Technical considerations when intervening with coronary device catheters in the vicinity of previously deployed stents

J. Aaron Grantham, Daniel J. Tiede, David Holmes

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

As stent use increases, interventional cardiologists are increasingly faced with patients that require procedures in the vicinity of previously deployed stents. We present two cases of side-branch interventions in the vicinity of previously deployed stents where devices were trapped by the stent. In each case, traction on the device resulted in stent dislodgment. The stents were successfully extracted and replaced without complications.

Original languageEnglish (US)
Pages (from-to)214-217
Number of pages4
JournalCatheterization and Cardiovascular Interventions
Volume52
Issue number2
DOIs
StatePublished - Feb 14 2001
Externally publishedYes

Fingerprint

Stents
Catheters
Equipment and Supplies
Traction

Keywords

  • Balloon angioplasty
  • Percutaneous coronary intervention
  • Stenting

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Radiology Nuclear Medicine and imaging

Cite this

Technical considerations when intervening with coronary device catheters in the vicinity of previously deployed stents. / Aaron Grantham, J.; Tiede, Daniel J.; Holmes, David.

In: Catheterization and Cardiovascular Interventions, Vol. 52, No. 2, 14.02.2001, p. 214-217.

Research output: Contribution to journalArticle

@article{344813f80a6b44a9b04c34d86c06d40d,
title = "Technical considerations when intervening with coronary device catheters in the vicinity of previously deployed stents",
abstract = "As stent use increases, interventional cardiologists are increasingly faced with patients that require procedures in the vicinity of previously deployed stents. We present two cases of side-branch interventions in the vicinity of previously deployed stents where devices were trapped by the stent. In each case, traction on the device resulted in stent dislodgment. The stents were successfully extracted and replaced without complications.",
keywords = "Balloon angioplasty, Percutaneous coronary intervention, Stenting",
author = "{Aaron Grantham}, J. and Tiede, {Daniel J.} and David Holmes",
year = "2001",
month = "2",
day = "14",
doi = "10.1002/1522-726X(200102)52:2<214::AID-CCD1051>3.0.CO;2-4",
language = "English (US)",
volume = "52",
pages = "214--217",
journal = "Catheterization and Cardiovascular Interventions",
issn = "1522-1946",
publisher = "Wiley-Liss Inc.",
number = "2",

}

TY - JOUR

T1 - Technical considerations when intervening with coronary device catheters in the vicinity of previously deployed stents

AU - Aaron Grantham, J.

AU - Tiede, Daniel J.

AU - Holmes, David

PY - 2001/2/14

Y1 - 2001/2/14

N2 - As stent use increases, interventional cardiologists are increasingly faced with patients that require procedures in the vicinity of previously deployed stents. We present two cases of side-branch interventions in the vicinity of previously deployed stents where devices were trapped by the stent. In each case, traction on the device resulted in stent dislodgment. The stents were successfully extracted and replaced without complications.

AB - As stent use increases, interventional cardiologists are increasingly faced with patients that require procedures in the vicinity of previously deployed stents. We present two cases of side-branch interventions in the vicinity of previously deployed stents where devices were trapped by the stent. In each case, traction on the device resulted in stent dislodgment. The stents were successfully extracted and replaced without complications.

KW - Balloon angioplasty

KW - Percutaneous coronary intervention

KW - Stenting

UR - http://www.scopus.com/inward/record.url?scp=0035141393&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0035141393&partnerID=8YFLogxK

U2 - 10.1002/1522-726X(200102)52:2<214::AID-CCD1051>3.0.CO;2-4

DO - 10.1002/1522-726X(200102)52:2<214::AID-CCD1051>3.0.CO;2-4

M3 - Article

C2 - 11170332

AN - SCOPUS:0035141393

VL - 52

SP - 214

EP - 217

JO - Catheterization and Cardiovascular Interventions

JF - Catheterization and Cardiovascular Interventions

SN - 1522-1946

IS - 2