Impaired detection of ventricular tachyarrhythmias by a rate-smoothing algorithm in dual-chamber implantable defibrillators: Intradevice interactions

Michael Glikson, Amy L. Beeman, David M. Luria, David L. Hayes, Paul Andrew Friedman

Research output: Contribution to journalArticle

21 Citations (Scopus)

Abstract

Introduction: Rate smoothing is an algorithm initially designed to prevent rapid changes in pacemaker rates. In this study, we sought to determine the potential of the rate-smoothing mechanism in preventing detection of ventricular tachyarrhythmias. Methods and Results: Clinical testing of rate smoothing was performed at the time of defibrillator arrhythmia induction in 16 patients with implantable defibrillators during 65 episodes of ventricular tachyarrhythmias. We also performed simulator-based testing to assess detection of ventricular tachycardia between 170 and 220 beats/min with systematic sequential change of rate-smoothing percent, AV delay, and maximal rate. During clinical testing of 54 ventricular fibrillation/polymorphic ventricular tachyarrhythmia episodes, there were no cases of nondetection and 3 episodes (5%) of minimally delayed detection. Of 10 monomorphic ventricular tachyarrhythmias, 6 had either delayed (2 cases) or absent (4 cases) detection. During simulator testing, complex interrelationships were demonstrated in AV delay, upper rate, and rate-smoothing percent in determining the severity of the effect on detection. Generally, long AV delay, higher upper rate, and smaller (more aggressive) rate smoothing were associated with increased risk of ventricular tachyarrhythmia underdetection. Importantly, use of parameters that impaired detection was always accompanied by a programmer warning message. Conclusion: Rate smoothing may result in delay or failure of ventricular tachycardia detection. It is important to consider warning messages when programming rate smoothing and to test for appropriate detection when rate smoothing is used despite warning messages.

Original languageEnglish (US)
Pages (from-to)312-318
Number of pages7
JournalJournal of Cardiovascular Electrophysiology
Volume13
Issue number4
StatePublished - 2002

Fingerprint

Implantable Defibrillators
Tachycardia
Ventricular Tachycardia
Defibrillators
Ventricular Fibrillation
Cardiac Arrhythmias

Keywords

  • Dual-chamber defibrillator
  • Implantable cardioverter defibrillator
  • Rate smoothing
  • Ventricular fibrillation detection
  • Ventricular tachycardia detection

ASJC Scopus subject areas

  • Physiology
  • Cardiology and Cardiovascular Medicine

Cite this

Impaired detection of ventricular tachyarrhythmias by a rate-smoothing algorithm in dual-chamber implantable defibrillators : Intradevice interactions. / Glikson, Michael; Beeman, Amy L.; Luria, David M.; Hayes, David L.; Friedman, Paul Andrew.

In: Journal of Cardiovascular Electrophysiology, Vol. 13, No. 4, 2002, p. 312-318.

Research output: Contribution to journalArticle

@article{f58fe3af303c4616a3769218760761de,
title = "Impaired detection of ventricular tachyarrhythmias by a rate-smoothing algorithm in dual-chamber implantable defibrillators: Intradevice interactions",
abstract = "Introduction: Rate smoothing is an algorithm initially designed to prevent rapid changes in pacemaker rates. In this study, we sought to determine the potential of the rate-smoothing mechanism in preventing detection of ventricular tachyarrhythmias. Methods and Results: Clinical testing of rate smoothing was performed at the time of defibrillator arrhythmia induction in 16 patients with implantable defibrillators during 65 episodes of ventricular tachyarrhythmias. We also performed simulator-based testing to assess detection of ventricular tachycardia between 170 and 220 beats/min with systematic sequential change of rate-smoothing percent, AV delay, and maximal rate. During clinical testing of 54 ventricular fibrillation/polymorphic ventricular tachyarrhythmia episodes, there were no cases of nondetection and 3 episodes (5{\%}) of minimally delayed detection. Of 10 monomorphic ventricular tachyarrhythmias, 6 had either delayed (2 cases) or absent (4 cases) detection. During simulator testing, complex interrelationships were demonstrated in AV delay, upper rate, and rate-smoothing percent in determining the severity of the effect on detection. Generally, long AV delay, higher upper rate, and smaller (more aggressive) rate smoothing were associated with increased risk of ventricular tachyarrhythmia underdetection. Importantly, use of parameters that impaired detection was always accompanied by a programmer warning message. Conclusion: Rate smoothing may result in delay or failure of ventricular tachycardia detection. It is important to consider warning messages when programming rate smoothing and to test for appropriate detection when rate smoothing is used despite warning messages.",
keywords = "Dual-chamber defibrillator, Implantable cardioverter defibrillator, Rate smoothing, Ventricular fibrillation detection, Ventricular tachycardia detection",
author = "Michael Glikson and Beeman, {Amy L.} and Luria, {David M.} and Hayes, {David L.} and Friedman, {Paul Andrew}",
year = "2002",
language = "English (US)",
volume = "13",
pages = "312--318",
journal = "Journal of Cardiovascular Electrophysiology",
issn = "1045-3873",
publisher = "Wiley-Blackwell",
number = "4",

