Image Artifact Management for Clinical Magnetic Resonance Imaging on a 7 T Scanner Using Single-Channel Radiofrequency Transmit Mode

Andrew J. Fagan, Kirk M. Welker, Kimberly K. Amrami, Matthew A. Frick, Robert E. Watson, Peter Kollasch, Venkata Chebrolu, Joel P. Felmlee

Research output: Contribution to journalArticle

Abstract

OBJECTIVES: The aim of this work was to devise mitigation strategies for addressing a range of image artifacts on a clinical 7 T magnetic resonance imaging scanner using the regulatory-approved single-channel radiofrequency transmit mode and vendor-supplied radiofrequency coils to facilitate clinical scanning within reasonable scan times. MATERIALS AND METHODS: Optimized imaging sequence protocols were developed for routine musculoskeletal knee and neurological imaging. Sources of severe image nonuniformities were identified, and mitigation strategies were devised. A range of custom-made high permittivity dielectric pads were used to compensate for B1 and B1 inhomogeneities, and also for magnetic susceptibility-induced signal dropouts particularly in the basal regions of the temporal lobes and in the cerebellum. RESULTS: Significant improvements in image uniformity were obtained using dielectric pads in the knee and brain. A combination of small voxels, reduced field of view B0 shimming, and high in-plane parallel imaging factors helped to minimize signal loss in areas of high susceptibility-induced field distortions. The high inherent signal-to-noise ratio at 7 T allowed for high receiver bandwidths and thin slices to minimize chemical shift artifacts. Intermittent artifacts due to radiofrequency inversion pulse limitations (power, bandwidth) were minimized with dielectric pads. A patient with 2 implanted metallic cranial fixation devices located within the radiofrequency transmit field was successfully imaged, with minimal image geometric distortions. CONCLUSIONS: Challenges relating to severe image artifacts at 7 T using single-channel radiofrequency transmit functionality in the knee and brain were overcome using the approaches described in this article. The resultant high diagnostic image quality paves the way for incorporation of this technology into the routine clinical workflow. Further developmental efforts are required to expand the range of applications to other anatomical areas, and to expand the evidence- and knowledge-base relating to the safety of scanning patients with implanted metallic devices.

Original languageEnglish (US)
Pages (from-to)781-791
Number of pages11
JournalInvestigative radiology
Volume54
Issue number12
DOIs
StatePublished - Dec 1 2019

Fingerprint

Artifacts
Magnetic Resonance Imaging
Knee
Equipment and Supplies
Knowledge Bases
Workflow
Brain
Signal-To-Noise Ratio
Temporal Lobe
Patient Safety
Cerebellum
Technology

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Image Artifact Management for Clinical Magnetic Resonance Imaging on a 7 T Scanner Using Single-Channel Radiofrequency Transmit Mode. / Fagan, Andrew J.; Welker, Kirk M.; Amrami, Kimberly K.; Frick, Matthew A.; Watson, Robert E.; Kollasch, Peter; Chebrolu, Venkata; Felmlee, Joel P.

In: Investigative radiology, Vol. 54, No. 12, 01.12.2019, p. 781-791.

Research output: Contribution to journalArticle

Fagan, Andrew J. ; Welker, Kirk M. ; Amrami, Kimberly K. ; Frick, Matthew A. ; Watson, Robert E. ; Kollasch, Peter ; Chebrolu, Venkata ; Felmlee, Joel P. / Image Artifact Management for Clinical Magnetic Resonance Imaging on a 7 T Scanner Using Single-Channel Radiofrequency Transmit Mode. In: Investigative radiology. 2019 ; Vol. 54, No. 12. pp. 781-791.
@article{d54fba5dde9e43dd8819ba893f40934d,
title = "Image Artifact Management for Clinical Magnetic Resonance Imaging on a 7 T Scanner Using Single-Channel Radiofrequency Transmit Mode",
abstract = "OBJECTIVES: The aim of this work was to devise mitigation strategies for addressing a range of image artifacts on a clinical 7 T magnetic resonance imaging scanner using the regulatory-approved single-channel radiofrequency transmit mode and vendor-supplied radiofrequency coils to facilitate clinical scanning within reasonable scan times. MATERIALS AND METHODS: Optimized imaging sequence protocols were developed for routine musculoskeletal knee and neurological imaging. Sources of severe image nonuniformities were identified, and mitigation strategies were devised. A range of custom-made high permittivity dielectric pads were used to compensate for B1 and B1 inhomogeneities, and also for magnetic susceptibility-induced signal dropouts particularly in the basal regions of the temporal lobes and in the cerebellum. RESULTS: Significant improvements in image uniformity were obtained using dielectric pads in the knee and brain. A combination of small voxels, reduced field of view B0 shimming, and high in-plane parallel imaging factors helped to minimize signal loss in areas of high susceptibility-induced field distortions. The high inherent signal-to-noise ratio at 7 T allowed for high receiver bandwidths and thin slices to minimize chemical shift artifacts. Intermittent artifacts due to radiofrequency inversion pulse limitations (power, bandwidth) were minimized with dielectric pads. A patient with 2 implanted metallic cranial fixation devices located within the radiofrequency transmit field was successfully imaged, with minimal image geometric distortions. CONCLUSIONS: Challenges relating to severe image artifacts at 7 T using single-channel radiofrequency transmit functionality in the knee and brain were overcome using the approaches described in this article. The resultant high diagnostic image quality paves the way for incorporation of this technology into the routine clinical workflow. Further developmental efforts are required to expand the range of applications to other anatomical areas, and to expand the evidence- and knowledge-base relating to the safety of scanning patients with implanted metallic devices.",
author = "Fagan, {Andrew J.} and Welker, {Kirk M.} and Amrami, {Kimberly K.} and Frick, {Matthew A.} and Watson, {Robert E.} and Peter Kollasch and Venkata Chebrolu and Felmlee, {Joel P.}",
year = "2019",
month = "12",
day = "1",
doi = "10.1097/RLI.0000000000000598",
language = "English (US)",
volume = "54",
pages = "781--791",
journal = "Investigative Radiology",
issn = "0020-9996",
publisher = "Lippincott Williams and Wilkins",
number = "12",

