Feasibility of dose reduction using novel denoising techniques for low kV (80 kV) CT enterography. Optimization and validation

Luís S. Guimarães, Joel Garland Fletcher, Lifeng Yu, James E. Huprich, Jeff L. Fidler, Armando Manduca, Juan Carlos Ramirez-Giraldo, David R. Holmes III, Cynthia H McCollough

Research output: Contribution to journalArticle

27 Citations (Scopus)

Abstract

Rationale and Objectives: The aim of this study was to optimize and validate projection-space denoising (PSDN) strategies for application to 80-kV computed tomographic (CT) data to achieve 50% dose reduction. Materials and Methods: Image data obtained at 80 kV (mean CT dose index volume, 7.9 mGy) from dual-source, dual-energy CT enterographic (CTE) exams in 42 patients were used. For each exam, nine 80 kV image data sets were reconstructed using PSDN (three levels of intensity) with or without image-based denoising and compared to commercial reconstruction kernels. For optimization, qualitative analysis selected optimal denoising strategies, with quantitative analysis measuring image contrast, noise, and sharpness (full width at half maximum bowel wall thickness, maximum CT number gradient). For validation, two radiologists examined image quality, comparing low-dose 80-kV optimally denoised images to full-dose mixed-voltage images. Results: PSDN algorithms generated the best 80-kV image quality (41 of 42 patients), while the commercial kernels produced the worst (39 of 42) (P < .001). Overall, 80-kV PSDN approaches resulted in higher contrast (mean, 332 vs 290 Hounsfield units), slightly less noise (mean, 20 vs 26 Hounsfield units), but slightly decreased image sharpness (relative bowel wall thickness, 1.069 vs 1.000) compared to full-dose mixed-voltage images. Mean image quality scores for full-dose CTE images were 4.9 compared to 4.5 for optimally denoised half-dose 80-kV CTE images and 3.1 for nondenoised 80-kV CTE images (P < .001). Conclusion: Optimized denoising strategies improve the quality of 80-kV CTE images such that CT data obtained at 50% of routine dose levels approaches the image quality of full-dose exams.

Original languageEnglish (US)
Pages (from-to)1203-1210
Number of pages8
JournalAcademic Radiology
Volume17
Issue number10
DOIs
StatePublished - Oct 2010

Fingerprint

Noise
Radiologists
Datasets

Keywords

  • Bilateral filtering
  • CT enterography, low-energy CT
  • Image denoising
  • Image noise
  • Image quality
  • Noise reduction
  • Projection-space algorithms
  • Radiation dose

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Feasibility of dose reduction using novel denoising techniques for low kV (80 kV) CT enterography. Optimization and validation. / Guimarães, Luís S.; Fletcher, Joel Garland; Yu, Lifeng; Huprich, James E.; Fidler, Jeff L.; Manduca, Armando; Ramirez-Giraldo, Juan Carlos; Holmes III, David R.; McCollough, Cynthia H.

In: Academic Radiology, Vol. 17, No. 10, 10.2010, p. 1203-1210.

Research output: Contribution to journalArticle

@article{bd85b79f2a144b8b9bc08f33d8da3e27,
title = "Feasibility of dose reduction using novel denoising techniques for low kV (80 kV) CT enterography. Optimization and validation",
abstract = "Rationale and Objectives: The aim of this study was to optimize and validate projection-space denoising (PSDN) strategies for application to 80-kV computed tomographic (CT) data to achieve 50{\%} dose reduction. Materials and Methods: Image data obtained at 80 kV (mean CT dose index volume, 7.9 mGy) from dual-source, dual-energy CT enterographic (CTE) exams in 42 patients were used. For each exam, nine 80 kV image data sets were reconstructed using PSDN (three levels of intensity) with or without image-based denoising and compared to commercial reconstruction kernels. For optimization, qualitative analysis selected optimal denoising strategies, with quantitative analysis measuring image contrast, noise, and sharpness (full width at half maximum bowel wall thickness, maximum CT number gradient). For validation, two radiologists examined image quality, comparing low-dose 80-kV optimally denoised images to full-dose mixed-voltage images. Results: PSDN algorithms generated the best 80-kV image quality (41 of 42 patients), while the commercial kernels produced the worst (39 of 42) (P < .001). Overall, 80-kV PSDN approaches resulted in higher contrast (mean, 332 vs 290 Hounsfield units), slightly less noise (mean, 20 vs 26 Hounsfield units), but slightly decreased image sharpness (relative bowel wall thickness, 1.069 vs 1.000) compared to full-dose mixed-voltage images. Mean image quality scores for full-dose CTE images were 4.9 compared to 4.5 for optimally denoised half-dose 80-kV CTE images and 3.1 for nondenoised 80-kV CTE images (P < .001). Conclusion: Optimized denoising strategies improve the quality of 80-kV CTE images such that CT data obtained at 50{\%} of routine dose levels approaches the image quality of full-dose exams.",
keywords = "Bilateral filtering, CT enterography, low-energy CT, Image denoising, Image noise, Image quality, Noise reduction, Projection-space algorithms, Radiation dose",
author = "Guimar{\~a}es, {Lu{\'i}s S.} and Fletcher, {Joel Garland} and Lifeng Yu and Huprich, {James E.} and Fidler, {Jeff L.} and Armando Manduca and Ramirez-Giraldo, {Juan Carlos} and {Holmes III}, {David R.} and McCollough, {Cynthia H}",
year = "2010",
month = "10",
doi = "10.1016/j.acra.2010.07.001",
language = "English (US)",
volume = "17",
pages = "1203--1210",
journal = "Academic Radiology",
issn = "1076-6332",
publisher = "Elsevier USA",
number = "10",

