Abstract
PURPOSE: To evaluate image quality, vessel conspicuity, and signal-to- noise ratio (S/N) of four two-dimensional magnetic resonance (MR) phase- contrast techniques: cine phase-contrast with respiratory compensation (CPC- R), cine phase-contrast without respiratory compensation (CPC), phase- contrast with two signals acquired (PC-2), and phase-contrast with eight signals acquired (PC-8). MATERIALS AND METHODS: MR angiographic images of the portal vein were obtained in 14 patients. RESULTS: Overall image quality was rated excellent in 10 of 14 patients (CPC-R), in one of 14 patients (CPC), in none of 14 patients (PC-2), and in one of 14 patients (PC-8). Views of 10 upper abdominal vessels were evaluated with each technique. The mean (± standard deviation) number of well-imaged vessels was 6 ± 2 (CPC-R), 3 ± 2 (CPC), 2 ± 2 (PC-2), and 3 ± 2 (PC-8). CPC-R images were significantly better in overall image quality and number of vessels seen (P < .01). Mean S/Ns were 10.8 ± 3.4 (CPC-R), 6.8 ± 2.7 (CPC), 4.5 ± 1.7 (PC-2), and 6.4 ± 2.1 (PC-8). S/N with CPC-R was significantly greater than with CPC, PC-2, or PC-8. CONCLUSION: CPC-R provided consistently higher quality angiographic images than were obtained with the other sequences studied.
Original language | English (US) |
---|---|
Pages (from-to) | 101-105 |
Number of pages | 5 |
Journal | Radiology |
Volume | 195 |
Issue number | 1 |
DOIs | |
State | Published - Jan 1 1995 |
Keywords
- Abdomen, MR
- Magnetic resonance (MR), phase imaging
- Magnetic resonance (MR), technology
- Magnetic resonance (MR), vascular studies
ASJC Scopus subject areas
- Radiology Nuclear Medicine and imaging