MR angiography fusion technique for treatment planning of intracranial arteriovenous malformations

Kiaran Patrick McGee, Vladimir Ivanovic, Joel P. Felmlee, Fredrick B. Meyer, Bruce E. Pollock, John III Huston

Research output: Contribution to journalArticle

7 Citations (Scopus)

Abstract

Purpose: To develop an image fusion technique using elliptical centric contrast-enhanced (CE) MR angiography (MRA) and three-dimensional (3D) time-of-flight (TOF) acquisitions for radiosurgery treatment planning of arteriovenous malformations (AVMs). Materials and Methods: CE and 3D-TOF MR angiograms with disparate in-plane fields of view (FOVs) were acquired, followed by k-space reformatting to provide equal voxel dimensions. Spatial domain addition was performed to provide a third, fused data volume. Spatial distortion was evaluated on an MRA phantom and provided slice-dependent and global distortion along the three physical dimensions of the MR scanner. In vivo validation was performed on 10 patients with intracranial AVMs prior to their conventional angiogram on the day of gamma knife radiosurgery. Results: Spatial distortion in the phantom within a volume of 14 × 14 × 3.2 cm 3 was less than ±1 mm (±1 standard deviation (SD)) for CE and 3D-TOF data sets. Fused data volumes were successfully generated for all 10 patients. Conclusion: Image fusion can be used to obtain high-resolution CE-MRA images of intracranial AVMs while keeping the fiducial markers needed for gamma knife radiosurgery planning. The spatial fidelity of these data is within the tolerance acceptable for daily quality control (QC) purposes and gamma knife treatment planning.

Original languageEnglish (US)
Pages (from-to)361-369
Number of pages9
JournalJournal of Magnetic Resonance Imaging
Volume23
Issue number3
DOIs
StatePublished - Mar 2006

Fingerprint

Planning Techniques
Intracranial Arteriovenous Malformations
Angiography
Radiosurgery
Fiducial Markers
Therapeutics
Arteriovenous Malformations
Quality Control
Contrast Media

Keywords

  • Arteriovenous malformations
  • Contrast-enhanced MRA
  • Distortion
  • Image fusion
  • Radiosurgery

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

MR angiography fusion technique for treatment planning of intracranial arteriovenous malformations. / McGee, Kiaran Patrick; Ivanovic, Vladimir; Felmlee, Joel P.; Meyer, Fredrick B.; Pollock, Bruce E.; Huston, John III.

In: Journal of Magnetic Resonance Imaging, Vol. 23, No. 3, 03.2006, p. 361-369.

Research output: Contribution to journalArticle

McGee, Kiaran Patrick ; Ivanovic, Vladimir ; Felmlee, Joel P. ; Meyer, Fredrick B. ; Pollock, Bruce E. ; Huston, John III. / MR angiography fusion technique for treatment planning of intracranial arteriovenous malformations. In: Journal of Magnetic Resonance Imaging. 2006 ; Vol. 23, No. 3. pp. 361-369.
@article{bc36e04fa2bc47d8ace3253888f82696,
title = "MR angiography fusion technique for treatment planning of intracranial arteriovenous malformations",
abstract = "Purpose: To develop an image fusion technique using elliptical centric contrast-enhanced (CE) MR angiography (MRA) and three-dimensional (3D) time-of-flight (TOF) acquisitions for radiosurgery treatment planning of arteriovenous malformations (AVMs). Materials and Methods: CE and 3D-TOF MR angiograms with disparate in-plane fields of view (FOVs) were acquired, followed by k-space reformatting to provide equal voxel dimensions. Spatial domain addition was performed to provide a third, fused data volume. Spatial distortion was evaluated on an MRA phantom and provided slice-dependent and global distortion along the three physical dimensions of the MR scanner. In vivo validation was performed on 10 patients with intracranial AVMs prior to their conventional angiogram on the day of gamma knife radiosurgery. Results: Spatial distortion in the phantom within a volume of 14 × 14 × 3.2 cm 3 was less than ±1 mm (±1 standard deviation (SD)) for CE and 3D-TOF data sets. Fused data volumes were successfully generated for all 10 patients. Conclusion: Image fusion can be used to obtain high-resolution CE-MRA images of intracranial AVMs while keeping the fiducial markers needed for gamma knife radiosurgery planning. The spatial fidelity of these data is within the tolerance acceptable for daily quality control (QC) purposes and gamma knife treatment planning.",
keywords = "Arteriovenous malformations, Contrast-enhanced MRA, Distortion, Image fusion, Radiosurgery",
author = "McGee, {Kiaran Patrick} and Vladimir Ivanovic and Felmlee, {Joel P.} and Meyer, {Fredrick B.} and Pollock, {Bruce E.} and Huston, {John III}",
year = "2006",
month = "3",
doi = "10.1002/jmri.20519",
language = "English (US)",
volume = "23",
pages = "361--369",
journal = "Journal of Magnetic Resonance Imaging",
issn = "1053-1807",
publisher = "John Wiley and Sons Inc.",
number = "3",

