TY - JOUR
T1 - Validation of an automated latex particle-enhanced immunoturbidimetric von Willebrand factor activity assay
AU - Chen, D.
AU - Tange, J. I.
AU - Meyers, B. J.
AU - Pruthi, R. K.
AU - Nichols, W. L.
AU - Heit, J. A.
PY - 2011/10
Y1 - 2011/10
N2 - Background:Laboratory diagnosis of von Willebrand disease (VWD) requires accurate measurement of plasma von Willebrand factor (VWF) activity. Objectives:To evaluate laboratory characteristics, diagnostic accuracy and testing utilities of an automated latex particle-enhanced immunoturbidimetric VWF assay (VWF:Lx) based on a monoclonal antibody recognizing the VWF-platelet glycoprotein (GP) Ib binding domain. Methods:Laboratory characteristics including lower detection limit, linearity, precision, sample stability, and method comparison between VWF:Lx and VWF ristocetin cofactor activity by platelet aggregometry (VWF:RCo) were examined. To assess VWF:Lx diagnostic accuracy, 492 patient plasma samples, including 40 previously characterized VWD patient samples, were tested for VWF antigen (VWF:Ag) and VWF:RCo by either aggregometry or flow cytometry, and VWF:Lx with supplemental VWF multimer analysis when indicated. Based on results of VWF:Ag, VWF:RCo and VWF multimer analysis, and available clinical information, samples were categorized as: normal; VWD types 1, 2A/B, 2M, or severe 1 vs. 2M; or acquired VWF abnormalities (AVWA) due to subtle loss of highest molecular weight multimers. Results:VWF:Lx had excellent laboratory characteristics and linear correlation with VWF:RCo (R 2=0.93). VWF:Lx accurately classified virtually all normal and VWD patient samples. Compared with VWF:RCo, VWF:Lx had superior sensitivity and specificity for distinguishing severe type 1 vs. 2M VWD and identifying AVWA. A proposed screening panel comprising VWF:Ag and VWF:Lx had 100% and 83% sensitivity for detecting VWD and AVWA, respectively. Conclusions:VWF:Lx has excellent laboratory characteristics and diagnostic accuracy compared with VWF:RCo, and can be used as part of an initial VWD screening panel and as a supplementary test.
AB - Background:Laboratory diagnosis of von Willebrand disease (VWD) requires accurate measurement of plasma von Willebrand factor (VWF) activity. Objectives:To evaluate laboratory characteristics, diagnostic accuracy and testing utilities of an automated latex particle-enhanced immunoturbidimetric VWF assay (VWF:Lx) based on a monoclonal antibody recognizing the VWF-platelet glycoprotein (GP) Ib binding domain. Methods:Laboratory characteristics including lower detection limit, linearity, precision, sample stability, and method comparison between VWF:Lx and VWF ristocetin cofactor activity by platelet aggregometry (VWF:RCo) were examined. To assess VWF:Lx diagnostic accuracy, 492 patient plasma samples, including 40 previously characterized VWD patient samples, were tested for VWF antigen (VWF:Ag) and VWF:RCo by either aggregometry or flow cytometry, and VWF:Lx with supplemental VWF multimer analysis when indicated. Based on results of VWF:Ag, VWF:RCo and VWF multimer analysis, and available clinical information, samples were categorized as: normal; VWD types 1, 2A/B, 2M, or severe 1 vs. 2M; or acquired VWF abnormalities (AVWA) due to subtle loss of highest molecular weight multimers. Results:VWF:Lx had excellent laboratory characteristics and linear correlation with VWF:RCo (R 2=0.93). VWF:Lx accurately classified virtually all normal and VWD patient samples. Compared with VWF:RCo, VWF:Lx had superior sensitivity and specificity for distinguishing severe type 1 vs. 2M VWD and identifying AVWA. A proposed screening panel comprising VWF:Ag and VWF:Lx had 100% and 83% sensitivity for detecting VWD and AVWA, respectively. Conclusions:VWF:Lx has excellent laboratory characteristics and diagnostic accuracy compared with VWF:RCo, and can be used as part of an initial VWD screening panel and as a supplementary test.
KW - Latex immunoassay
KW - Ristocetin cofactor activity
KW - Von Willebrand disease
KW - Von Willebrand factor activity
UR - http://www.scopus.com/inward/record.url?scp=80053417668&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=80053417668&partnerID=8YFLogxK
U2 - 10.1111/j.1538-7836.2011.04460.x
DO - 10.1111/j.1538-7836.2011.04460.x
M3 - Article
C2 - 21824283
AN - SCOPUS:80053417668
SN - 1538-7933
VL - 9
SP - 1993
EP - 2002
JO - Journal of Thrombosis and Haemostasis
JF - Journal of Thrombosis and Haemostasis
IS - 10
ER -