Rescue retrieval of AngioGuard embolic capture system after failure of capture sheath retrieval in carotid angioplasty and stenting

Wilson P. Daugherty, J. B. White, H. J. Cloft, D. F. Kallmes

Research output: Contribution to journalArticlepeer-review

11 Scopus citations

Abstract

This report describes the cases of 2 patients with failed retrieval of an AngioGuard distal protection device after carotid stent placement with rescue retrieval by use of a vertebral catheter. An AngioGuard was deployed before carotid angioplasty and stent placement. Attempted removal of the device with the provided recapture sheath was unsuccessful. A 5F 120-cm vertebral catheter was used to recapture and remove the AngioGuard. This technique is a simple and readily available solution for the retrieval of failed removal of a protection device.

Original languageEnglish (US)
Pages (from-to)1594-1595
Number of pages2
JournalAmerican Journal of Neuroradiology
Volume29
Issue number8
DOIs
StatePublished - Sep 2008

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Clinical Neurology

Fingerprint

Dive into the research topics of 'Rescue retrieval of AngioGuard embolic capture system after failure of capture sheath retrieval in carotid angioplasty and stenting'. Together they form a unique fingerprint.

Cite this