Real-time strain rate imaging: Validation of peak compression and expansion rates by a tissue-mimicking phantom

Marek Belohlavek, Virginia B. Bartleson, Mark E. Zobitz

Research output: Contribution to journalArticle

39 Citations (Scopus)

Abstract

Strain rate imaging is a new modality in echocardiography intended for analysis of left ventricular function. This modality extends ultrasonographic techniques for analysis of tissue velocities by providing information about rates of local myocardial compression and expansion. Cyclic cardiac deformation is a complex process. Precision and accuracy of real-time strain rate (rtSR) measurements have not been studied under controlled laboratory conditions. Using a cyclically compressed tissue-mimicking gelatin phantom, we compared rtSR values to corresponding strain rate values calculated off line from local tissue velocities measured by Doppler echocardiography. We tested a clinically relevant range of strain rates (0.5- 3.5 sec-1) and different angles of insonation. Initial tests showed high precision (r ≥ 0.973, P < 0.001), but the assessment of accuracy (bias ≤ 0.559 sec-1) suggested a trend toward systematic underestimation of the reference values. We suspected a confounding influence of a clutter filter and repeated the tests with the filter inactive. The resulting accuracy improved tenfold (bias ≤ 0.045 sec-1), and the systematic underestimation was no longer present. We conclude that the rtSR is precise and accurate for a range of the tested values.

Original languageEnglish (US)
Pages (from-to)565-571
Number of pages7
JournalEchocardiography
Volume18
Issue number7
StatePublished - 2001

Fingerprint

Doppler Echocardiography
Gelatin
Left Ventricular Function
Echocardiography
Reference Values

Keywords

  • Clutter filter
  • Strain rate imaging
  • Tissue Doppler imaging
  • Ultrasound image processing

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Radiology Nuclear Medicine and imaging

Cite this

Real-time strain rate imaging : Validation of peak compression and expansion rates by a tissue-mimicking phantom. / Belohlavek, Marek; Bartleson, Virginia B.; Zobitz, Mark E.

In: Echocardiography, Vol. 18, No. 7, 2001, p. 565-571.

Research output: Contribution to journalArticle

@article{f2f862cb07fa4830a5c5ab0d44b56a72,
title = "Real-time strain rate imaging: Validation of peak compression and expansion rates by a tissue-mimicking phantom",
abstract = "Strain rate imaging is a new modality in echocardiography intended for analysis of left ventricular function. This modality extends ultrasonographic techniques for analysis of tissue velocities by providing information about rates of local myocardial compression and expansion. Cyclic cardiac deformation is a complex process. Precision and accuracy of real-time strain rate (rtSR) measurements have not been studied under controlled laboratory conditions. Using a cyclically compressed tissue-mimicking gelatin phantom, we compared rtSR values to corresponding strain rate values calculated off line from local tissue velocities measured by Doppler echocardiography. We tested a clinically relevant range of strain rates (0.5- 3.5 sec-1) and different angles of insonation. Initial tests showed high precision (r ≥ 0.973, P < 0.001), but the assessment of accuracy (bias ≤ 0.559 sec-1) suggested a trend toward systematic underestimation of the reference values. We suspected a confounding influence of a clutter filter and repeated the tests with the filter inactive. The resulting accuracy improved tenfold (bias ≤ 0.045 sec-1), and the systematic underestimation was no longer present. We conclude that the rtSR is precise and accurate for a range of the tested values.",
keywords = "Clutter filter, Strain rate imaging, Tissue Doppler imaging, Ultrasound image processing",
author = "Marek Belohlavek and Bartleson, {Virginia B.} and Zobitz, {Mark E.}",
year = "2001",
language = "English (US)",
volume = "18",
pages = "565--571",
journal = "Echocardiography",
issn = "0742-2822",
publisher = "Wiley-Blackwell",
number = "7",

}

TY - JOUR

T1 - Real-time strain rate imaging

T2 - Validation of peak compression and expansion rates by a tissue-mimicking phantom

AU - Belohlavek, Marek

AU - Bartleson, Virginia B.

AU - Zobitz, Mark E.

PY - 2001

Y1 - 2001

N2 - Strain rate imaging is a new modality in echocardiography intended for analysis of left ventricular function. This modality extends ultrasonographic techniques for analysis of tissue velocities by providing information about rates of local myocardial compression and expansion. Cyclic cardiac deformation is a complex process. Precision and accuracy of real-time strain rate (rtSR) measurements have not been studied under controlled laboratory conditions. Using a cyclically compressed tissue-mimicking gelatin phantom, we compared rtSR values to corresponding strain rate values calculated off line from local tissue velocities measured by Doppler echocardiography. We tested a clinically relevant range of strain rates (0.5- 3.5 sec-1) and different angles of insonation. Initial tests showed high precision (r ≥ 0.973, P < 0.001), but the assessment of accuracy (bias ≤ 0.559 sec-1) suggested a trend toward systematic underestimation of the reference values. We suspected a confounding influence of a clutter filter and repeated the tests with the filter inactive. The resulting accuracy improved tenfold (bias ≤ 0.045 sec-1), and the systematic underestimation was no longer present. We conclude that the rtSR is precise and accurate for a range of the tested values.

AB - Strain rate imaging is a new modality in echocardiography intended for analysis of left ventricular function. This modality extends ultrasonographic techniques for analysis of tissue velocities by providing information about rates of local myocardial compression and expansion. Cyclic cardiac deformation is a complex process. Precision and accuracy of real-time strain rate (rtSR) measurements have not been studied under controlled laboratory conditions. Using a cyclically compressed tissue-mimicking gelatin phantom, we compared rtSR values to corresponding strain rate values calculated off line from local tissue velocities measured by Doppler echocardiography. We tested a clinically relevant range of strain rates (0.5- 3.5 sec-1) and different angles of insonation. Initial tests showed high precision (r ≥ 0.973, P < 0.001), but the assessment of accuracy (bias ≤ 0.559 sec-1) suggested a trend toward systematic underestimation of the reference values. We suspected a confounding influence of a clutter filter and repeated the tests with the filter inactive. The resulting accuracy improved tenfold (bias ≤ 0.045 sec-1), and the systematic underestimation was no longer present. We conclude that the rtSR is precise and accurate for a range of the tested values.

KW - Clutter filter

KW - Strain rate imaging

KW - Tissue Doppler imaging

KW - Ultrasound image processing

UR - http://www.scopus.com/inward/record.url?scp=0035659793&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0035659793&partnerID=8YFLogxK

M3 - Article

C2 - 11737965

AN - SCOPUS:0035659793

VL - 18

SP - 565

EP - 571

JO - Echocardiography

JF - Echocardiography

SN - 0742-2822

IS - 7

ER -