Image-based gradient non-linearity characterization to determine higher-order spherical harmonic coefficients for improved spatial position accuracy in magnetic resonance imaging

Paul T. Weavers, Shengzhen Tao, Joshua D Trazasko, Yunhong Shu, Erik J. Tryggestad, Jeffrey L. Gunter, Kiaran Patrick McGee, Daniel V. Litwiller, Ken Pin Hwang, Matthew A Bernstein

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

Purpose Spatial position accuracy in magnetic resonance imaging (MRI) is an important concern for a variety of applications, including radiation therapy planning, surgical planning, and longitudinal studies of morphologic changes to study neurodegenerative diseases. Spatial accuracy is strongly influenced by gradient linearity. This work presents a method for characterizing the gradient non-linearity fields on a per-system basis, and using this information to provide improved and higher-order (9th vs. 5th) spherical harmonic coefficients for better spatial accuracy in MRI. Methods A large fiducial phantom containing 5229 water-filled spheres in a grid pattern is scanned with the MR system, and the positions all the fiducials are measured and compared to the corresponding ground truth fiducial positions as reported from a computed tomography (CT) scan of the object. Systematic errors from off-resonance (i.e., B0) effects are minimized with the use of increased receiver bandwidth (± 125 kHz) and two acquisitions with reversed readout gradient polarity. The spherical harmonic coefficients are estimated using an iterative process, and can be subsequently used to correct for gradient non-linearity. Test-retest stability was assessed with five repeated measurements on a single scanner, and cross-scanner variation on four different, identically-configured 3 T wide-bore systems. Results A decrease in the root-mean-square error (RMSE) over a 50 cm diameter spherical volume from 1.80 mm to 0.77 mm is reported here in the case of replacing the vendor's standard 5th order spherical harmonic coefficients with custom fitted 9th order coefficients, and from 1.5 mm to 1 mm by extending custom fitted 5th order correction to the 9th order. Minimum RMSE varied between scanners, but was stable with repeated measurements in the same scanner. Conclusions The results suggest that the proposed methods may be used on a per-system basis to more accurately calibrate MR gradient non-linearity coefficients when compared to vendor standard corrections.

Original languageEnglish (US)
Pages (from-to)54-62
Number of pages9
JournalMagnetic Resonance Imaging
Volume38
DOIs
StatePublished - May 1 2017

Fingerprint

Magnetic resonance
Mean square error
Magnetic Resonance Imaging
Neurodegenerative diseases
Imaging techniques
Planning
Systematic errors
Radiotherapy
Tomography
Bandwidth
Neurodegenerative Diseases
Longitudinal Studies
Water

Keywords

  • Calibration
  • Gradient non-linearity
  • MRI
  • Treatment planning

ASJC Scopus subject areas

  • Biophysics
  • Biomedical Engineering
  • Radiology Nuclear Medicine and imaging

Cite this

Image-based gradient non-linearity characterization to determine higher-order spherical harmonic coefficients for improved spatial position accuracy in magnetic resonance imaging. / Weavers, Paul T.; Tao, Shengzhen; Trazasko, Joshua D; Shu, Yunhong; Tryggestad, Erik J.; Gunter, Jeffrey L.; McGee, Kiaran Patrick; Litwiller, Daniel V.; Hwang, Ken Pin; Bernstein, Matthew A.

In: Magnetic Resonance Imaging, Vol. 38, 01.05.2017, p. 54-62.

Research output: Contribution to journalArticle

@article{a664f7e44a9e467084c7be3950a64541,
title = "Image-based gradient non-linearity characterization to determine higher-order spherical harmonic coefficients for improved spatial position accuracy in magnetic resonance imaging",
abstract = "Purpose Spatial position accuracy in magnetic resonance imaging (MRI) is an important concern for a variety of applications, including radiation therapy planning, surgical planning, and longitudinal studies of morphologic changes to study neurodegenerative diseases. Spatial accuracy is strongly influenced by gradient linearity. This work presents a method for characterizing the gradient non-linearity fields on a per-system basis, and using this information to provide improved and higher-order (9th vs. 5th) spherical harmonic coefficients for better spatial accuracy in MRI. Methods A large fiducial phantom containing 5229 water-filled spheres in a grid pattern is scanned with the MR system, and the positions all the fiducials are measured and compared to the corresponding ground truth fiducial positions as reported from a computed tomography (CT) scan of the object. Systematic errors from off-resonance (i.e., B0) effects are minimized with the use of increased receiver bandwidth (± 125 kHz) and two acquisitions with reversed readout gradient polarity. The spherical harmonic coefficients are estimated using an iterative process, and can be subsequently used to correct for gradient non-linearity. Test-retest stability was assessed with five repeated measurements on a single scanner, and cross-scanner variation on four different, identically-configured 3 T wide-bore systems. Results A decrease in the root-mean-square error (RMSE) over a 50 cm diameter spherical volume from 1.80 mm to 0.77 mm is reported here in the case of replacing the vendor's standard 5th order spherical harmonic coefficients with custom fitted 9th order coefficients, and from 1.5 mm to 1 mm by extending custom fitted 5th order correction to the 9th order. Minimum RMSE varied between scanners, but was stable with repeated measurements in the same scanner. Conclusions The results suggest that the proposed methods may be used on a per-system basis to more accurately calibrate MR gradient non-linearity coefficients when compared to vendor standard corrections.",
keywords = "Calibration, Gradient non-linearity, MRI, Treatment planning",
author = "Weavers, {Paul T.} and Shengzhen Tao and Trazasko, {Joshua D} and Yunhong Shu and Tryggestad, {Erik J.} and Gunter, {Jeffrey L.} and McGee, {Kiaran Patrick} and Litwiller, {Daniel V.} and Hwang, {Ken Pin} and Bernstein, {Matthew A}",
year = "2017",
month = "5",
day = "1",
doi = "10.1016/j.mri.2016.12.020",
language = "English (US)",
volume = "38",
pages = "54--62",
journal = "Magnetic Resonance Imaging",
issn = "0730-725X",
publisher = "Elsevier Inc.",