}

TY - JOUR

T1 - Image Artifact Management for Clinical Magnetic Resonance Imaging on a 7 T Scanner Using Single-Channel Radiofrequency Transmit Mode

AU - Fagan, Andrew J.

AU - Welker, Kirk M.

AU - Amrami, Kimberly K.

AU - Frick, Matthew A.

AU - Watson, Robert E.

AU - Kollasch, Peter

AU - Chebrolu, Venkata

AU - Felmlee, Joel P.

PY - 2019/12/1

Y1 - 2019/12/1

N2 - OBJECTIVES: The aim of this work was to devise mitigation strategies for addressing a range of image artifacts on a clinical 7 T magnetic resonance imaging scanner using the regulatory-approved single-channel radiofrequency transmit mode and vendor-supplied radiofrequency coils to facilitate clinical scanning within reasonable scan times. MATERIALS AND METHODS: Optimized imaging sequence protocols were developed for routine musculoskeletal knee and neurological imaging. Sources of severe image nonuniformities were identified, and mitigation strategies were devised. A range of custom-made high permittivity dielectric pads were used to compensate for B1 and B1 inhomogeneities, and also for magnetic susceptibility-induced signal dropouts particularly in the basal regions of the temporal lobes and in the cerebellum. RESULTS: Significant improvements in image uniformity were obtained using dielectric pads in the knee and brain. A combination of small voxels, reduced field of view B0 shimming, and high in-plane parallel imaging factors helped to minimize signal loss in areas of high susceptibility-induced field distortions. The high inherent signal-to-noise ratio at 7 T allowed for high receiver bandwidths and thin slices to minimize chemical shift artifacts. Intermittent artifacts due to radiofrequency inversion pulse limitations (power, bandwidth) were minimized with dielectric pads. A patient with 2 implanted metallic cranial fixation devices located within the radiofrequency transmit field was successfully imaged, with minimal image geometric distortions. CONCLUSIONS: Challenges relating to severe image artifacts at 7 T using single-channel radiofrequency transmit functionality in the knee and brain were overcome using the approaches described in this article. The resultant high diagnostic image quality paves the way for incorporation of this technology into the routine clinical workflow. Further developmental efforts are required to expand the range of applications to other anatomical areas, and to expand the evidence- and knowledge-base relating to the safety of scanning patients with implanted metallic devices.

AB - OBJECTIVES: The aim of this work was to devise mitigation strategies for addressing a range of image artifacts on a clinical 7 T magnetic resonance imaging scanner using the regulatory-approved single-channel radiofrequency transmit mode and vendor-supplied radiofrequency coils to facilitate clinical scanning within reasonable scan times. MATERIALS AND METHODS: Optimized imaging sequence protocols were developed for routine musculoskeletal knee and neurological imaging. Sources of severe image nonuniformities were identified, and mitigation strategies were devised. A range of custom-made high permittivity dielectric pads were used to compensate for B1 and B1 inhomogeneities, and also for magnetic susceptibility-induced signal dropouts particularly in the basal regions of the temporal lobes and in the cerebellum. RESULTS: Significant improvements in image uniformity were obtained using dielectric pads in the knee and brain. A combination of small voxels, reduced field of view B0 shimming, and high in-plane parallel imaging factors helped to minimize signal loss in areas of high susceptibility-induced field distortions. The high inherent signal-to-noise ratio at 7 T allowed for high receiver bandwidths and thin slices to minimize chemical shift artifacts. Intermittent artifacts due to radiofrequency inversion pulse limitations (power, bandwidth) were minimized with dielectric pads. A patient with 2 implanted metallic cranial fixation devices located within the radiofrequency transmit field was successfully imaged, with minimal image geometric distortions. CONCLUSIONS: Challenges relating to severe image artifacts at 7 T using single-channel radiofrequency transmit functionality in the knee and brain were overcome using the approaches described in this article. The resultant high diagnostic image quality paves the way for incorporation of this technology into the routine clinical workflow. Further developmental efforts are required to expand the range of applications to other anatomical areas, and to expand the evidence- and knowledge-base relating to the safety of scanning patients with implanted metallic devices.

UR - http://www.scopus.com/inward/record.url?scp=85074551510&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85074551510&partnerID=8YFLogxK

U2 - 10.1097/RLI.0000000000000598

DO - 10.1097/RLI.0000000000000598

M3 - Article

C2 - 31503079

AN - SCOPUS:85074551510

VL - 54

SP - 781

EP - 791

JO - Investigative Radiology

JF - Investigative Radiology

SN - 0020-9996

IS - 12

ER -