Performance goals for internal quality control of multichannel haematology analysers

G. G. Klee

Research output: Contribution to journalArticlepeer-review

10 Scopus citations

Abstract

Performance goals for analytic bias and analytic imprecision are presented for multichannel haematology analysers. These goals represent a compromize between limits that would cause minimal degradation in the information content for clinical decisions and limits that are achievable by current instruments. For midrange measurements the maximal limits for total analytic bias are ± 14.6% for WBC, ± 5.3% for RBC, ± 4.5% for Hb concentration, ± 4.2% for MCV, and ± 22% for platelet counts. The limits include error inherent in calibrator value assignment, instrument calibration, instrument drift, and instrument non-linearity as well as the quality control systems necessary to monitor each of these processes. If analytic biases are held within the proposed limits, the clinical specificities of the assays will be decreased by less than 7% for all assays except that platelet counts will have a decrease in specificity of about 15%. Analytic imprecision has minimal direct effect on clinical decisions because the influence of within-person and across-person biological variation dominates. However, analytic imprecision is a major factor in the control of analytic bias and thereby remains an important quality control factor.

Original languageEnglish (US)
Pages (from-to)65-74
Number of pages10
JournalClinical and Laboratory Haematology, Supplement
Volume12
Issue number1
StatePublished - 1990

Keywords

  • Accuracy
  • Analytical bias
  • Calibration
  • Calibrator
  • Linearity Precision
  • Quality control
  • Sensitivity
  • Specificity
  • Value

ASJC Scopus subject areas

  • Hematology

Fingerprint

Dive into the research topics of 'Performance goals for internal quality control of multichannel haematology analysers'. Together they form a unique fingerprint.

Cite this