Reproducible volar partial lacerations in flexor tendons: A new device for biomechanical studies

Lionel Erhard, Fredrick M. Schultz, Mark E. Zobitz, Chunfeng D Zhao, Peter C Amadio, Kai N. An

Research output: Contribution to journalArticle

7 Citations (Scopus)

Abstract

A device was designed to create clinically relevant, precise partial flexor digitorum profundus tendon lacerations for in vitro studies but can be adapted for in vivo studies. This caliper-based system utilizes a direct measurement of the tendon height and assumes an elliptical cross section to select the depth of the cut. The accuracy was tested on 60 cuts on 12 human tendons lacerated to an expected 50% or 75% of their cross-sectional area, based on the assumption that the cross-section was elliptical in shape. The cuts were made in portions of the tendon that varied in cross-sectional shape and size. The cut surface of the laceration was colored with Methylene blue and then the laceration was completed. The tendon cross-section was digitally imaged and the respective areas of the stained and unstained regions were evaluated using image-processing software. The mean lacerated areas were 52% (SD 5%) and 73% (SD 6%) for targeted lacerations of 50% and 75%, respectively. The device thus appears to be accurate within an acceptable 5% margin of error from the expected area, and adaptable to intra- and inter-tendinous size variations.

Original languageEnglish (US)
Pages (from-to)999-1002
Number of pages4
JournalJournal of Biomechanics
Volume35
Issue number7
DOIs
StatePublished - 2002

Fingerprint

Lacerations
Tendons
Equipment and Supplies
Methylene Blue
Image processing
Software

Keywords

  • Device
  • Flexor tendon
  • Partial laceration

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine

Cite this

Reproducible volar partial lacerations in flexor tendons : A new device for biomechanical studies. / Erhard, Lionel; Schultz, Fredrick M.; Zobitz, Mark E.; Zhao, Chunfeng D; Amadio, Peter C; An, Kai N.

In: Journal of Biomechanics, Vol. 35, No. 7, 2002, p. 999-1002.

Research output: Contribution to journalArticle

Erhard, Lionel ; Schultz, Fredrick M. ; Zobitz, Mark E. ; Zhao, Chunfeng D ; Amadio, Peter C ; An, Kai N. / Reproducible volar partial lacerations in flexor tendons : A new device for biomechanical studies. In: Journal of Biomechanics. 2002 ; Vol. 35, No. 7. pp. 999-1002.
@article{5a2e9c505bfe42a0ba01353f4d6b5125,
title = "Reproducible volar partial lacerations in flexor tendons: A new device for biomechanical studies",
abstract = "A device was designed to create clinically relevant, precise partial flexor digitorum profundus tendon lacerations for in vitro studies but can be adapted for in vivo studies. This caliper-based system utilizes a direct measurement of the tendon height and assumes an elliptical cross section to select the depth of the cut. The accuracy was tested on 60 cuts on 12 human tendons lacerated to an expected 50{\%} or 75{\%} of their cross-sectional area, based on the assumption that the cross-section was elliptical in shape. The cuts were made in portions of the tendon that varied in cross-sectional shape and size. The cut surface of the laceration was colored with Methylene blue and then the laceration was completed. The tendon cross-section was digitally imaged and the respective areas of the stained and unstained regions were evaluated using image-processing software. The mean lacerated areas were 52{\%} (SD 5{\%}) and 73{\%} (SD 6{\%}) for targeted lacerations of 50{\%} and 75{\%}, respectively. The device thus appears to be accurate within an acceptable 5{\%} margin of error from the expected area, and adaptable to intra- and inter-tendinous size variations.",
keywords = "Device, Flexor tendon, Partial laceration",
author = "Lionel Erhard and Schultz, {Fredrick M.} and Zobitz, {Mark E.} and Zhao, {Chunfeng D} and Amadio, {Peter C} and An, {Kai N.}",
year = "2002",
doi = "10.1016/S0021-9290(02)00038-6",
language = "English (US)",
volume = "35",
pages = "999--1002",
journal = "Journal of Biomechanics",
issn = "0021-9290",
publisher = "Elsevier Limited",
number = "7",

}

TY - JOUR

T1 - Reproducible volar partial lacerations in flexor tendons

T2 - A new device for biomechanical studies

AU - Erhard, Lionel

AU - Schultz, Fredrick M.

AU - Zobitz, Mark E.

AU - Zhao, Chunfeng D

AU - Amadio, Peter C

AU - An, Kai N.

PY - 2002

Y1 - 2002

N2 - A device was designed to create clinically relevant, precise partial flexor digitorum profundus tendon lacerations for in vitro studies but can be adapted for in vivo studies. This caliper-based system utilizes a direct measurement of the tendon height and assumes an elliptical cross section to select the depth of the cut. The accuracy was tested on 60 cuts on 12 human tendons lacerated to an expected 50% or 75% of their cross-sectional area, based on the assumption that the cross-section was elliptical in shape. The cuts were made in portions of the tendon that varied in cross-sectional shape and size. The cut surface of the laceration was colored with Methylene blue and then the laceration was completed. The tendon cross-section was digitally imaged and the respective areas of the stained and unstained regions were evaluated using image-processing software. The mean lacerated areas were 52% (SD 5%) and 73% (SD 6%) for targeted lacerations of 50% and 75%, respectively. The device thus appears to be accurate within an acceptable 5% margin of error from the expected area, and adaptable to intra- and inter-tendinous size variations.

AB - A device was designed to create clinically relevant, precise partial flexor digitorum profundus tendon lacerations for in vitro studies but can be adapted for in vivo studies. This caliper-based system utilizes a direct measurement of the tendon height and assumes an elliptical cross section to select the depth of the cut. The accuracy was tested on 60 cuts on 12 human tendons lacerated to an expected 50% or 75% of their cross-sectional area, based on the assumption that the cross-section was elliptical in shape. The cuts were made in portions of the tendon that varied in cross-sectional shape and size. The cut surface of the laceration was colored with Methylene blue and then the laceration was completed. The tendon cross-section was digitally imaged and the respective areas of the stained and unstained regions were evaluated using image-processing software. The mean lacerated areas were 52% (SD 5%) and 73% (SD 6%) for targeted lacerations of 50% and 75%, respectively. The device thus appears to be accurate within an acceptable 5% margin of error from the expected area, and adaptable to intra- and inter-tendinous size variations.

KW - Device

KW - Flexor tendon

KW - Partial laceration

UR - http://www.scopus.com/inward/record.url?scp=0036277442&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0036277442&partnerID=8YFLogxK

U2 - 10.1016/S0021-9290(02)00038-6

DO - 10.1016/S0021-9290(02)00038-6

M3 - Article

C2 - 12052405

AN - SCOPUS:0036277442

VL - 35

SP - 999

EP - 1002

JO - Journal of Biomechanics