The intersection of two ultrasound beams with slightly different frequencies results in generation of a localized radiation force and stimulates emission of audio signals from targeted objects. Vibro-acoustography uses this phenomenon to probe elastic properties of objects. Vibro-acoustography of contrast microbubbles in degassed water produced quantitative flow measurements from analysis of their acoustic emission. We used a dual-beam transducer generating bursts of 40-kHz vibrations. The vibrations resulted from interference of 3.48-MHz and 3.52-MHz confocal beams intersecting at the center of a thin plastic conduit. We tested flows of 13, 48, 85, and 120 mL/min of contrast microbubbles at concentrations from 1.2 × 105 to 6 × 105 bubbles/mL. The amplitude of the acoustic emission was linear with microbubble concentrations up to a value of 3.6 × 105 bubbles/mL. A replenishment method for microbubble contrast and flow rate analysis was used with radiation force bursts deployed at 0.05, 0.1, 0.2, 0.5, 1, and 2-second pulsing intervals. The relation between the pulsing intervals and the peak amplitude was fitted by an exponential curve and a rate constant calculated for each tested flow rate. The rate constant values were linearly correlated with the tested flows. The vibro-acoustography method provides objective, quantitative, and highly-localized assessment of flow using contrast microbubbles.

Original languageEnglish (US)
Pages (from-to)249-256
Number of pages8
JournalUltrasonic Imaging
Volume23
Issue number4
StatePublished - 2001

Fingerprint

Elasticity Imaging Techniques
Microbubbles
Acoustics
low frequencies
acoustics
acoustic emission
bursts
Vibration
bubbles
flow velocity
intervals
replenishment
vibration
audio signals
flow measurement
Radiation
radiation
intersections
transducers
Transducers

Keywords

  • Contrast media
  • Flow
  • Perfusion
  • Ultrasonics

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Acoustics and Ultrasonics

Cite this

Vibro-acoustography : Quantification of flow with highly-localized low-frequency acoustic force. / Belohlavek, Marek; Asanuma, Toshihiko; Kinnick, Randall R.; Greenleaf, James F.

In: Ultrasonic Imaging, Vol. 23, No. 4, 2001, p. 249-256.

Research output: Contribution to journalArticle

@article{5a711085d8874d8cbfb597857c62c3c9,
title = "Vibro-acoustography: Quantification of flow with highly-localized low-frequency acoustic force",
abstract = "The intersection of two ultrasound beams with slightly different frequencies results in generation of a localized radiation force and stimulates emission of audio signals from targeted objects. Vibro-acoustography uses this phenomenon to probe elastic properties of objects. Vibro-acoustography of contrast microbubbles in degassed water produced quantitative flow measurements from analysis of their acoustic emission. We used a dual-beam transducer generating bursts of 40-kHz vibrations. The vibrations resulted from interference of 3.48-MHz and 3.52-MHz confocal beams intersecting at the center of a thin plastic conduit. We tested flows of 13, 48, 85, and 120 mL/min of contrast microbubbles at concentrations from 1.2 × 105 to 6 × 105 bubbles/mL. The amplitude of the acoustic emission was linear with microbubble concentrations up to a value of 3.6 × 105 bubbles/mL. A replenishment method for microbubble contrast and flow rate analysis was used with radiation force bursts deployed at 0.05, 0.1, 0.2, 0.5, 1, and 2-second pulsing intervals. The relation between the pulsing intervals and the peak amplitude was fitted by an exponential curve and a rate constant calculated for each tested flow rate. The rate constant values were linearly correlated with the tested flows. The vibro-acoustography method provides objective, quantitative, and highly-localized assessment of flow using contrast microbubbles.",
keywords = "Contrast media, Flow, Perfusion, Ultrasonics",
author = "Marek Belohlavek and Toshihiko Asanuma and Kinnick, {Randall R.} and Greenleaf, {James F}",
year = "2001",
language = "English (US)",
volume = "23",
pages = "249--256",
journal = "Ultrasonic Imaging",
issn = "0161-7346",
publisher = "Dynamedia Inc.",
number = "4",

}

TY - JOUR

T1 - Vibro-acoustography

T2 - Quantification of flow with highly-localized low-frequency acoustic force

AU - Belohlavek, Marek

AU - Asanuma, Toshihiko

AU - Kinnick, Randall R.

AU - Greenleaf, James F

PY - 2001

Y1 - 2001

N2 - The intersection of two ultrasound beams with slightly different frequencies results in generation of a localized radiation force and stimulates emission of audio signals from targeted objects. Vibro-acoustography uses this phenomenon to probe elastic properties of objects. Vibro-acoustography of contrast microbubbles in degassed water produced quantitative flow measurements from analysis of their acoustic emission. We used a dual-beam transducer generating bursts of 40-kHz vibrations. The vibrations resulted from interference of 3.48-MHz and 3.52-MHz confocal beams intersecting at the center of a thin plastic conduit. We tested flows of 13, 48, 85, and 120 mL/min of contrast microbubbles at concentrations from 1.2 × 105 to 6 × 105 bubbles/mL. The amplitude of the acoustic emission was linear with microbubble concentrations up to a value of 3.6 × 105 bubbles/mL. A replenishment method for microbubble contrast and flow rate analysis was used with radiation force bursts deployed at 0.05, 0.1, 0.2, 0.5, 1, and 2-second pulsing intervals. The relation between the pulsing intervals and the peak amplitude was fitted by an exponential curve and a rate constant calculated for each tested flow rate. The rate constant values were linearly correlated with the tested flows. The vibro-acoustography method provides objective, quantitative, and highly-localized assessment of flow using contrast microbubbles.

AB - The intersection of two ultrasound beams with slightly different frequencies results in generation of a localized radiation force and stimulates emission of audio signals from targeted objects. Vibro-acoustography uses this phenomenon to probe elastic properties of objects. Vibro-acoustography of contrast microbubbles in degassed water produced quantitative flow measurements from analysis of their acoustic emission. We used a dual-beam transducer generating bursts of 40-kHz vibrations. The vibrations resulted from interference of 3.48-MHz and 3.52-MHz confocal beams intersecting at the center of a thin plastic conduit. We tested flows of 13, 48, 85, and 120 mL/min of contrast microbubbles at concentrations from 1.2 × 105 to 6 × 105 bubbles/mL. The amplitude of the acoustic emission was linear with microbubble concentrations up to a value of 3.6 × 105 bubbles/mL. A replenishment method for microbubble contrast and flow rate analysis was used with radiation force bursts deployed at 0.05, 0.1, 0.2, 0.5, 1, and 2-second pulsing intervals. The relation between the pulsing intervals and the peak amplitude was fitted by an exponential curve and a rate constant calculated for each tested flow rate. The rate constant values were linearly correlated with the tested flows. The vibro-acoustography method provides objective, quantitative, and highly-localized assessment of flow using contrast microbubbles.

KW - Contrast media

KW - Flow

KW - Perfusion

KW - Ultrasonics

UR - http://www.scopus.com/inward/record.url?scp=0035739686&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0035739686&partnerID=8YFLogxK

M3 - Article

C2 - 12051277

AN - SCOPUS:0035739686

VL - 23

SP - 249

EP - 256

JO - Ultrasonic Imaging

JF - Ultrasonic Imaging

SN - 0161-7346

IS - 4

ER -