Minimizing susceptibility-induced BOLD sensitivity loss in multi-band accelerated fMRI using point spread function mapping and gradient reversal

Myung Ho In, Daehun Kang, Hang Joon Jo, Uten Yarach, Nolan K. Meyer, Joshua D. Trzasko, John Huston, Matt A. Bernstein, Yunhong Shu

Research output: Contribution to journalArticlepeer-review

Abstract

Objective. Interleaved reverse-gradient fMRI (RG-fMRI) with a point-spread-function (PSF) mapping-based distortion correction scheme has the potential to minimize signal loss in echo-planar-imaging (EPI). In this work, the RG-fMRI is further improved by imaging protocol optimization and application of reverse Fourier acquisition. Approach. Multi-band imaging was adapted for RG-fMRI to improve the temporal and spatial resolution. To better understand signal dropouts in forward and reverse EPIs, a simple theoretical relationship between echo shift and geometric distortion was derived and validated by the reliable measurements using PSF mapping method. After examining practical imaging protocols for RG-fMRI in three subjects on both a conventional whole-body and a high-performance compact 3 T, the results were compared and the feasibility to further improve the RG-fMRI scheme were explored. High-resolution breath-holding RG-fMRI was conducted with nine subjects on the compact 3 T and the fMRI reliability improvement in high susceptibility brain regions was demonstrated. Finally, reverse Fourier acquisition was applied to RG-fMRI, and its benefit was assessed by a simulation study based on the breath-holding RG-fMRI data. Main results. The temporal and spatial resolution of the multi-band RG-fMRI became feasible for whole-brain fMRI. Echo shift measurements from PSF mapping well estimated signal dropout effects in the EPI pair and were useful to further improve the RG-fMRI scheme. Breath-holding RG-fMRI demonstrated improved fMRI reliability in high susceptibility brain regions. Reverse partial Fourier acquisition omitting the late echoes could further improve the temporal or spatial resolution for RG-fMRI without noticeable signal degradation and spatial resolution loss. Significance. With the improved imaging scheme, RG-fMRI could reliably investigate the functional mechanisms of the human brain in the temporal and frontal areas suffering from susceptibility-induced functional sensitivity loss.

Original languageEnglish (US)
Article number025012
JournalPhysics in medicine and biology
Volume68
Issue number2
DOIs
StatePublished - Jan 21 2023

Keywords

  • EPI distortion
  • geometric distortion
  • point spread function
  • reverse gradient approach
  • signal loss
  • susceptibility artifacts

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Minimizing susceptibility-induced BOLD sensitivity loss in multi-band accelerated fMRI using point spread function mapping and gradient reversal'. Together they form a unique fingerprint.

Cite this