Undersampled elliptical centric view-order for improved spatial resolution in contrast-enhanced MR angiography

Ananth J. Madhuranthakam, Houchun H. Hu, Andrew V. Barger, Clifton R Haider, David G. Kruger, James Glockner, John III Huston, Stephen J Riederer

Research output: Contribution to journalArticle

27 Citations (Scopus)

Abstract

Although contrast-enhanced MR angiography (CE-MRA) has been successfully developed into a routine clinical imaging technique, there is still need for improved spatial resolution in a given acquisition time. Undersampled projection reconstruction (PR) techniques maintain spatial resolution with reduced scan times, and the elliptical centric (EC) view order provides high quality arterial phase images without venous contamination. In this work, we present a hybrid elliptical centric-projection reconstruction (EC-PR) technique to provide spatial resolution improvement over standard EC in a given time. The k-space sampling was performed by undersampling the periphery of the k Y-kZ phase encoding plane of an EC view order in a PR like manner. The sampled views were maintained on a rectilinear grid, and thus reconstructed by standard 3DFT. The non-sampled views were compensated either by zero-filling or performing a 2D homodyne reconstruction. Compared to a fully sampled k-space, the EC-PR sequence acquired in the same scan time provides a resolution improvement of about two, as shown by point spread function analysis and phantom experiments. The hypothesis that EC-PR provides improved resolution while retaining diagnostically adequate SNR was tested in 11 CEMRA studies of the popliteal and carotid arteries and shown to be true (P < 0.03).

Original languageEnglish (US)
Pages (from-to)50-58
Number of pages9
JournalMagnetic Resonance in Medicine
Volume55
Issue number1
DOIs
StatePublished - Jan 2006

Fingerprint

Angiography
Popliteal Artery
Carotid Arteries

Keywords

  • 2D homodyne
  • Elliptical-centric
  • MR angiography
  • Partial fourier
  • Projection reconstruction

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Undersampled elliptical centric view-order for improved spatial resolution in contrast-enhanced MR angiography. / Madhuranthakam, Ananth J.; Hu, Houchun H.; Barger, Andrew V.; Haider, Clifton R; Kruger, David G.; Glockner, James; Huston, John III; Riederer, Stephen J.

In: Magnetic Resonance in Medicine, Vol. 55, No. 1, 01.2006, p. 50-58.

Research output: Contribution to journalArticle

@article{d3082f6e74934aa5adc1c2d0710e7e01,
title = "Undersampled elliptical centric view-order for improved spatial resolution in contrast-enhanced MR angiography",
abstract = "Although contrast-enhanced MR angiography (CE-MRA) has been successfully developed into a routine clinical imaging technique, there is still need for improved spatial resolution in a given acquisition time. Undersampled projection reconstruction (PR) techniques maintain spatial resolution with reduced scan times, and the elliptical centric (EC) view order provides high quality arterial phase images without venous contamination. In this work, we present a hybrid elliptical centric-projection reconstruction (EC-PR) technique to provide spatial resolution improvement over standard EC in a given time. The k-space sampling was performed by undersampling the periphery of the k Y-kZ phase encoding plane of an EC view order in a PR like manner. The sampled views were maintained on a rectilinear grid, and thus reconstructed by standard 3DFT. The non-sampled views were compensated either by zero-filling or performing a 2D homodyne reconstruction. Compared to a fully sampled k-space, the EC-PR sequence acquired in the same scan time provides a resolution improvement of about two, as shown by point spread function analysis and phantom experiments. The hypothesis that EC-PR provides improved resolution while retaining diagnostically adequate SNR was tested in 11 CEMRA studies of the popliteal and carotid arteries and shown to be true (P < 0.03).",
keywords = "2D homodyne, Elliptical-centric, MR angiography, Partial fourier, Projection reconstruction",
author = "Madhuranthakam, {Ananth J.} and Hu, {Houchun H.} and Barger, {Andrew V.} and Haider, {Clifton R} and Kruger, {David G.} and James Glockner and Huston, {John III} and Riederer, {Stephen J}",
year = "2006",
month = "1",
doi = "10.1002/mrm.20726",
language = "English (US)",
volume = "55",
pages = "50--58",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",
number = "1",

}

TY - JOUR

T1 - Undersampled elliptical centric view-order for improved spatial resolution in contrast-enhanced MR angiography

AU - Madhuranthakam, Ananth J.

AU - Hu, Houchun H.

AU - Barger, Andrew V.

AU - Haider, Clifton R

AU - Kruger, David G.

AU - Glockner, James

AU - Huston, John III

AU - Riederer, Stephen J

PY - 2006/1

Y1 - 2006/1

N2 - Although contrast-enhanced MR angiography (CE-MRA) has been successfully developed into a routine clinical imaging technique, there is still need for improved spatial resolution in a given acquisition time. Undersampled projection reconstruction (PR) techniques maintain spatial resolution with reduced scan times, and the elliptical centric (EC) view order provides high quality arterial phase images without venous contamination. In this work, we present a hybrid elliptical centric-projection reconstruction (EC-PR) technique to provide spatial resolution improvement over standard EC in a given time. The k-space sampling was performed by undersampling the periphery of the k Y-kZ phase encoding plane of an EC view order in a PR like manner. The sampled views were maintained on a rectilinear grid, and thus reconstructed by standard 3DFT. The non-sampled views were compensated either by zero-filling or performing a 2D homodyne reconstruction. Compared to a fully sampled k-space, the EC-PR sequence acquired in the same scan time provides a resolution improvement of about two, as shown by point spread function analysis and phantom experiments. The hypothesis that EC-PR provides improved resolution while retaining diagnostically adequate SNR was tested in 11 CEMRA studies of the popliteal and carotid arteries and shown to be true (P < 0.03).

AB - Although contrast-enhanced MR angiography (CE-MRA) has been successfully developed into a routine clinical imaging technique, there is still need for improved spatial resolution in a given acquisition time. Undersampled projection reconstruction (PR) techniques maintain spatial resolution with reduced scan times, and the elliptical centric (EC) view order provides high quality arterial phase images without venous contamination. In this work, we present a hybrid elliptical centric-projection reconstruction (EC-PR) technique to provide spatial resolution improvement over standard EC in a given time. The k-space sampling was performed by undersampling the periphery of the k Y-kZ phase encoding plane of an EC view order in a PR like manner. The sampled views were maintained on a rectilinear grid, and thus reconstructed by standard 3DFT. The non-sampled views were compensated either by zero-filling or performing a 2D homodyne reconstruction. Compared to a fully sampled k-space, the EC-PR sequence acquired in the same scan time provides a resolution improvement of about two, as shown by point spread function analysis and phantom experiments. The hypothesis that EC-PR provides improved resolution while retaining diagnostically adequate SNR was tested in 11 CEMRA studies of the popliteal and carotid arteries and shown to be true (P < 0.03).

KW - 2D homodyne

KW - Elliptical-centric

KW - MR angiography

KW - Partial fourier

KW - Projection reconstruction

UR - http://www.scopus.com/inward/record.url?scp=30344458822&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=30344458822&partnerID=8YFLogxK

U2 - 10.1002/mrm.20726

DO - 10.1002/mrm.20726

M3 - Article

C2 - 16315207

AN - SCOPUS:30344458822

VL - 55

SP - 50

EP - 58

JO - Magnetic Resonance in Medicine

JF - Magnetic Resonance in Medicine

SN - 0740-3194

IS - 1

ER -