Abstract

An imaging method is introduced in which the signal in MR images is affected by the stiffness distribution in the object being imaged. Intravoxel phase dispersion (IVPD) that occurs during MR elastography (MRE) acquisitions decreases the signal in soft regions more than in stiff regions due to changes in shear wave amplitude and wavelength. The IVPD effect is enhanced by lowpass filtering the MR k-space data with a circular Gaussian lowpass filter. A processing method is introduced to take the time series of MRE magnitude images with IVPD and produce a final stiffness-weighted image (SWI) by calculating the minimum signal at each pixel from a small number of temporal samples. The SWI technique is demonstrated in phantom studies as well as in the case of a preserved postmortem breast tissue specimen with a stiff lesion created by focused ultrasound ablation to mimic a breast cancer. When free of significant sources of depth-dependent wave attenuation, interference, and boundary effects, SWI is a simple, fast, qualitative technique that does not require the use of phase unwrapping or inversion algorithms for localizing stiff regions in an object.

Original languageEnglish (US)
Pages (from-to)59-67
Number of pages9
JournalMagnetic Resonance in Medicine
Volume55
Issue number1
DOIs
StatePublished - Jan 2006

Fingerprint

Elasticity Imaging Techniques
Magnetic Resonance Imaging
Breast
Breast Neoplasms

Keywords

  • Elasticity
  • Intravoxel phase dispersion
  • Magnetic resonance elastography
  • Stiffness-weighted imaging

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Stiffness-weighted magnetic resonance imaging. / Glaser, Kevin J.; Felmlee, Joel P.; Manduca, Armando; Mariappan, Yogesh Kannan; Ehman, Richard Lorne.

In: Magnetic Resonance in Medicine, Vol. 55, No. 1, 01.2006, p. 59-67.

Research output: Contribution to journalArticle

Glaser, Kevin J. ; Felmlee, Joel P. ; Manduca, Armando ; Mariappan, Yogesh Kannan ; Ehman, Richard Lorne. / Stiffness-weighted magnetic resonance imaging. In: Magnetic Resonance in Medicine. 2006 ; Vol. 55, No. 1. pp. 59-67.
@article{622be1d4fd06493882e165410ebcb65c,
title = "Stiffness-weighted magnetic resonance imaging",
abstract = "An imaging method is introduced in which the signal in MR images is affected by the stiffness distribution in the object being imaged. Intravoxel phase dispersion (IVPD) that occurs during MR elastography (MRE) acquisitions decreases the signal in soft regions more than in stiff regions due to changes in shear wave amplitude and wavelength. The IVPD effect is enhanced by lowpass filtering the MR k-space data with a circular Gaussian lowpass filter. A processing method is introduced to take the time series of MRE magnitude images with IVPD and produce a final stiffness-weighted image (SWI) by calculating the minimum signal at each pixel from a small number of temporal samples. The SWI technique is demonstrated in phantom studies as well as in the case of a preserved postmortem breast tissue specimen with a stiff lesion created by focused ultrasound ablation to mimic a breast cancer. When free of significant sources of depth-dependent wave attenuation, interference, and boundary effects, SWI is a simple, fast, qualitative technique that does not require the use of phase unwrapping or inversion algorithms for localizing stiff regions in an object.",
keywords = "Elasticity, Intravoxel phase dispersion, Magnetic resonance elastography, Stiffness-weighted imaging",
author = "Glaser, {Kevin J.} and Felmlee, {Joel P.} and Armando Manduca and Mariappan, {Yogesh Kannan} and Ehman, {Richard Lorne}",
year = "2006",
month = "1",
doi = "10.1002/mrm.20748",
language = "English (US)",
volume = "55",
pages = "59--67",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",
number = "1",

}

TY - JOUR

T1 - Stiffness-weighted magnetic resonance imaging

AU - Glaser, Kevin J.

AU - Felmlee, Joel P.

AU - Manduca, Armando

AU - Mariappan, Yogesh Kannan

AU - Ehman, Richard Lorne

PY - 2006/1

Y1 - 2006/1

N2 - An imaging method is introduced in which the signal in MR images is affected by the stiffness distribution in the object being imaged. Intravoxel phase dispersion (IVPD) that occurs during MR elastography (MRE) acquisitions decreases the signal in soft regions more than in stiff regions due to changes in shear wave amplitude and wavelength. The IVPD effect is enhanced by lowpass filtering the MR k-space data with a circular Gaussian lowpass filter. A processing method is introduced to take the time series of MRE magnitude images with IVPD and produce a final stiffness-weighted image (SWI) by calculating the minimum signal at each pixel from a small number of temporal samples. The SWI technique is demonstrated in phantom studies as well as in the case of a preserved postmortem breast tissue specimen with a stiff lesion created by focused ultrasound ablation to mimic a breast cancer. When free of significant sources of depth-dependent wave attenuation, interference, and boundary effects, SWI is a simple, fast, qualitative technique that does not require the use of phase unwrapping or inversion algorithms for localizing stiff regions in an object.

AB - An imaging method is introduced in which the signal in MR images is affected by the stiffness distribution in the object being imaged. Intravoxel phase dispersion (IVPD) that occurs during MR elastography (MRE) acquisitions decreases the signal in soft regions more than in stiff regions due to changes in shear wave amplitude and wavelength. The IVPD effect is enhanced by lowpass filtering the MR k-space data with a circular Gaussian lowpass filter. A processing method is introduced to take the time series of MRE magnitude images with IVPD and produce a final stiffness-weighted image (SWI) by calculating the minimum signal at each pixel from a small number of temporal samples. The SWI technique is demonstrated in phantom studies as well as in the case of a preserved postmortem breast tissue specimen with a stiff lesion created by focused ultrasound ablation to mimic a breast cancer. When free of significant sources of depth-dependent wave attenuation, interference, and boundary effects, SWI is a simple, fast, qualitative technique that does not require the use of phase unwrapping or inversion algorithms for localizing stiff regions in an object.

KW - Elasticity

KW - Intravoxel phase dispersion

KW - Magnetic resonance elastography

KW - Stiffness-weighted imaging

UR - http://www.scopus.com/inward/record.url?scp=30344470786&partnerID=8YFLogxK