Head and Neck Radiation Dose and Radiation Safety for Interventional Physicians

Kenneth Fetterly, Beth Ann Schueler, Michael Grams, Glenn Sturchio, Malcolm Bell, Rajiv Gulati

Research output: Contribution to journalArticle

11 Citations (Scopus)

Abstract

Objectives The first aim of this study was to assess the magnitude of radiation dose to tissues of the head and neck of physicians performing x-ray-guided interventional procedures. The second aim was to assess protection of tissues of the head offered by select wearable radiation safety devices. Background Radiation dose to tissues of the head and neck is of significant interest to practicing interventional physicians. However, methods to estimate radiation dose are not generally available, and furthermore, some of the available research relating to protection of these tissues is misleading. Methods Using a single representative geometry, scatter radiation dose to a humanoid phantom was measured using radiochromic film and normalized by the radiation dose to the left collar of the radioprotective thorax apron. Radiation protection offered by leaded glasses and by a radioabsorbent surgical cap was measured. Results In the test geometry, average radiation doses to the unprotected brain, carotid arteries, and ocular lenses were 8.4%, 17%, and 50% of the dose measured at the left collar, respectively. Two representative types of leaded glasses reduced dose to the ocular lens on the side of the physician from which the scatter originates by 27% to 62% but offered no protection to the contralateral eye. The radioabsorbent surgical cap reduced brain dose by only 3.3%. Conclusions A method by which interventional physicians can estimate dose to head and neck tissues on the basis of their personal dosimeter readings is described. Radiation protection of the ocular lenses by leaded glasses may be incomplete, and protection of the brain by a radioabsorbent surgical cap was minimal.

Original languageEnglish (US)
Pages (from-to)520-528
Number of pages9
JournalJACC: Cardiovascular Interventions
Volume10
Issue number5
DOIs
StatePublished - Mar 13 2017

Fingerprint

Neck
Head
Radiation
Safety
Crystalline Lens
Physicians
Glass
Brain
Protective Devices
Background Radiation
Radiation Protection
Carotid Arteries
Reading
Thorax
X-Rays
Research

Keywords

  • head and neck
  • interventional fluoroscopy
  • occupational radiation dose
  • radiation protection

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this

Head and Neck Radiation Dose and Radiation Safety for Interventional Physicians. / Fetterly, Kenneth; Schueler, Beth Ann; Grams, Michael; Sturchio, Glenn; Bell, Malcolm; Gulati, Rajiv.

In: JACC: Cardiovascular Interventions, Vol. 10, No. 5, 13.03.2017, p. 520-528.

Research output: Contribution to journalArticle

Fetterly, Kenneth ; Schueler, Beth Ann ; Grams, Michael ; Sturchio, Glenn ; Bell, Malcolm ; Gulati, Rajiv. / Head and Neck Radiation Dose and Radiation Safety for Interventional Physicians. In: JACC: Cardiovascular Interventions. 2017 ; Vol. 10, No. 5. pp. 520-528.
@article{16c0849be2e049038cc913b6c7e1a1ac,
title = "Head and Neck Radiation Dose and Radiation Safety for Interventional Physicians",
abstract = "Objectives The first aim of this study was to assess the magnitude of radiation dose to tissues of the head and neck of physicians performing x-ray-guided interventional procedures. The second aim was to assess protection of tissues of the head offered by select wearable radiation safety devices. Background Radiation dose to tissues of the head and neck is of significant interest to practicing interventional physicians. However, methods to estimate radiation dose are not generally available, and furthermore, some of the available research relating to protection of these tissues is misleading. Methods Using a single representative geometry, scatter radiation dose to a humanoid phantom was measured using radiochromic film and normalized by the radiation dose to the left collar of the radioprotective thorax apron. Radiation protection offered by leaded glasses and by a radioabsorbent surgical cap was measured. Results In the test geometry, average radiation doses to the unprotected brain, carotid arteries, and ocular lenses were 8.4{\%}, 17{\%}, and 50{\%} of the dose measured at the left collar, respectively. Two representative types of leaded glasses reduced dose to the ocular lens on the side of the physician from which the scatter originates by 27{\%} to 62{\%} but offered no protection to the contralateral eye. The radioabsorbent surgical cap reduced brain dose by only 3.3{\%}. Conclusions A method by which interventional physicians can estimate dose to head and neck tissues on the basis of their personal dosimeter readings is described. Radiation protection of the ocular lenses by leaded glasses may be incomplete, and protection of the brain by a radioabsorbent surgical cap was minimal.",
keywords = "head and neck, interventional fluoroscopy, occupational radiation dose, radiation protection",
author = "Kenneth Fetterly and Schueler, {Beth Ann} and Michael Grams and Glenn Sturchio and Malcolm Bell and Rajiv Gulati",
year = "2017",
month = "3",
day = "13",
doi = "10.1016/j.jcin.2016.11.026",
language = "English (US)",
volume = "10",
pages = "520--528",
journal = "JACC: Cardiovascular Interventions",
issn = "1936-8798",
publisher = "Elsevier Inc.",
number = "5",