}

TY - JOUR

T1 - MR angiography fusion technique for treatment planning of intracranial arteriovenous malformations

AU - McGee, Kiaran Patrick

AU - Ivanovic, Vladimir

AU - Felmlee, Joel P.

AU - Meyer, Fredrick B.

AU - Pollock, Bruce E.

AU - Huston, John III

PY - 2006/3

Y1 - 2006/3

N2 - Purpose: To develop an image fusion technique using elliptical centric contrast-enhanced (CE) MR angiography (MRA) and three-dimensional (3D) time-of-flight (TOF) acquisitions for radiosurgery treatment planning of arteriovenous malformations (AVMs). Materials and Methods: CE and 3D-TOF MR angiograms with disparate in-plane fields of view (FOVs) were acquired, followed by k-space reformatting to provide equal voxel dimensions. Spatial domain addition was performed to provide a third, fused data volume. Spatial distortion was evaluated on an MRA phantom and provided slice-dependent and global distortion along the three physical dimensions of the MR scanner. In vivo validation was performed on 10 patients with intracranial AVMs prior to their conventional angiogram on the day of gamma knife radiosurgery. Results: Spatial distortion in the phantom within a volume of 14 × 14 × 3.2 cm 3 was less than ±1 mm (±1 standard deviation (SD)) for CE and 3D-TOF data sets. Fused data volumes were successfully generated for all 10 patients. Conclusion: Image fusion can be used to obtain high-resolution CE-MRA images of intracranial AVMs while keeping the fiducial markers needed for gamma knife radiosurgery planning. The spatial fidelity of these data is within the tolerance acceptable for daily quality control (QC) purposes and gamma knife treatment planning.

AB - Purpose: To develop an image fusion technique using elliptical centric contrast-enhanced (CE) MR angiography (MRA) and three-dimensional (3D) time-of-flight (TOF) acquisitions for radiosurgery treatment planning of arteriovenous malformations (AVMs). Materials and Methods: CE and 3D-TOF MR angiograms with disparate in-plane fields of view (FOVs) were acquired, followed by k-space reformatting to provide equal voxel dimensions. Spatial domain addition was performed to provide a third, fused data volume. Spatial distortion was evaluated on an MRA phantom and provided slice-dependent and global distortion along the three physical dimensions of the MR scanner. In vivo validation was performed on 10 patients with intracranial AVMs prior to their conventional angiogram on the day of gamma knife radiosurgery. Results: Spatial distortion in the phantom within a volume of 14 × 14 × 3.2 cm 3 was less than ±1 mm (±1 standard deviation (SD)) for CE and 3D-TOF data sets. Fused data volumes were successfully generated for all 10 patients. Conclusion: Image fusion can be used to obtain high-resolution CE-MRA images of intracranial AVMs while keeping the fiducial markers needed for gamma knife radiosurgery planning. The spatial fidelity of these data is within the tolerance acceptable for daily quality control (QC) purposes and gamma knife treatment planning.

KW - Arteriovenous malformations

KW - Contrast-enhanced MRA

KW - Distortion

KW - Image fusion

KW - Radiosurgery

UR - http://www.scopus.com/inward/record.url?scp=33644833349&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33644833349&partnerID=8YFLogxK

U2 - 10.1002/jmri.20519

DO - 10.1002/jmri.20519

M3 - Article

C2 - 16475146

AN - SCOPUS:33644833349

VL - 23

SP - 361

EP - 369

JO - Journal of Magnetic Resonance Imaging

JF - Journal of Magnetic Resonance Imaging

SN - 1053-1807

IS - 3

ER -