Blood flow velocity measurements: A comparison of 25 clinical ultrasonographic units

D. S. Groth, F. E. Zink, J. P. Felmlee, J. M. Kofler, E. M. James, J. R. Lindsey, W. Pavlicek

Research output: Contribution to journalArticlepeer-review

8 Scopus citations

Abstract

A blood-mimicking flow phantom was used to evaluate the precision of velocity measurements acquired using 25 pulsed Doppler ultrasonographic units from four vendors. Measurements were made at four constant flow rates (12 to 50 cm/s peak velocity). The average standard deviation values of the peak and time-averaged velocities among all units and all flow rates were found to be 7 and 9% of the mean, respectively, while the corresponding values for a subgroup of 20 identical units were 5 and 8%. Considered in conjunction with other published data, this suggests that units should be calibrated to an institutional standard at the time of acceptance testing.

Original languageEnglish (US)
Pages (from-to)273-277
Number of pages5
JournalJournal of Ultrasound in Medicine
Volume14
Issue number4
DOIs
StatePublished - 1995

Keywords

  • Calibration
  • Quality assurance
  • Ultrasonographic Doppler studies
  • Ultrasonographic utilization

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging