Mitral prosthesis malfunction. Comparative Doppler echocardiographic studies on mitral prostheses before and after replacement

U. Nellessen, T. Masuyama, C. P. Appleton, T. Tye, R. L. Popp

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

To assess the influence of mitral prosthesis malfunction on various Doppler echocardiographic indexes, we studied the changes in the peak mitral flow velocity during early diastolic filling phase (V(max)), the mean transprosthesis pressure drop from the simplified Bernoulli equation, the mitral valve area by the pressure half-time method, and the left ventricular isovolumic relaxation time in 15 patients before and after replacement of the malfunctioning mitral prosthesis using continuous wave Doppler echocardiography. Examination of the 15 replaced prostheses revealed a torn or perforated leaflet in 12 valves and a sewing ring dehiscence in one valve. Additional restricted leaflet motion (classified as mild obstruction) was seen in three of these 13 valves. In the remaining two valves, severe prosthesis obstruction was noted. Changes in the Doppler indexes between the preoperative and postoperative study were present in all patients regarding V(max) (mean, 2.2 ± 0.3 versus 1.6 ± 0.2 m/sec; p < 0.001), mean gradient (mean, 9 ± 5 versus 5 ± 0.8 mm Hg; p < 0.001), and isovolumic relaxation time (mean, 47 ± 12 msec versus 80 ± 13 msec; p < 0.001. The mean mitral valve area remained virtually unchanged (2.3 ± 0.9 versus 2.6 ± 0.3 cm2; p = NS) but increased postoperatively in each patient with preoperative mild or severe prosthesis obstruction without concomitant aortic regurgitation. Our conclusion is that the peak mitral flow velocity, the mean gradient, and the isovolumic relaxation time are useful parameters in the differentiation of normal and abnormal mitral prosthesis function but may not define the underlying lesion. The determination of the mitral valve area facilitates detection of even mild mitral prosthesis obstruction, but, contrary to theoretical considerations, these data suggest this parameter is not significantly influenced by mitral regurgitation.

Original languageEnglish (US)
Pages (from-to)330-336
Number of pages7
JournalCirculation
Volume79
Issue number2
StatePublished - 1989
Externally publishedYes

Fingerprint

Prostheses and Implants
Mitral Valve
Pressure
Aortic Valve Insufficiency
Doppler Echocardiography
Mitral Valve Insufficiency

ASJC Scopus subject areas

  • Physiology
  • Cardiology and Cardiovascular Medicine

Cite this

Nellessen, U., Masuyama, T., Appleton, C. P., Tye, T., & Popp, R. L. (1989). Mitral prosthesis malfunction. Comparative Doppler echocardiographic studies on mitral prostheses before and after replacement. Circulation, 79(2), 330-336.

Mitral prosthesis malfunction. Comparative Doppler echocardiographic studies on mitral prostheses before and after replacement. / Nellessen, U.; Masuyama, T.; Appleton, C. P.; Tye, T.; Popp, R. L.

In: Circulation, Vol. 79, No. 2, 1989, p. 330-336.

Research output: Contribution to journalArticle

Nellessen, U, Masuyama, T, Appleton, CP, Tye, T & Popp, RL 1989, 'Mitral prosthesis malfunction. Comparative Doppler echocardiographic studies on mitral prostheses before and after replacement', Circulation, vol. 79, no. 2, pp. 330-336.
Nellessen, U. ; Masuyama, T. ; Appleton, C. P. ; Tye, T. ; Popp, R. L. / Mitral prosthesis malfunction. Comparative Doppler echocardiographic studies on mitral prostheses before and after replacement. In: Circulation. 1989 ; Vol. 79, No. 2. pp. 330-336.
@article{10ae338f638b400a92bcb1abc300c921,
title = "Mitral prosthesis malfunction. Comparative Doppler echocardiographic studies on mitral prostheses before and after replacement",
abstract = "To assess the influence of mitral prosthesis malfunction on various Doppler echocardiographic indexes, we studied the changes in the peak mitral flow velocity during early diastolic filling phase (V(max)), the mean transprosthesis pressure drop from the simplified Bernoulli equation, the mitral valve area by the pressure half-time method, and the left ventricular isovolumic relaxation time in 15 patients before and after replacement of the malfunctioning mitral prosthesis using continuous wave Doppler echocardiography. Examination of the 15 replaced prostheses revealed a torn or perforated leaflet in 12 valves and a sewing ring dehiscence in one valve. Additional restricted leaflet motion (classified as mild obstruction) was seen in three of these 13 valves. In the remaining two valves, severe prosthesis obstruction was noted. Changes in the Doppler indexes between the preoperative and postoperative study were present in all patients regarding V(max) (mean, 2.2 ± 0.3 versus 1.6 ± 0.2 m/sec; p < 0.001), mean gradient (mean, 9 ± 5 versus 5 ± 0.8 mm Hg; p < 0.001), and isovolumic relaxation time (mean, 47 ± 12 msec versus 80 ± 13 msec; p < 0.001. The mean mitral valve area remained virtually unchanged (2.3 ± 0.9 versus 2.6 ± 0.3 cm2; p = NS) but increased postoperatively in each patient with preoperative mild or severe prosthesis obstruction without concomitant aortic regurgitation. Our conclusion is that the peak mitral flow velocity, the mean gradient, and the isovolumic relaxation time are useful parameters in the differentiation of normal and abnormal mitral prosthesis function but may not define the underlying lesion. The determination of the mitral valve area facilitates detection of even mild mitral prosthesis obstruction, but, contrary to theoretical considerations, these data suggest this parameter is not significantly influenced by mitral regurgitation.",
author = "U. Nellessen and T. Masuyama and Appleton, {C. P.} and T. Tye and Popp, {R. L.}",
year = "1989",
language = "English (US)",
volume = "79",
pages = "330--336",
journal = "Circulation",
issn = "0009-7322",
publisher = "Lippincott Williams and Wilkins",
number = "2",

