Effects of patellar thickness on compression and shear forces in total knee arthroplasty

Calvin S. Oishi, Kenton R. Kaufman, Steven E. Irby, Clifford W. Colwell

Research output: Contribution to journalArticlepeer-review

48 Scopus citations

Abstract

Ten unembalmed cadaveric knees were dissected to the capsule preserving the ligamentous and tendinous structures. A posterior cruciate ligament sparing total condylar knee arthroplasty was implanted routinely. A force transducer that measured compression force and shear was implanted into each patella. For each knee, 3 thicknesses of the patellar composite (osteotomized patella, transducer, polyethylene component) were evaluated: (1) precut patellar thickness, (2) precut plus 2 mm, and (3) precut plus 4 mm. The knees were tested in an Oxford Knee Testing Rig, which allowed dynamic testing with 6° of freedom. Values of patellar forces were obtained throughout a range of motion of 0° to 95° flexion. At higher flexion angles (45° and above), the total patellofemoral shear forces for the precut plus 2 mm and the precut plus 4 mm composites were altered significantly from the precut thickness. Increasing the patellar thickness results in a significant increase in shear forces, potentially leading to early loosening of the component or increased wear or both. Therefore, bone conserving resections that increase the patellar composite thickness above the precut thickness should be avoided.

Original languageEnglish (US)
Pages (from-to)283-290
Number of pages8
JournalClinical orthopaedics and related research
Volume331
DOIs
StatePublished - 1996

ASJC Scopus subject areas

  • Surgery
  • Orthopedics and Sports Medicine

Fingerprint

Dive into the research topics of 'Effects of patellar thickness on compression and shear forces in total knee arthroplasty'. Together they form a unique fingerprint.

Cite this