}

TY - JOUR

T1 - Image-based gradient non-linearity characterization to determine higher-order spherical harmonic coefficients for improved spatial position accuracy in magnetic resonance imaging

AU - Weavers, Paul T.

AU - Tao, Shengzhen

AU - Trazasko, Joshua D

AU - Shu, Yunhong

AU - Tryggestad, Erik J.

AU - Gunter, Jeffrey L.

AU - McGee, Kiaran Patrick

AU - Litwiller, Daniel V.

AU - Hwang, Ken Pin

AU - Bernstein, Matthew A

PY - 2017/5/1

Y1 - 2017/5/1

N2 - Purpose Spatial position accuracy in magnetic resonance imaging (MRI) is an important concern for a variety of applications, including radiation therapy planning, surgical planning, and longitudinal studies of morphologic changes to study neurodegenerative diseases. Spatial accuracy is strongly influenced by gradient linearity. This work presents a method for characterizing the gradient non-linearity fields on a per-system basis, and using this information to provide improved and higher-order (9th vs. 5th) spherical harmonic coefficients for better spatial accuracy in MRI. Methods A large fiducial phantom containing 5229 water-filled spheres in a grid pattern is scanned with the MR system, and the positions all the fiducials are measured and compared to the corresponding ground truth fiducial positions as reported from a computed tomography (CT) scan of the object. Systematic errors from off-resonance (i.e., B0) effects are minimized with the use of increased receiver bandwidth (± 125 kHz) and two acquisitions with reversed readout gradient polarity. The spherical harmonic coefficients are estimated using an iterative process, and can be subsequently used to correct for gradient non-linearity. Test-retest stability was assessed with five repeated measurements on a single scanner, and cross-scanner variation on four different, identically-configured 3 T wide-bore systems. Results A decrease in the root-mean-square error (RMSE) over a 50 cm diameter spherical volume from 1.80 mm to 0.77 mm is reported here in the case of replacing the vendor's standard 5th order spherical harmonic coefficients with custom fitted 9th order coefficients, and from 1.5 mm to 1 mm by extending custom fitted 5th order correction to the 9th order. Minimum RMSE varied between scanners, but was stable with repeated measurements in the same scanner. Conclusions The results suggest that the proposed methods may be used on a per-system basis to more accurately calibrate MR gradient non-linearity coefficients when compared to vendor standard corrections.

AB - Purpose Spatial position accuracy in magnetic resonance imaging (MRI) is an important concern for a variety of applications, including radiation therapy planning, surgical planning, and longitudinal studies of morphologic changes to study neurodegenerative diseases. Spatial accuracy is strongly influenced by gradient linearity. This work presents a method for characterizing the gradient non-linearity fields on a per-system basis, and using this information to provide improved and higher-order (9th vs. 5th) spherical harmonic coefficients for better spatial accuracy in MRI. Methods A large fiducial phantom containing 5229 water-filled spheres in a grid pattern is scanned with the MR system, and the positions all the fiducials are measured and compared to the corresponding ground truth fiducial positions as reported from a computed tomography (CT) scan of the object. Systematic errors from off-resonance (i.e., B0) effects are minimized with the use of increased receiver bandwidth (± 125 kHz) and two acquisitions with reversed readout gradient polarity. The spherical harmonic coefficients are estimated using an iterative process, and can be subsequently used to correct for gradient non-linearity. Test-retest stability was assessed with five repeated measurements on a single scanner, and cross-scanner variation on four different, identically-configured 3 T wide-bore systems. Results A decrease in the root-mean-square error (RMSE) over a 50 cm diameter spherical volume from 1.80 mm to 0.77 mm is reported here in the case of replacing the vendor's standard 5th order spherical harmonic coefficients with custom fitted 9th order coefficients, and from 1.5 mm to 1 mm by extending custom fitted 5th order correction to the 9th order. Minimum RMSE varied between scanners, but was stable with repeated measurements in the same scanner. Conclusions The results suggest that the proposed methods may be used on a per-system basis to more accurately calibrate MR gradient non-linearity coefficients when compared to vendor standard corrections.

KW - Calibration

KW - Gradient non-linearity

KW - MRI

KW - Treatment planning

UR - http://www.scopus.com/inward/record.url?scp=85007530434&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85007530434&partnerID=8YFLogxK

U2 - 10.1016/j.mri.2016.12.020

DO - 10.1016/j.mri.2016.12.020

M3 - Article

C2 - 28034637

AN - SCOPUS:85007530434

VL - 38

SP - 54

EP - 62

JO - Magnetic Resonance Imaging

JF - Magnetic Resonance Imaging

SN - 0730-725X

ER -