}

TY - JOUR

T1 - Mitral prosthesis malfunction. Comparative Doppler echocardiographic studies on mitral prostheses before and after replacement

AU - Nellessen, U.

AU - Masuyama, T.

AU - Appleton, C. P.

AU - Tye, T.

AU - Popp, R. L.

PY - 1989

Y1 - 1989

N2 - To assess the influence of mitral prosthesis malfunction on various Doppler echocardiographic indexes, we studied the changes in the peak mitral flow velocity during early diastolic filling phase (V(max)), the mean transprosthesis pressure drop from the simplified Bernoulli equation, the mitral valve area by the pressure half-time method, and the left ventricular isovolumic relaxation time in 15 patients before and after replacement of the malfunctioning mitral prosthesis using continuous wave Doppler echocardiography. Examination of the 15 replaced prostheses revealed a torn or perforated leaflet in 12 valves and a sewing ring dehiscence in one valve. Additional restricted leaflet motion (classified as mild obstruction) was seen in three of these 13 valves. In the remaining two valves, severe prosthesis obstruction was noted. Changes in the Doppler indexes between the preoperative and postoperative study were present in all patients regarding V(max) (mean, 2.2 ± 0.3 versus 1.6 ± 0.2 m/sec; p < 0.001), mean gradient (mean, 9 ± 5 versus 5 ± 0.8 mm Hg; p < 0.001), and isovolumic relaxation time (mean, 47 ± 12 msec versus 80 ± 13 msec; p < 0.001. The mean mitral valve area remained virtually unchanged (2.3 ± 0.9 versus 2.6 ± 0.3 cm2; p = NS) but increased postoperatively in each patient with preoperative mild or severe prosthesis obstruction without concomitant aortic regurgitation. Our conclusion is that the peak mitral flow velocity, the mean gradient, and the isovolumic relaxation time are useful parameters in the differentiation of normal and abnormal mitral prosthesis function but may not define the underlying lesion. The determination of the mitral valve area facilitates detection of even mild mitral prosthesis obstruction, but, contrary to theoretical considerations, these data suggest this parameter is not significantly influenced by mitral regurgitation.

AB - To assess the influence of mitral prosthesis malfunction on various Doppler echocardiographic indexes, we studied the changes in the peak mitral flow velocity during early diastolic filling phase (V(max)), the mean transprosthesis pressure drop from the simplified Bernoulli equation, the mitral valve area by the pressure half-time method, and the left ventricular isovolumic relaxation time in 15 patients before and after replacement of the malfunctioning mitral prosthesis using continuous wave Doppler echocardiography. Examination of the 15 replaced prostheses revealed a torn or perforated leaflet in 12 valves and a sewing ring dehiscence in one valve. Additional restricted leaflet motion (classified as mild obstruction) was seen in three of these 13 valves. In the remaining two valves, severe prosthesis obstruction was noted. Changes in the Doppler indexes between the preoperative and postoperative study were present in all patients regarding V(max) (mean, 2.2 ± 0.3 versus 1.6 ± 0.2 m/sec; p < 0.001), mean gradient (mean, 9 ± 5 versus 5 ± 0.8 mm Hg; p < 0.001), and isovolumic relaxation time (mean, 47 ± 12 msec versus 80 ± 13 msec; p < 0.001. The mean mitral valve area remained virtually unchanged (2.3 ± 0.9 versus 2.6 ± 0.3 cm2; p = NS) but increased postoperatively in each patient with preoperative mild or severe prosthesis obstruction without concomitant aortic regurgitation. Our conclusion is that the peak mitral flow velocity, the mean gradient, and the isovolumic relaxation time are useful parameters in the differentiation of normal and abnormal mitral prosthesis function but may not define the underlying lesion. The determination of the mitral valve area facilitates detection of even mild mitral prosthesis obstruction, but, contrary to theoretical considerations, these data suggest this parameter is not significantly influenced by mitral regurgitation.

UR - http://www.scopus.com/inward/record.url?scp=0024535054&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0024535054&partnerID=8YFLogxK

M3 - Article

C2 - 2914351

AN - SCOPUS:0024535054

VL - 79

SP - 330

EP - 336

JO - Circulation

JF - Circulation

SN - 0009-7322

IS - 2

ER -