}

TY - JOUR

T1 - Feasibility of dose reduction using novel denoising techniques for low kV (80 kV) CT enterography. Optimization and validation

AU - Guimarães, Luís S.

AU - Fletcher, Joel Garland

AU - Yu, Lifeng

AU - Huprich, James E.

AU - Fidler, Jeff L.

AU - Manduca, Armando

AU - Ramirez-Giraldo, Juan Carlos

AU - Holmes III, David R.

AU - McCollough, Cynthia H

PY - 2010/10

Y1 - 2010/10

N2 - Rationale and Objectives: The aim of this study was to optimize and validate projection-space denoising (PSDN) strategies for application to 80-kV computed tomographic (CT) data to achieve 50% dose reduction. Materials and Methods: Image data obtained at 80 kV (mean CT dose index volume, 7.9 mGy) from dual-source, dual-energy CT enterographic (CTE) exams in 42 patients were used. For each exam, nine 80 kV image data sets were reconstructed using PSDN (three levels of intensity) with or without image-based denoising and compared to commercial reconstruction kernels. For optimization, qualitative analysis selected optimal denoising strategies, with quantitative analysis measuring image contrast, noise, and sharpness (full width at half maximum bowel wall thickness, maximum CT number gradient). For validation, two radiologists examined image quality, comparing low-dose 80-kV optimally denoised images to full-dose mixed-voltage images. Results: PSDN algorithms generated the best 80-kV image quality (41 of 42 patients), while the commercial kernels produced the worst (39 of 42) (P < .001). Overall, 80-kV PSDN approaches resulted in higher contrast (mean, 332 vs 290 Hounsfield units), slightly less noise (mean, 20 vs 26 Hounsfield units), but slightly decreased image sharpness (relative bowel wall thickness, 1.069 vs 1.000) compared to full-dose mixed-voltage images. Mean image quality scores for full-dose CTE images were 4.9 compared to 4.5 for optimally denoised half-dose 80-kV CTE images and 3.1 for nondenoised 80-kV CTE images (P < .001). Conclusion: Optimized denoising strategies improve the quality of 80-kV CTE images such that CT data obtained at 50% of routine dose levels approaches the image quality of full-dose exams.

AB - Rationale and Objectives: The aim of this study was to optimize and validate projection-space denoising (PSDN) strategies for application to 80-kV computed tomographic (CT) data to achieve 50% dose reduction. Materials and Methods: Image data obtained at 80 kV (mean CT dose index volume, 7.9 mGy) from dual-source, dual-energy CT enterographic (CTE) exams in 42 patients were used. For each exam, nine 80 kV image data sets were reconstructed using PSDN (three levels of intensity) with or without image-based denoising and compared to commercial reconstruction kernels. For optimization, qualitative analysis selected optimal denoising strategies, with quantitative analysis measuring image contrast, noise, and sharpness (full width at half maximum bowel wall thickness, maximum CT number gradient). For validation, two radiologists examined image quality, comparing low-dose 80-kV optimally denoised images to full-dose mixed-voltage images. Results: PSDN algorithms generated the best 80-kV image quality (41 of 42 patients), while the commercial kernels produced the worst (39 of 42) (P < .001). Overall, 80-kV PSDN approaches resulted in higher contrast (mean, 332 vs 290 Hounsfield units), slightly less noise (mean, 20 vs 26 Hounsfield units), but slightly decreased image sharpness (relative bowel wall thickness, 1.069 vs 1.000) compared to full-dose mixed-voltage images. Mean image quality scores for full-dose CTE images were 4.9 compared to 4.5 for optimally denoised half-dose 80-kV CTE images and 3.1 for nondenoised 80-kV CTE images (P < .001). Conclusion: Optimized denoising strategies improve the quality of 80-kV CTE images such that CT data obtained at 50% of routine dose levels approaches the image quality of full-dose exams.

KW - Bilateral filtering

KW - CT enterography, low-energy CT

KW - Image denoising

KW - Image noise

KW - Image quality

KW - Noise reduction

KW - Projection-space algorithms

KW - Radiation dose

UR - http://www.scopus.com/inward/record.url?scp=77956464589&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77956464589&partnerID=8YFLogxK

U2 - 10.1016/j.acra.2010.07.001

DO - 10.1016/j.acra.2010.07.001

M3 - Article

C2 - 20832023

AN - SCOPUS:77956464589

VL - 17

SP - 1203

EP - 1210

JO - Academic Radiology

JF - Academic Radiology

SN - 1076-6332

IS - 10

ER -