Flow quantification using fast cine phase-contrast MR imaging, conventional cine phase-contrast MR imaging, and Doppler sonography: In vitro and in vivo validation

Vivian S. Lee, Charles E. Spritzer, Barbara A. Carroll, Luther G. Pool, Matthew A Bernstein, Sheila K. Heinle, James R. MacFall

Research output: Contribution to journalArticle

87 Citations (Scopus)

Abstract

OBJECTIVE. Our purpose was to assess the accuracy of measurements of flow velocity and volume flow rate in an in vitro phantom and in healthy human volunteers using a cardiac-gated, segmented K-space, fast cine phase- contrast (PC) MR imaging technique with view sharing (fast PC). We compared this method with conventional cine PC MR imaging and Doppler sonography. SUBJECTS AND METHODS. Pulsatile flow was generated in a flow phantom that consisted of a cylindric robe having various degrees of tapered stenosis. Phase-encoded velocity maps were obtained using cine PC and fast PC MR imaging. Doppler sonography was also performed. Measurements of aortic and pulmonary artery peak systolic and minimum diastolic velocity and volume flow rate were then compared in eight healthy volunteers using the three imaging techniques. RESULTS. We found excellent agreement between fast PC and cine PC measurements of peak systolic velocity when regions of interest were drawn to exclude vessel margins (r > .99 for phantom studies, and r = .80 for human studies). Correlation between minimum diastolic velocity measurements by MR imaging was limited by noise that resulted from high encoding velocity settings. However, such correlation improved with signal averaging. When compared with predicted values of volume flow rates, both time PC (r > .99) and fast PC (r = .97) MR imaging were more accurate than Doppler sonography (r = .78) in vitro. Measurements of cardiac output were adversely affected by low signal to noise, especially during diastole; estimates based on systolic forward flow resulted in better agreement between the two MR imaging methods. CONCLUSION. Fast PC MR flow quantification may prove to be a useful adjunct to routine MR studies for measurements of peak flow velocity. However, estimates of volume flow rate using fast PC MR imaging are limited because of increased noise during low diastolic flow as well as edge artifacts.

Original languageEnglish (US)
Pages (from-to)1125-1131
Number of pages7
JournalAmerican Journal of Roentgenology
Volume169
Issue number4
StatePublished - Oct 1997
Externally publishedYes

Fingerprint

Doppler Ultrasonography
Noise
Healthy Volunteers
Pulsatile Flow
Diastole
Cardiac Output
Artifacts
Pulmonary Artery
Pathologic Constriction
In Vitro Techniques

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Flow quantification using fast cine phase-contrast MR imaging, conventional cine phase-contrast MR imaging, and Doppler sonography : In vitro and in vivo validation. / Lee, Vivian S.; Spritzer, Charles E.; Carroll, Barbara A.; Pool, Luther G.; Bernstein, Matthew A; Heinle, Sheila K.; MacFall, James R.

In: American Journal of Roentgenology, Vol. 169, No. 4, 10.1997, p. 1125-1131.

Research output: Contribution to journalArticle

Lee, Vivian S. ; Spritzer, Charles E. ; Carroll, Barbara A. ; Pool, Luther G. ; Bernstein, Matthew A ; Heinle, Sheila K. ; MacFall, James R. / Flow quantification using fast cine phase-contrast MR imaging, conventional cine phase-contrast MR imaging, and Doppler sonography : In vitro and in vivo validation. In: American Journal of Roentgenology. 1997 ; Vol. 169, No. 4. pp. 1125-1131.
@article{7cfba25bcfc34ba6aa3085889f2878a1,
title = "Flow quantification using fast cine phase-contrast MR imaging, conventional cine phase-contrast MR imaging, and Doppler sonography: In vitro and in vivo validation",
abstract = "OBJECTIVE. Our purpose was to assess the accuracy of measurements of flow velocity and volume flow rate in an in vitro phantom and in healthy human volunteers using a cardiac-gated, segmented K-space, fast cine phase- contrast (PC) MR imaging technique with view sharing (fast PC). We compared this method with conventional cine PC MR imaging and Doppler sonography. SUBJECTS AND METHODS. Pulsatile flow was generated in a flow phantom that consisted of a cylindric robe having various degrees of tapered stenosis. Phase-encoded velocity maps were obtained using cine PC and fast PC MR imaging. Doppler sonography was also performed. Measurements of aortic and pulmonary artery peak systolic and minimum diastolic velocity and volume flow rate were then compared in eight healthy volunteers using the three imaging techniques. RESULTS. We found excellent agreement between fast PC and cine PC measurements of peak systolic velocity when regions of interest were drawn to exclude vessel margins (r > .99 for phantom studies, and r = .80 for human studies). Correlation between minimum diastolic velocity measurements by MR imaging was limited by noise that resulted from high encoding velocity settings. However, such correlation improved with signal averaging. When compared with predicted values of volume flow rates, both time PC (r > .99) and fast PC (r = .97) MR imaging were more accurate than Doppler sonography (r = .78) in vitro. Measurements of cardiac output were adversely affected by low signal to noise, especially during diastole; estimates based on systolic forward flow resulted in better agreement between the two MR imaging methods. CONCLUSION. Fast PC MR flow quantification may prove to be a useful adjunct to routine MR studies for measurements of peak flow velocity. However, estimates of volume flow rate using fast PC MR imaging are limited because of increased noise during low diastolic flow as well as edge artifacts.",
author = "Lee, {Vivian S.} and Spritzer, {Charles E.} and Carroll, {Barbara A.} and Pool, {Luther G.} and Bernstein, {Matthew A} and Heinle, {Sheila K.} and MacFall, {James R.}",
year = "1997",
month = "10",
language = "English (US)",
volume = "169",
pages = "1125--1131",
journal = "American Journal of Roentgenology",
issn = "0361-803X",
publisher = "American Roentgen Ray Society",
number = "4",

