Metal artifact reduction from reformatted projections for hip prostheses in multislice helical computed tomography techniques and initial clinical results

Lifeng Yu, Hua Li, M. S. Jan Mueller, James M. Kofler, Xin Liu, Andrew N. Primak, Joel G. Fletcher, Luis S. Guimaraes, Thanila Macedo, Cynthia H. McColloagh

Research output: Contribution to journalArticlepeer-review

76 Scopus citations

Abstract

Purpose: Hip prosthesis is one of the most common typos of metal implants and can cause significant artifacts in computed tomography (CT) examinations. The puipose of this work was to develop a projection-based method for reducing metal artifacts caused by hip prostheses in multislice helical CT. Method and Materials: The proposed method is based on a novel concept, reformatted projection, which is formed by combining the projection data at the same view angle over the full longitudinal scan range. Detection and segmentation of the metal were performed on each reformatted projection image. Two dimensional interpolation based on Delaunay triangulation was used to fill voids left after removal of the metal in the reformatted projection. The corrected data were then reconstructed using a commercially available algorithm. The main advantage of this method is that both the detection of the metal objects and the interpolations are performed on complete reformatted projections with the entire metal region present, which is particularly useful for long hip prostheses. Twenty clinical abdominal/pelvis exams with hip prostheses were corrected and clinically evaluated. Results: The overall image quality and the conspicuity in some critical organs were significantly improved compared with the uncorrected images: overall quality (P = 0.0024): bladder base (P = 0.0027), and rectum (P = 0.0078). The average noise level in the bladder base was reduced from 86.7 HU to 36.2 HU. In 17 of 20 cases, the radiologists preferred either coronal (13) or axial (4) views of the corrected images. Conclusions: A novel method for reducing metal artifact in multislice helical CT was developed. Initial clinical results showed that the proposed method can effectively reduce the artifacts caused by metal implants for the cases of unilateral and bilateral hip prothesis.

Original languageEnglish (US)
Pages (from-to)691-696
Number of pages6
JournalInvestigative radiology
Volume44
Issue number11
DOIs
StatePublished - Nov 2009

Keywords

  • Computed tomography (ct)
  • Metal artifact reduction
  • Multi-slice helical ct

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Metal artifact reduction from reformatted projections for hip prostheses in multislice helical computed tomography techniques and initial clinical results'. Together they form a unique fingerprint.

Cite this