Estimating head and neck tissue dose from x-ray scatter to physicians performing x-ray guided cardiovascular procedures: A phantom study

Kenneth A. Fetterly, Beth A. Schueler, Michael P. Grams, Glenn M. Sturchio

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

Physicians performing x-ray guided interventional procedures have a keen interest in radiation safety. Radiation dose to tissues and organs of the head and neck are of particular interest because they are not routinely protected by wearable radiation safety devices. This study was conducted to facilitate estimation of radiation dose to tissues of the head and neck of interventional physicians based on the dose recorded by a personal dosimeter worn on the left collar. Scatter beam qualities maximum energy and HVL were measured for 40 scatter beams emitting from an anthropomorphic patient phantom. Variables of the scatter beams included scatter angle (35° and 90°), primary beam peak tube potential (60, 80, 100, and 120 kVp), and 5 Cu spectral filter thicknesses (0-0.9 mm). Four reference scatter beam qualities were selected to represent the range of scatter beams realized in a typical practice. A general radiographic x-ray tube was tuned to produce scatter-equivalent radiographic beams and used to simultaneously expose the head and neck of an anthropomorphic operator phantom and radiochromic film. The geometric relationship between the x-ray source of the scatter-equivalent beams and the operator phantom was set to mimic that between a patient and physician performing an invasive cardiovascular procedure. Dose to the exterior surface of the operator phantom was measured with both 3 × 3 cm2 pieces of film and personal dosimeters positioned at the location of the left collar. All films were scanned with a calibrated flatbed scanner, which converted the film's reflective density to dose. Films from the transverse planes of the operator phantom provided 2D maps of the dose distribution within the phantom. These dose maps were normalized by the dose at the left collar, providing 2D percent of left collar dose (LCD) maps. The percent LCD maps were overlain with bony anatomy CT images of the operator phantom and estimates of percent LCD to the left, right and whole brain, brain stem, lenses of the eyes, and carotid arteries were calculated. Per expectation, results indicated greater percent dose to superficial versus deep tissues and increasing percent dose to deep tissues with increasing scatter-equivalent beam energy and HVL. The results enable estimation of the scatter dose to tissues of the head and neck of interventional physicians based on occupational dose measured by a personal dosimeter worn at the collar outside the protective apron.

Original languageEnglish (US)
Pages (from-to)43-58
Number of pages16
JournalJournal of Radiological Protection
Volume37
Issue number1
DOIs
StatePublished - Mar 1 2017

Fingerprint

Neck
Head
X-Rays
Physicians
Radiation
Protective Devices
Crystalline Lens
Carotid Arteries
Brain Stem
Anatomy
tissue
dose
brain
Safety
Brain
safety
Radiation Dosimeters
scanner
anatomy
energy

Keywords

  • occupational radiation dose
  • radiochromic film dosimetry
  • scatter radiation

ASJC Scopus subject areas

  • Waste Management and Disposal
  • Public Health, Environmental and Occupational Health

Cite this

Estimating head and neck tissue dose from x-ray scatter to physicians performing x-ray guided cardiovascular procedures : A phantom study. / Fetterly, Kenneth A.; Schueler, Beth A.; Grams, Michael P.; Sturchio, Glenn M.

In: Journal of Radiological Protection, Vol. 37, No. 1, 01.03.2017, p. 43-58.

Research output: Contribution to journalArticle

Fetterly, Kenneth A. ; Schueler, Beth A. ; Grams, Michael P. ; Sturchio, Glenn M. / Estimating head and neck tissue dose from x-ray scatter to physicians performing x-ray guided cardiovascular procedures : A phantom study. In: Journal of Radiological Protection. 2017 ; Vol. 37, No. 1. pp. 43-58.
@article{02c5f25996e44a00a768f8832f68cdab,
title = "Estimating head and neck tissue dose from x-ray scatter to physicians performing x-ray guided cardiovascular procedures: A phantom study",
abstract = "Physicians performing x-ray guided interventional procedures have a keen interest in radiation safety. Radiation dose to tissues and organs of the head and neck are of particular interest because they are not routinely protected by wearable radiation safety devices. This study was conducted to facilitate estimation of radiation dose to tissues of the head and neck of interventional physicians based on the dose recorded by a personal dosimeter worn on the left collar. Scatter beam qualities maximum energy and HVL were measured for 40 scatter beams emitting from an anthropomorphic patient phantom. Variables of the scatter beams included scatter angle (35° and 90°), primary beam peak tube potential (60, 80, 100, and 120 kVp), and 5 Cu spectral filter thicknesses (0-0.9 mm). Four reference scatter beam qualities were selected to represent the range of scatter beams realized in a typical practice. A general radiographic x-ray tube was tuned to produce scatter-equivalent radiographic beams and used to simultaneously expose the head and neck of an anthropomorphic operator phantom and radiochromic film. The geometric relationship between the x-ray source of the scatter-equivalent beams and the operator phantom was set to mimic that between a patient and physician performing an invasive cardiovascular procedure. Dose to the exterior surface of the operator phantom was measured with both 3 × 3 cm2 pieces of film and personal dosimeters positioned at the location of the left collar. All films were scanned with a calibrated flatbed scanner, which converted the film's reflective density to dose. Films from the transverse planes of the operator phantom provided 2D maps of the dose distribution within the phantom. These dose maps were normalized by the dose at the left collar, providing 2D percent of left collar dose (LCD) maps. The percent LCD maps were overlain with bony anatomy CT images of the operator phantom and estimates of percent LCD to the left, right and whole brain, brain stem, lenses of the eyes, and carotid arteries were calculated. Per expectation, results indicated greater percent dose to superficial versus deep tissues and increasing percent dose to deep tissues with increasing scatter-equivalent beam energy and HVL. The results enable estimation of the scatter dose to tissues of the head and neck of interventional physicians based on occupational dose measured by a personal dosimeter worn at the collar outside the protective apron.",
keywords = "occupational radiation dose, radiochromic film dosimetry, scatter radiation",
author = "Fetterly, {Kenneth A.} and Schueler, {Beth A.} and Grams, {Michael P.} and Sturchio, {Glenn M.}",
year = "2017",
month = "3",
day = "1",
doi = "10.1088/1361-6498/37/1/43",
language = "English (US)",
volume = "37",
pages = "43--58",
journal = "Journal of Radiological Protection",
issn = "0952-4746",
publisher = "IOP Publishing Ltd.",
number = "1",

