Evaluating a Convolutional Neural Network Noise Reduction Method When Applied to CT Images Reconstructed Differently Than Training Data

Nathan R. Huber, Andrew D. Missert, Lifeng Yu, Shuai Leng, Cynthia H. McCollough

Research output: Contribution to journalArticlepeer-review

Abstract

Objective The aim of this study was to evaluate a narrowly trained convolutional neural network (CNN) denoising algorithm when applied to images reconstructed differently than training data set. Methods A residual CNN was trained using 10 noise inserted examinations. Training images were reconstructed with 275 mm of field of view (FOV), medium smooth kernel (D30), and 3 mm of thickness. Six examinations were reserved for testing; these were reconstructed with 100 to 450 mm of FOV, smooth to sharp kernels, and 1 to 5 mm of thickness. Results When test and training reconstruction settings were not matched, there was either reduced denoising efficiency or resolution degradation. Denoising efficiency was reduced when FOV was decreased or a smoother kernel was used. Resolution loss occurred when the network was applied to an increased FOV, sharper kernel, or decreased image thickness. Conclusions The CNN denoising performance was degraded with variations in FOV, kernel, or decreased thickness. Denoising performance was not affected by increased thickness.

Original languageEnglish (US)
Pages (from-to)544-551
Number of pages8
JournalJournal of computer assisted tomography
Volume45
Issue number4
DOIs
StatePublished - 2021

Keywords

  • CNN
  • deep learning
  • denoising
  • noise reduction
  • reconstruction

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Evaluating a Convolutional Neural Network Noise Reduction Method When Applied to CT Images Reconstructed Differently Than Training Data'. Together they form a unique fingerprint.

Cite this