Effects of patellar thickness on compression and shear forces in total knee arthroplasty

Calvin S. Oishi, Kenton R Kaufman, Steven E. Irby, Clifford W. Colwell

Research output: Contribution to journalArticle

45 Citations (Scopus)

Abstract

Ten unembalmed cadaveric knees were dissected to the capsule preserving the ligamentous and tendinous structures. A posterior cruciate ligament sparing total condylar knee arthroplasty was implanted routinely. A force transducer that measured compression force and shear was implanted into each patella. For each knee, 3 thicknesses of the patellar composite (osteotomized patella, transducer, polyethylene component) were evaluated: (1) precut patellar thickness, (2) precut plus 2 mm, and (3) precut plus 4 mm. The knees were tested in an Oxford Knee Testing Rig, which allowed dynamic testing with 6° of freedom. Values of patellar forces were obtained throughout a range of motion of 0° to 95° flexion. At higher flexion angles (45° and above), the total patellofemoral shear forces for the precut plus 2 mm and the precut plus 4 mm composites were altered significantly from the precut thickness. Increasing the patellar thickness results in a significant increase in shear forces, potentially leading to early loosening of the component or increased wear or both. Therefore, bone conserving resections that increase the patellar composite thickness above the precut thickness should be avoided.

Original languageEnglish (US)
Pages (from-to)283-290
Number of pages8
JournalClinical Orthopaedics and Related Research
Issue number331
StatePublished - 1996
Externally publishedYes

Fingerprint

Knee Replacement Arthroplasties
Knee
Patella
Transducers
Posterior Cruciate Ligament
Polyethylene
Articular Range of Motion
Capsules
Bone and Bones

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine
  • Surgery

Cite this

Effects of patellar thickness on compression and shear forces in total knee arthroplasty. / Oishi, Calvin S.; Kaufman, Kenton R; Irby, Steven E.; Colwell, Clifford W.

In: Clinical Orthopaedics and Related Research, No. 331, 1996, p. 283-290.

Research output: Contribution to journalArticle

@article{a618a277b4eb4f56b248ca7eed7c8121,
title = "Effects of patellar thickness on compression and shear forces in total knee arthroplasty",
abstract = "Ten unembalmed cadaveric knees were dissected to the capsule preserving the ligamentous and tendinous structures. A posterior cruciate ligament sparing total condylar knee arthroplasty was implanted routinely. A force transducer that measured compression force and shear was implanted into each patella. For each knee, 3 thicknesses of the patellar composite (osteotomized patella, transducer, polyethylene component) were evaluated: (1) precut patellar thickness, (2) precut plus 2 mm, and (3) precut plus 4 mm. The knees were tested in an Oxford Knee Testing Rig, which allowed dynamic testing with 6° of freedom. Values of patellar forces were obtained throughout a range of motion of 0° to 95° flexion. At higher flexion angles (45° and above), the total patellofemoral shear forces for the precut plus 2 mm and the precut plus 4 mm composites were altered significantly from the precut thickness. Increasing the patellar thickness results in a significant increase in shear forces, potentially leading to early loosening of the component or increased wear or both. Therefore, bone conserving resections that increase the patellar composite thickness above the precut thickness should be avoided.",
author = "Oishi, {Calvin S.} and Kaufman, {Kenton R} and Irby, {Steven E.} and Colwell, {Clifford W.}",
year = "1996",
language = "English (US)",
pages = "283--290",
journal = "Clinical Orthopaedics and Related Research",
issn = "0009-921X",
publisher = "Springer New York",
number = "331",

}

TY - JOUR

T1 - Effects of patellar thickness on compression and shear forces in total knee arthroplasty

AU - Oishi, Calvin S.

AU - Kaufman, Kenton R

AU - Irby, Steven E.

AU - Colwell, Clifford W.

PY - 1996

Y1 - 1996

N2 - Ten unembalmed cadaveric knees were dissected to the capsule preserving the ligamentous and tendinous structures. A posterior cruciate ligament sparing total condylar knee arthroplasty was implanted routinely. A force transducer that measured compression force and shear was implanted into each patella. For each knee, 3 thicknesses of the patellar composite (osteotomized patella, transducer, polyethylene component) were evaluated: (1) precut patellar thickness, (2) precut plus 2 mm, and (3) precut plus 4 mm. The knees were tested in an Oxford Knee Testing Rig, which allowed dynamic testing with 6° of freedom. Values of patellar forces were obtained throughout a range of motion of 0° to 95° flexion. At higher flexion angles (45° and above), the total patellofemoral shear forces for the precut plus 2 mm and the precut plus 4 mm composites were altered significantly from the precut thickness. Increasing the patellar thickness results in a significant increase in shear forces, potentially leading to early loosening of the component or increased wear or both. Therefore, bone conserving resections that increase the patellar composite thickness above the precut thickness should be avoided.

AB - Ten unembalmed cadaveric knees were dissected to the capsule preserving the ligamentous and tendinous structures. A posterior cruciate ligament sparing total condylar knee arthroplasty was implanted routinely. A force transducer that measured compression force and shear was implanted into each patella. For each knee, 3 thicknesses of the patellar composite (osteotomized patella, transducer, polyethylene component) were evaluated: (1) precut patellar thickness, (2) precut plus 2 mm, and (3) precut plus 4 mm. The knees were tested in an Oxford Knee Testing Rig, which allowed dynamic testing with 6° of freedom. Values of patellar forces were obtained throughout a range of motion of 0° to 95° flexion. At higher flexion angles (45° and above), the total patellofemoral shear forces for the precut plus 2 mm and the precut plus 4 mm composites were altered significantly from the precut thickness. Increasing the patellar thickness results in a significant increase in shear forces, potentially leading to early loosening of the component or increased wear or both. Therefore, bone conserving resections that increase the patellar composite thickness above the precut thickness should be avoided.

UR - http://www.scopus.com/inward/record.url?scp=0029999570&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0029999570&partnerID=8YFLogxK

M3 - Article

C2 - 8895651

AN - SCOPUS:0029999570

SP - 283

EP - 290

JO - Clinical Orthopaedics and Related Research

JF - Clinical Orthopaedics and Related Research

SN - 0009-921X

IS - 331

ER -