}

TY - JOUR

T1 - Head and Neck Radiation Dose and Radiation Safety for Interventional Physicians

AU - Fetterly, Kenneth

AU - Schueler, Beth Ann

AU - Grams, Michael

AU - Sturchio, Glenn

AU - Bell, Malcolm

AU - Gulati, Rajiv

PY - 2017/3/13

Y1 - 2017/3/13

N2 - Objectives The first aim of this study was to assess the magnitude of radiation dose to tissues of the head and neck of physicians performing x-ray-guided interventional procedures. The second aim was to assess protection of tissues of the head offered by select wearable radiation safety devices. Background Radiation dose to tissues of the head and neck is of significant interest to practicing interventional physicians. However, methods to estimate radiation dose are not generally available, and furthermore, some of the available research relating to protection of these tissues is misleading. Methods Using a single representative geometry, scatter radiation dose to a humanoid phantom was measured using radiochromic film and normalized by the radiation dose to the left collar of the radioprotective thorax apron. Radiation protection offered by leaded glasses and by a radioabsorbent surgical cap was measured. Results In the test geometry, average radiation doses to the unprotected brain, carotid arteries, and ocular lenses were 8.4%, 17%, and 50% of the dose measured at the left collar, respectively. Two representative types of leaded glasses reduced dose to the ocular lens on the side of the physician from which the scatter originates by 27% to 62% but offered no protection to the contralateral eye. The radioabsorbent surgical cap reduced brain dose by only 3.3%. Conclusions A method by which interventional physicians can estimate dose to head and neck tissues on the basis of their personal dosimeter readings is described. Radiation protection of the ocular lenses by leaded glasses may be incomplete, and protection of the brain by a radioabsorbent surgical cap was minimal.

AB - Objectives The first aim of this study was to assess the magnitude of radiation dose to tissues of the head and neck of physicians performing x-ray-guided interventional procedures. The second aim was to assess protection of tissues of the head offered by select wearable radiation safety devices. Background Radiation dose to tissues of the head and neck is of significant interest to practicing interventional physicians. However, methods to estimate radiation dose are not generally available, and furthermore, some of the available research relating to protection of these tissues is misleading. Methods Using a single representative geometry, scatter radiation dose to a humanoid phantom was measured using radiochromic film and normalized by the radiation dose to the left collar of the radioprotective thorax apron. Radiation protection offered by leaded glasses and by a radioabsorbent surgical cap was measured. Results In the test geometry, average radiation doses to the unprotected brain, carotid arteries, and ocular lenses were 8.4%, 17%, and 50% of the dose measured at the left collar, respectively. Two representative types of leaded glasses reduced dose to the ocular lens on the side of the physician from which the scatter originates by 27% to 62% but offered no protection to the contralateral eye. The radioabsorbent surgical cap reduced brain dose by only 3.3%. Conclusions A method by which interventional physicians can estimate dose to head and neck tissues on the basis of their personal dosimeter readings is described. Radiation protection of the ocular lenses by leaded glasses may be incomplete, and protection of the brain by a radioabsorbent surgical cap was minimal.

KW - head and neck

KW - interventional fluoroscopy

KW - occupational radiation dose

KW - radiation protection

UR - http://www.scopus.com/inward/record.url?scp=85014537894&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85014537894&partnerID=8YFLogxK

U2 - 10.1016/j.jcin.2016.11.026

DO - 10.1016/j.jcin.2016.11.026

M3 - Article

VL - 10

SP - 520

EP - 528

JO - JACC: Cardiovascular Interventions

JF - JACC: Cardiovascular Interventions

SN - 1936-8798

IS - 5

ER -