T1 - Prediction of human observer performance in a 2-alternative forced choice low-contrast detection task using channelized Hotelling observer
T2 - Impact of radiation dose and reconstruction algorithms
AU - Yu, Lifeng
AU - Leng, Shuai
AU - Chen, Lingyun
AU - Kofler, James M.
AU - Carter, Rickey E.
AU - McCollough, Cynthia H.
N1 - Funding Information:
This work was supported in part by National Institutes of Health (NIH) Grant No. R01 EB071095 from the National Institute of Biomedical Imaging and Bioengineering. CHM received research support from Siemens Healthcare. The authors would like to thank Dr. Matthew Kupinski for his help on model observers and Ms. Kristina Nunez for her assistance with paper preparation. Investigators interested in use of the data described in this study should contact the authors.
PY - 2013/4
Y1 - 2013/4
N2 - Purpose: Efficient optimization of CT protocols demands a quantitative approach to predicting human observer performance on specific tasks at various scan and reconstruction settings. The goal of this work was to investigate how well a channelized Hotelling observer (CHO) can predict human observer performance on 2-alternative forced choice (2AFC) lesion-detection tasks at various dose levels and two different reconstruction algorithms: a filtered-backprojection (FBP) and an iterative reconstruction (IR) method. Methods: A 35 × 26 cm2 torso-shaped phantom filled with water was used to simulate an average-sized patient. Three rods with different diameters (small: 3 mm; medium: 5 mm; large: 9 mm) were placed in the center region of the phantom to simulate small, medium, and large lesions. The contrast relative to background was -15 HU at 120 kV. The phantom was scanned 100 times using automatic exposure control each at 60, 120, 240, 360, and 480 quality reference mAs on a 128-slice scanner. After removing the three rods, the water phantom was again scanned 100 times to provide signal-absent background images at the exact same locations. By extracting regions of interest around the three rods and on the signal-absent images, the authors generated 21 2AFC studies. Each 2AFC study had 100 trials, with each trial consisting of a signal-present image and a signal-absent image side-by-side in randomized order. In total, 2100 trials were presented to both the model and human observers. Four medical physicists acted as human observers. For the model observer, the authors used a CHO with Gabor channels, which involves six channel passbands, five orientations, and two phases, leading to a total of 60 channels. The performance predicted by the CHO was compared with that obtained by four medical physicists at each 2AFC study. Results: The human and model observers were highly correlated at each dose level for each lesion size for both FBP and IR. The Pearsons product-moment correlation coefficients were 0.986 95 confidence interval (CI): 0.958-0.996 for FBP and 0.985 (95 CI: 0.863-0.998) for IR. Bland-Altman plots showed excellent agreement for all dose levels and lesions sizes with a mean absolute difference of 1.0 ± 1.1 for FBP and 2.1 ± 3.3 for IR. Conclusions: Human observer performance on a 2AFC lesion detection task in CT with a uniform background can be accurately predicted by a CHO model observer at different radiation dose levels and for both FBP and IR methods.
AB - Purpose: Efficient optimization of CT protocols demands a quantitative approach to predicting human observer performance on specific tasks at various scan and reconstruction settings. The goal of this work was to investigate how well a channelized Hotelling observer (CHO) can predict human observer performance on 2-alternative forced choice (2AFC) lesion-detection tasks at various dose levels and two different reconstruction algorithms: a filtered-backprojection (FBP) and an iterative reconstruction (IR) method. Methods: A 35 × 26 cm2 torso-shaped phantom filled with water was used to simulate an average-sized patient. Three rods with different diameters (small: 3 mm; medium: 5 mm; large: 9 mm) were placed in the center region of the phantom to simulate small, medium, and large lesions. The contrast relative to background was -15 HU at 120 kV. The phantom was scanned 100 times using automatic exposure control each at 60, 120, 240, 360, and 480 quality reference mAs on a 128-slice scanner. After removing the three rods, the water phantom was again scanned 100 times to provide signal-absent background images at the exact same locations. By extracting regions of interest around the three rods and on the signal-absent images, the authors generated 21 2AFC studies. Each 2AFC study had 100 trials, with each trial consisting of a signal-present image and a signal-absent image side-by-side in randomized order. In total, 2100 trials were presented to both the model and human observers. Four medical physicists acted as human observers. For the model observer, the authors used a CHO with Gabor channels, which involves six channel passbands, five orientations, and two phases, leading to a total of 60 channels. The performance predicted by the CHO was compared with that obtained by four medical physicists at each 2AFC study. Results: The human and model observers were highly correlated at each dose level for each lesion size for both FBP and IR. The Pearsons product-moment correlation coefficients were 0.986 95 confidence interval (CI): 0.958-0.996 for FBP and 0.985 (95 CI: 0.863-0.998) for IR. Bland-Altman plots showed excellent agreement for all dose levels and lesions sizes with a mean absolute difference of 1.0 ± 1.1 for FBP and 2.1 ± 3.3 for IR. Conclusions: Human observer performance on a 2AFC lesion detection task in CT with a uniform background can be accurately predicted by a CHO model observer at different radiation dose levels and for both FBP and IR methods.
KW - computed tomography (CT)
KW - image quality
KW - iterative reconstruction (IR)
KW - model observer
KW - radiation dose
UR - http://www.scopus.com/inward/record.url?scp=84876160428&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84876160428&partnerID=8YFLogxK
U2 - 10.1118/1.4794498
DO - 10.1118/1.4794498
M3 - Article
C2 - 23556902
AN - SCOPUS:84876160428
SN - 0094-2405
VL - 40
JO - Medical Physics
JF - Medical Physics
IS - 4
M1 - 041908
ER -