N2 - By varying its thickness to compensate for the different path length through the patient as a function of fan angle, a pre-patient bowtie filter modulates flux distribution to reduce patient dose, scatter, and detector dynamic range, and to improve image quality. A dynamic bowtie filter is superior to its traditional, static counterpart in its ability to adjust its thickness along different fan and view angles to suit a specific patient and task. Among the proposed dynamic bowtie designs, the piecewise-linear and the digital beam attenuators offer more flexibility than conventional filters, but rely on analog positioning of a limited number of wedges. In this work, we introduce a new approach with digital control, called the fluid-filled dynamic bowtie filter. It is a two-dimensional array of small binary elements (channels filled or unfilled with attenuating liquid) in which the cumulative thickness along the x-ray path contributes to the bowtie"™s total attenuation. Using simulated data from a pelvic scan, the performance is compared with the piecewise-linear attenuator. The fluid-filled design better matches the desired target attenuation profile and delivers a 4.2x reduction in dynamic range. The variance of the reconstruction (or noise map) can also be more homogeneous. In minimizing peak variance, the fluid-filled attenuator shows a 3% improvement. From the initial simulation results, the proposed design has more control over the flux distribution as a function of both fan and view angles.
AB - By varying its thickness to compensate for the different path length through the patient as a function of fan angle, a pre-patient bowtie filter modulates flux distribution to reduce patient dose, scatter, and detector dynamic range, and to improve image quality. A dynamic bowtie filter is superior to its traditional, static counterpart in its ability to adjust its thickness along different fan and view angles to suit a specific patient and task. Among the proposed dynamic bowtie designs, the piecewise-linear and the digital beam attenuators offer more flexibility than conventional filters, but rely on analog positioning of a limited number of wedges. In this work, we introduce a new approach with digital control, called the fluid-filled dynamic bowtie filter. It is a two-dimensional array of small binary elements (channels filled or unfilled with attenuating liquid) in which the cumulative thickness along the x-ray path contributes to the bowtie"™s total attenuation. Using simulated data from a pelvic scan, the performance is compared with the piecewise-linear attenuator. The fluid-filled design better matches the desired target attenuation profile and delivers a 4.2x reduction in dynamic range. The variance of the reconstruction (or noise map) can also be more homogeneous. In minimizing peak variance, the fluid-filled attenuator shows a 3% improvement. From the initial simulation results, the proposed design has more control over the flux distribution as a function of both fan and view angles.
KW - Dynamic bowtie
KW - Dynamic range reduction
KW - Fluence field modulation
KW - Peak variance minimization
KW - Photon counting detector
UR - http://www.scopus.com/inward/record.url?scp=84943329381&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84943329381&partnerID=8YFLogxK
U2 - 10.1117/12.2081673
DO - 10.1117/12.2081673
M3 - Conference contribution
AN - SCOPUS:84943329381
T3 - Progress in Biomedical Optics and Imaging - Proceedings of SPIE
BT - Medical Imaging 2015
A2 - Hoeschen, Christoph
A2 - Kontos, Despina
A2 - Hoeschen, Christoph
PB - SPIE
Y2 - 22 February 2015 through 25 February 2015
ER -