}

TY - JOUR

T1 - Estimating head and neck tissue dose from x-ray scatter to physicians performing x-ray guided cardiovascular procedures

T2 - A phantom study

AU - Fetterly, Kenneth A.

AU - Schueler, Beth A.

AU - Grams, Michael P.

AU - Sturchio, Glenn M.

PY - 2017/3/1

Y1 - 2017/3/1

N2 - Physicians performing x-ray guided interventional procedures have a keen interest in radiation safety. Radiation dose to tissues and organs of the head and neck are of particular interest because they are not routinely protected by wearable radiation safety devices. This study was conducted to facilitate estimation of radiation dose to tissues of the head and neck of interventional physicians based on the dose recorded by a personal dosimeter worn on the left collar. Scatter beam qualities maximum energy and HVL were measured for 40 scatter beams emitting from an anthropomorphic patient phantom. Variables of the scatter beams included scatter angle (35° and 90°), primary beam peak tube potential (60, 80, 100, and 120 kVp), and 5 Cu spectral filter thicknesses (0-0.9 mm). Four reference scatter beam qualities were selected to represent the range of scatter beams realized in a typical practice. A general radiographic x-ray tube was tuned to produce scatter-equivalent radiographic beams and used to simultaneously expose the head and neck of an anthropomorphic operator phantom and radiochromic film. The geometric relationship between the x-ray source of the scatter-equivalent beams and the operator phantom was set to mimic that between a patient and physician performing an invasive cardiovascular procedure. Dose to the exterior surface of the operator phantom was measured with both 3 × 3 cm2 pieces of film and personal dosimeters positioned at the location of the left collar. All films were scanned with a calibrated flatbed scanner, which converted the film's reflective density to dose. Films from the transverse planes of the operator phantom provided 2D maps of the dose distribution within the phantom. These dose maps were normalized by the dose at the left collar, providing 2D percent of left collar dose (LCD) maps. The percent LCD maps were overlain with bony anatomy CT images of the operator phantom and estimates of percent LCD to the left, right and whole brain, brain stem, lenses of the eyes, and carotid arteries were calculated. Per expectation, results indicated greater percent dose to superficial versus deep tissues and increasing percent dose to deep tissues with increasing scatter-equivalent beam energy and HVL. The results enable estimation of the scatter dose to tissues of the head and neck of interventional physicians based on occupational dose measured by a personal dosimeter worn at the collar outside the protective apron.

AB - Physicians performing x-ray guided interventional procedures have a keen interest in radiation safety. Radiation dose to tissues and organs of the head and neck are of particular interest because they are not routinely protected by wearable radiation safety devices. This study was conducted to facilitate estimation of radiation dose to tissues of the head and neck of interventional physicians based on the dose recorded by a personal dosimeter worn on the left collar. Scatter beam qualities maximum energy and HVL were measured for 40 scatter beams emitting from an anthropomorphic patient phantom. Variables of the scatter beams included scatter angle (35° and 90°), primary beam peak tube potential (60, 80, 100, and 120 kVp), and 5 Cu spectral filter thicknesses (0-0.9 mm). Four reference scatter beam qualities were selected to represent the range of scatter beams realized in a typical practice. A general radiographic x-ray tube was tuned to produce scatter-equivalent radiographic beams and used to simultaneously expose the head and neck of an anthropomorphic operator phantom and radiochromic film. The geometric relationship between the x-ray source of the scatter-equivalent beams and the operator phantom was set to mimic that between a patient and physician performing an invasive cardiovascular procedure. Dose to the exterior surface of the operator phantom was measured with both 3 × 3 cm2 pieces of film and personal dosimeters positioned at the location of the left collar. All films were scanned with a calibrated flatbed scanner, which converted the film's reflective density to dose. Films from the transverse planes of the operator phantom provided 2D maps of the dose distribution within the phantom. These dose maps were normalized by the dose at the left collar, providing 2D percent of left collar dose (LCD) maps. The percent LCD maps were overlain with bony anatomy CT images of the operator phantom and estimates of percent LCD to the left, right and whole brain, brain stem, lenses of the eyes, and carotid arteries were calculated. Per expectation, results indicated greater percent dose to superficial versus deep tissues and increasing percent dose to deep tissues with increasing scatter-equivalent beam energy and HVL. The results enable estimation of the scatter dose to tissues of the head and neck of interventional physicians based on occupational dose measured by a personal dosimeter worn at the collar outside the protective apron.

KW - occupational radiation dose

KW - radiochromic film dosimetry

KW - scatter radiation

UR - http://www.scopus.com/inward/record.url?scp=85014575890&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85014575890&partnerID=8YFLogxK

U2 - 10.1088/1361-6498/37/1/43

DO - 10.1088/1361-6498/37/1/43

M3 - Article

C2 - 27941226

AN - SCOPUS:85014575890

VL - 37

SP - 43

EP - 58

JO - Journal of Radiological Protection

JF - Journal of Radiological Protection

SN - 0952-4746

IS - 1

ER -