truth of the lesion to be analyzed in real patient data. In this work, we developed and validated a framework to insert digital lesions of arbitrary chemical composition into patient projection data acquired on a dual-source, dual-energy CT system.METHODS: A model that takes into account beam-hardening effects was developed to predict the CT number of objects with known chemical composition. The model utilizes information about the x-ray energy spectra, the patient/phantom attenuation, and the x-ray detector energy response. The beam-hardening model was validated on samples of iodine (I) and calcium (Ca) for a second-generation dual-source, dual-energy CT scanner for all tube potentials available and a wide range of patient sizes. The seven most prevalent mineral components of renal stones were modeled and digital stones were created with CT numbers computed for each patient/phantom size and x-ray energy spectra using the developed beam-hardening model. Each digital stone was inserted in the dual-energy projection data of a water phantom scanned on a dual-source scanner and reconstructed with the routine algorithms in use in our practice. The geometry of the forward projection for dual-energy data was validated by comparing CT number accuracy and high-contrast resolution of simulated dual-energy CT data of the ACR phantom with experimentally acquired data.RESULTS: The beam-hardening model and forward projection method accurately predicted the CT number of I and Ca over a wide range of tube potentials and phantom sizes. The images reconstructed after the insertion of digital kidney stones were consistent with the images reconstructed from the scanner, and the CT number ratios for different kidney stone types were consistent with data in the literature. A sample application of the proposed tool was also demonstrated.CONCLUSION: A framework was developed and validated for the creation of digital objects of known mineral composition, and for inserting the digital objects into projection data from a commercial dual-source, dual-energy CT scanner. Among other applications, it will allow a systematic investigation of the impact of scan and reconstruction parameters on kidney stone dual-energy properties under rigorously controlled conditions.",
keywords = "computed tomography (CT), dual-energy CT, lesion insertion, material decomposition, projection domain",
author = "Andrea Ferrero and Baiyu Chen and Zhoubo Li and Lifeng Yu and McCollough, {Cynthia H}",
year = "2017",
month = "5",
day = "1",
doi = "10.1002/mp.12185",
language = "English (US)",
volume = "44",
pages = "1655--1660",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "5",

}

TY - JOUR

T1 - Technical Note

T2 - Insertion of digital lesions in the projection domain for dual-source, dual-energy CT

AU - Ferrero, Andrea

AU - Chen, Baiyu

AU - Li, Zhoubo

AU - Yu, Lifeng

AU - McCollough, Cynthia H

PY - 2017/5/1

Y1 - 2017/5/1

N2 - PURPOSE: To compare algorithms performing material decomposition and classification in dual-energy CT, it is desirable to know the ground truth of the lesion to be analyzed in real patient data. In this work, we developed and validated a framework to insert digital lesions of arbitrary chemical composition into patient projection data acquired on a dual-source, dual-energy CT system.METHODS: A model that takes into account beam-hardening effects was developed to predict the CT number of objects with known chemical composition. The model utilizes information about the x-ray energy spectra, the patient/phantom attenuation, and the x-ray detector energy response. The beam-hardening model was validated on samples of iodine (I) and calcium (Ca) for a second-generation dual-source, dual-energy CT scanner for all tube potentials available and a wide range of patient sizes. The seven most prevalent mineral components of renal stones were modeled and digital stones were created with CT numbers computed for each patient/phantom size and x-ray energy spectra using the developed beam-hardening model. Each digital stone was inserted in the dual-energy projection data of a water phantom scanned on a dual-source scanner and reconstructed with the routine algorithms in use in our practice. The geometry of the forward projection for dual-energy data was validated by comparing CT number accuracy and high-contrast resolution of simulated dual-energy CT data of the ACR phantom with experimentally acquired data.RESULTS: The beam-hardening model and forward projection method accurately predicted the CT number of I and Ca over a wide range of tube potentials and phantom sizes. The images reconstructed after the insertion of digital kidney stones were consistent with the images reconstructed from the scanner, and the CT number ratios for different kidney stone types were consistent with data in the literature. A sample application of the proposed tool was also demonstrated.CONCLUSION: A framework was developed and validated for the creation of digital objects of known mineral composition, and for inserting the digital objects into projection data from a commercial dual-source, dual-energy CT scanner. Among other applications, it will allow a systematic investigation of the impact of scan and reconstruction parameters on kidney stone dual-energy properties under rigorously controlled conditions.

AB - PURPOSE: To compare algorithms performing material decomposition and classification in dual-energy CT, it is desirable to know the ground truth of the lesion to be analyzed in real patient data. In this work, we developed and validated a framework to insert digital lesions of arbitrary chemical composition into patient projection data acquired on a dual-source, dual-energy CT system.METHODS: A model that takes into account beam-hardening effects was developed to predict the CT number of objects with known chemical composition. The model utilizes information about the x-ray energy spectra, the patient/phantom attenuation, and the x-ray detector energy response. The beam-hardening model was validated on samples of iodine (I) and calcium (Ca) for a second-generation dual-source, dual-energy CT scanner for all tube potentials available and a wide range of patient sizes. The seven most prevalent mineral components of renal stones were modeled and digital stones were created with CT numbers computed for each patient/phantom size and x-ray energy spectra using the developed beam-hardening model. Each digital stone was inserted in the dual-energy projection data of a water phantom scanned on a dual-source scanner and reconstructed with the routine algorithms in use in our practice. The geometry of the forward projection for dual-energy data was validated by comparing CT number accuracy and high-contrast resolution of simulated dual-energy CT data of the ACR phantom with experimentally acquired data.RESULTS: The beam-hardening model and forward projection method accurately predicted the CT number of I and Ca over a wide range of tube potentials and phantom sizes. The images reconstructed after the insertion of digital kidney stones were consistent with the images reconstructed from the scanner, and the CT number ratios for different kidney stone types were consistent with data in the literature. A sample application of the proposed tool was also demonstrated.CONCLUSION: A framework was developed and validated for the creation of digital objects of known mineral composition, and for inserting the digital objects into projection data from a commercial dual-source, dual-energy CT scanner. Among other applications, it will allow a systematic investigation of the impact of scan and reconstruction parameters on kidney stone dual-energy properties under rigorously controlled conditions.

KW - computed tomography (CT)

KW - dual-energy CT

KW - lesion insertion

KW - material decomposition

KW - projection domain

UR - http://www.scopus.com/inward/record.url?scp=85050300174&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85050300174&partnerID=8YFLogxK

U2 - 10.1002/mp.12185

DO - 10.1002/mp.12185

M3 - Article

C2 - 28241103

AN - SCOPUS:85050300174

VL - 44

SP - 1655

EP - 1660

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 5

ER -