}

TY - JOUR

T1 - Flow quantification using fast cine phase-contrast MR imaging, conventional cine phase-contrast MR imaging, and Doppler sonography

T2 - In vitro and in vivo validation

AU - Lee, Vivian S.

AU - Spritzer, Charles E.

AU - Carroll, Barbara A.

AU - Pool, Luther G.

AU - Bernstein, Matthew A

AU - Heinle, Sheila K.

AU - MacFall, James R.

PY - 1997/10

Y1 - 1997/10

N2 - OBJECTIVE. Our purpose was to assess the accuracy of measurements of flow velocity and volume flow rate in an in vitro phantom and in healthy human volunteers using a cardiac-gated, segmented K-space, fast cine phase- contrast (PC) MR imaging technique with view sharing (fast PC). We compared this method with conventional cine PC MR imaging and Doppler sonography. SUBJECTS AND METHODS. Pulsatile flow was generated in a flow phantom that consisted of a cylindric robe having various degrees of tapered stenosis. Phase-encoded velocity maps were obtained using cine PC and fast PC MR imaging. Doppler sonography was also performed. Measurements of aortic and pulmonary artery peak systolic and minimum diastolic velocity and volume flow rate were then compared in eight healthy volunteers using the three imaging techniques. RESULTS. We found excellent agreement between fast PC and cine PC measurements of peak systolic velocity when regions of interest were drawn to exclude vessel margins (r > .99 for phantom studies, and r = .80 for human studies). Correlation between minimum diastolic velocity measurements by MR imaging was limited by noise that resulted from high encoding velocity settings. However, such correlation improved with signal averaging. When compared with predicted values of volume flow rates, both time PC (r > .99) and fast PC (r = .97) MR imaging were more accurate than Doppler sonography (r = .78) in vitro. Measurements of cardiac output were adversely affected by low signal to noise, especially during diastole; estimates based on systolic forward flow resulted in better agreement between the two MR imaging methods. CONCLUSION. Fast PC MR flow quantification may prove to be a useful adjunct to routine MR studies for measurements of peak flow velocity. However, estimates of volume flow rate using fast PC MR imaging are limited because of increased noise during low diastolic flow as well as edge artifacts.

AB - OBJECTIVE. Our purpose was to assess the accuracy of measurements of flow velocity and volume flow rate in an in vitro phantom and in healthy human volunteers using a cardiac-gated, segmented K-space, fast cine phase- contrast (PC) MR imaging technique with view sharing (fast PC). We compared this method with conventional cine PC MR imaging and Doppler sonography. SUBJECTS AND METHODS. Pulsatile flow was generated in a flow phantom that consisted of a cylindric robe having various degrees of tapered stenosis. Phase-encoded velocity maps were obtained using cine PC and fast PC MR imaging. Doppler sonography was also performed. Measurements of aortic and pulmonary artery peak systolic and minimum diastolic velocity and volume flow rate were then compared in eight healthy volunteers using the three imaging techniques. RESULTS. We found excellent agreement between fast PC and cine PC measurements of peak systolic velocity when regions of interest were drawn to exclude vessel margins (r > .99 for phantom studies, and r = .80 for human studies). Correlation between minimum diastolic velocity measurements by MR imaging was limited by noise that resulted from high encoding velocity settings. However, such correlation improved with signal averaging. When compared with predicted values of volume flow rates, both time PC (r > .99) and fast PC (r = .97) MR imaging were more accurate than Doppler sonography (r = .78) in vitro. Measurements of cardiac output were adversely affected by low signal to noise, especially during diastole; estimates based on systolic forward flow resulted in better agreement between the two MR imaging methods. CONCLUSION. Fast PC MR flow quantification may prove to be a useful adjunct to routine MR studies for measurements of peak flow velocity. However, estimates of volume flow rate using fast PC MR imaging are limited because of increased noise during low diastolic flow as well as edge artifacts.

UR - http://www.scopus.com/inward/record.url?scp=0030922875&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0030922875&partnerID=8YFLogxK

M3 - Article

C2 - 9308476

AN - SCOPUS:0030922875

VL - 169

SP - 1125

EP - 1131

JO - American Journal of Roentgenology

JF - American Journal of Roentgenology

SN - 0361-803X

IS - 4

ER -