Improved centric phase encoding orders for three-dimensional magnetization-prepared MR angiography

Alan H. Wilman, Stephen J. Riederer

Research output: Contribution to journalArticlepeer-review

67 Scopus citations

Abstract

New centric phase encoding orders are introduced for transient state three-dimensional MR imaging. Unlike previous centric orders, these variants are not fixed with respect to differences in the relative y and z fields of view, but are tailored to account for the differential degrees of gradient dephasing in the k(y) and k(z) phase encoding directions and thus maintain the highest weighting for the most central k-space values. The new centric orders are compared with previous methods in both simulation and experimental magnetizationprepared MR angiography of the carotid and renal arteries. In comparison with other tested centric orders, the new methods provided better retention of true vessel shape caused by isotropic blurring across the phase encoding plane. In addition, the new methods provided increased contrast, with the degree of improvement over previous methods depending on vessel orientation, size, and degree of inflow.

Original languageEnglish (US)
Pages (from-to)384-392
Number of pages9
JournalMagnetic Resonance in Medicine
Volume36
Issue number3
DOIs
StatePublished - Sep 1996

Keywords

  • MR angiography
  • centric phase encode order
  • magnetization preparation

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Improved centric phase encoding orders for three-dimensional magnetization-prepared MR angiography'. Together they form a unique fingerprint.