TY - JOUR
T1 - Three-dimensional ultrasound for image-guided mapping and intervention
T2 - methods, quantitative validation, and clinical feasibility of a novel multimodality image mapping system.
AU - Okumura, Yasuo
AU - Henz, Benhur D.
AU - Johnson, Susan B.
AU - Bunch, T. Jared
AU - O'Brien, Christine J.
AU - Hodge, David O.
AU - Altman, Andres
AU - Govari, Assaf
AU - Packer, Douglas L.
PY - 2008/6/1
Y1 - 2008/6/1
N2 - BACKGROUND: Multiple factors create discrepancies between electroanatomic maps and merged, preacquired computed tomographic images used in guiding atrial fibrillation ablation. Therefore, a Carto-based 3D ultrasound image system (Biosense Webster Inc) was validated in an animal model and tested in 15 atrial fibrillation patients. METHODS AND RESULTS: Twelve dogs underwent evaluation using a newly developed Carto-based 3D ultrasound system. After fiducial clip markers were percutaneously implanted at critical locations in each cardiac chamber, 3D ultrasound geometries, derived from a family of 2D intracardiac echocardiographic images, were constructed. Point-source error of 3D ultrasound-derived geometries, assessed by actual real-time 2D intracardiac echocardiographic clip sites, was 2.1+/-1.1 mm for atrial and 2.4+/-1.2 mm for ventricular sites. These errors were significantly less than the variance on CartoMerge computed tomographic images (atria: 3.3+/-1.6 mm; ventricles: 4.8+/-2.0 mm; P<0.001 for both). Target ablation at each clip, guided only by 3D ultrasound-derived geometry, resulted in lesions within 1.1+/-1.1 mm of the actual clips. Pulmonary vein ablation guided by 3D ultrasound-derived geometry resulted in circumferential ablative lesions. Mapping in 15 patients produced modestly smaller 3D ultrasound versus electroanatomic map left atrial volumes (98+/-24 cm(3) versus 109+/-25 cm(3), P<0.05). Three-dimensional ultrasound-guided pulmonary vein isolation and linear ablation in these patients were successfully performed with confirmation of pulmonary vein entrance/exit block. CONCLUSIONS: These data demonstrate that 3D ultrasound images seamlessly yield anatomically accurate chamber geometries. Image volumes from the ultrasound system are more accurate than possible with CartoMerge computed tomographic imaging. This clinical study also demonstrates the initial feasibility of this guidance system for ablation in patients with atrial fibrillation.
AB - BACKGROUND: Multiple factors create discrepancies between electroanatomic maps and merged, preacquired computed tomographic images used in guiding atrial fibrillation ablation. Therefore, a Carto-based 3D ultrasound image system (Biosense Webster Inc) was validated in an animal model and tested in 15 atrial fibrillation patients. METHODS AND RESULTS: Twelve dogs underwent evaluation using a newly developed Carto-based 3D ultrasound system. After fiducial clip markers were percutaneously implanted at critical locations in each cardiac chamber, 3D ultrasound geometries, derived from a family of 2D intracardiac echocardiographic images, were constructed. Point-source error of 3D ultrasound-derived geometries, assessed by actual real-time 2D intracardiac echocardiographic clip sites, was 2.1+/-1.1 mm for atrial and 2.4+/-1.2 mm for ventricular sites. These errors were significantly less than the variance on CartoMerge computed tomographic images (atria: 3.3+/-1.6 mm; ventricles: 4.8+/-2.0 mm; P<0.001 for both). Target ablation at each clip, guided only by 3D ultrasound-derived geometry, resulted in lesions within 1.1+/-1.1 mm of the actual clips. Pulmonary vein ablation guided by 3D ultrasound-derived geometry resulted in circumferential ablative lesions. Mapping in 15 patients produced modestly smaller 3D ultrasound versus electroanatomic map left atrial volumes (98+/-24 cm(3) versus 109+/-25 cm(3), P<0.05). Three-dimensional ultrasound-guided pulmonary vein isolation and linear ablation in these patients were successfully performed with confirmation of pulmonary vein entrance/exit block. CONCLUSIONS: These data demonstrate that 3D ultrasound images seamlessly yield anatomically accurate chamber geometries. Image volumes from the ultrasound system are more accurate than possible with CartoMerge computed tomographic imaging. This clinical study also demonstrates the initial feasibility of this guidance system for ablation in patients with atrial fibrillation.
UR - http://www.scopus.com/inward/record.url?scp=53049085427&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=53049085427&partnerID=8YFLogxK
U2 - 10.1161/CIRCEP.108.769935
DO - 10.1161/CIRCEP.108.769935
M3 - Article
C2 - 19808401
AN - SCOPUS:53049085427
SN - 1941-3149
VL - 1
SP - 110
EP - 119
JO - Circulation: Arrhythmia and Electrophysiology
JF - Circulation: Arrhythmia and Electrophysiology
IS - 2
ER -