Ferromagnetic bare metal stent for endothelial cell capture and retention

Susheil Uthamaraj, Brandon J. Tefft, Ota Hlinomaz, Gurpreet S. Sandhu, Dan Dragomir-Daescu

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

Rapid endothelialization of cardiovascular stents is needed to reduce stent thrombosis and to avoid anti-platelet therapy which can reduce bleeding risk. The feasibility of using magnetic forces to capture and retain endothelial outgrowth cells (EOC) labeled with super paramagnetic iron oxide nanoparticles (SPION) has been shown previously. But this technique requires the development of a mechanically functional stent from a magnetic and biocompatible material followed by in-vitro and in-vivo testing to prove rapid endothelialization. We developed a weakly ferromagnetic stent from 2205 duplex stainless steel using computer aided design (CAD) and its design was further refined using finite element analysis (FEA). The final design of the stent exhibited a principal strain below the fracture limit of the material during mechanical crimping and expansion. One hundred stents were manufactured and a subset of them was used for mechanical testing, retained magnetic field measurements, in-vitro cell capture studies, and in-vivo implantation studies. Ten stents were tested for deployment to verify if they sustained crimping and expansion cycle without failure. Another 10 stents were magnetized using a strong neodymium magnet and their retained magnetic field was measured. The stents showed that the retained magnetism was sufficient to capture SPION-labeled EOC in our in-vitro studies. SPION-labeled EOC capture and retention was verified in large animal models by implanting 1 magnetized stent and 1 non-magnetized control stent in each of 4 pigs. The stented arteries were explanted after 7 days and analyzed histologically. The weakly magnetic stents developed in this study were capable of attracting and retaining SPION-labeled endothelial cells which can promote rapid healing.

Original languageEnglish (US)
Article numbere53100
JournalJournal of Visualized Experiments
Volume2015
Issue number103
DOIs
StatePublished - Sep 18 2015

Keywords

  • 2205 stainless steel
  • Bioengineering
  • CAD
  • Cardiovascular stents
  • Endothelialization
  • FEA
  • Issue 103
  • Magnetic stent
  • Rapid healing

ASJC Scopus subject areas

  • General Neuroscience
  • General Chemical Engineering
  • General Biochemistry, Genetics and Molecular Biology
  • General Immunology and Microbiology

Fingerprint

Dive into the research topics of 'Ferromagnetic bare metal stent for endothelial cell capture and retention'. Together they form a unique fingerprint.

Cite this