}

TY - JOUR

T1 - Impaired detection of ventricular tachyarrhythmias by a rate-smoothing algorithm in dual-chamber implantable defibrillators

T2 - Intradevice interactions

AU - Glikson, Michael

AU - Beeman, Amy L.

AU - Luria, David M.

AU - Hayes, David L.

AU - Friedman, Paul Andrew

PY - 2002

Y1 - 2002

N2 - Introduction: Rate smoothing is an algorithm initially designed to prevent rapid changes in pacemaker rates. In this study, we sought to determine the potential of the rate-smoothing mechanism in preventing detection of ventricular tachyarrhythmias. Methods and Results: Clinical testing of rate smoothing was performed at the time of defibrillator arrhythmia induction in 16 patients with implantable defibrillators during 65 episodes of ventricular tachyarrhythmias. We also performed simulator-based testing to assess detection of ventricular tachycardia between 170 and 220 beats/min with systematic sequential change of rate-smoothing percent, AV delay, and maximal rate. During clinical testing of 54 ventricular fibrillation/polymorphic ventricular tachyarrhythmia episodes, there were no cases of nondetection and 3 episodes (5%) of minimally delayed detection. Of 10 monomorphic ventricular tachyarrhythmias, 6 had either delayed (2 cases) or absent (4 cases) detection. During simulator testing, complex interrelationships were demonstrated in AV delay, upper rate, and rate-smoothing percent in determining the severity of the effect on detection. Generally, long AV delay, higher upper rate, and smaller (more aggressive) rate smoothing were associated with increased risk of ventricular tachyarrhythmia underdetection. Importantly, use of parameters that impaired detection was always accompanied by a programmer warning message. Conclusion: Rate smoothing may result in delay or failure of ventricular tachycardia detection. It is important to consider warning messages when programming rate smoothing and to test for appropriate detection when rate smoothing is used despite warning messages.

AB - Introduction: Rate smoothing is an algorithm initially designed to prevent rapid changes in pacemaker rates. In this study, we sought to determine the potential of the rate-smoothing mechanism in preventing detection of ventricular tachyarrhythmias. Methods and Results: Clinical testing of rate smoothing was performed at the time of defibrillator arrhythmia induction in 16 patients with implantable defibrillators during 65 episodes of ventricular tachyarrhythmias. We also performed simulator-based testing to assess detection of ventricular tachycardia between 170 and 220 beats/min with systematic sequential change of rate-smoothing percent, AV delay, and maximal rate. During clinical testing of 54 ventricular fibrillation/polymorphic ventricular tachyarrhythmia episodes, there were no cases of nondetection and 3 episodes (5%) of minimally delayed detection. Of 10 monomorphic ventricular tachyarrhythmias, 6 had either delayed (2 cases) or absent (4 cases) detection. During simulator testing, complex interrelationships were demonstrated in AV delay, upper rate, and rate-smoothing percent in determining the severity of the effect on detection. Generally, long AV delay, higher upper rate, and smaller (more aggressive) rate smoothing were associated with increased risk of ventricular tachyarrhythmia underdetection. Importantly, use of parameters that impaired detection was always accompanied by a programmer warning message. Conclusion: Rate smoothing may result in delay or failure of ventricular tachycardia detection. It is important to consider warning messages when programming rate smoothing and to test for appropriate detection when rate smoothing is used despite warning messages.

KW - Dual-chamber defibrillator

KW - Implantable cardioverter defibrillator

KW - Rate smoothing

KW - Ventricular fibrillation detection

KW - Ventricular tachycardia detection

UR - http://www.scopus.com/inward/record.url?scp=0036107976&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0036107976&partnerID=8YFLogxK

M3 - Article

C2 - 12033344

AN - SCOPUS:0036107976

VL - 13

SP - 312

EP - 318

JO - Journal of Cardiovascular Electrophysiology

JF - Journal of Cardiovascular Electrophysiology

SN - 1045-3873

IS - 4

ER -