Abstract

Objectives This study sought to evaluate the Medtronic Evolut PRO Transcatheter Aortic Valve System in patients with severe symptomatic aortic stenosis. Background A next-generation self-expanding transcatheter aortic valve was designed with an external pericardial wrap with the intent to reduce paravalvular leak while maintaining the benefits of a low-profile, self-expanding, and repositionable supra-annular valve. Methods The Medtronic Evolut PRO Clinical Study included 60 patients undergoing transcatheter aortic valve replacement with the Evolut PRO valve at 8 investigational sites in the United States. Clinical outcomes at 30 days were evaluated using Valve Academic Research Consortium-2 criteria. The 2 primary safety endpoints were the incidence of all-cause mortality at 30 days and the incidence of disabling stroke at 30 days. The primary efficacy endpoint was the proportion of patients with no or trace prosthetic valve regurgitation at 30 days. An independent echocardiographic core laboratory (Mayo Clinic, Rochester, Minnesota) was used to adjudicate all echocardiographic assessments. Results All 60 patients received the Evolut PRO valve. At 30 days, 1 patient (1.7%) died and 1 patient (1.7%) experienced a nonfatal disabling stroke. Paravalvular regurgitation at 30 days was absent or trace in 72.4% of patients and was mild in the remainder of patients, with no patients having worse than mild paravavlular leak. The mean atrioventricular gradient was 6.4 ± 2.1 mm Hg and effective orifice area was 2.0 ± 0.5 cm2 at 30 days. Conclusions The safety and efficacy results of this study support the use of the Evolut PRO System for the treatment of severe symptomatic aortic stenosis in patients who are at increased surgical risk, resulting in excellent hemodynamics and minimal paravalvular leak (The Medtronic TAVR 2.0 US Clinical Study; NCT02738853)

Original languageEnglish (US)
Pages (from-to)181-191
Number of pages11
JournalJACC: Cardiovascular Interventions
Volume11
Issue number2
DOIs
StatePublished - Jan 22 2018

Fingerprint

Aortic Valve
Aortic Valve Stenosis
Stroke
Safety
Incidence
Hemodynamics
Mortality

Keywords

  • aortic stenosis
  • paravalvular leak
  • transcatheter aortic valve replacement

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this

Forrest, J. K., Mangi, A. A., Popma, J. J., Khabbaz, K., Reardon, M. J., Kleiman, N. S., ... Williams, M. R. (2018). Early Outcomes With the Evolut PRO Repositionable Self-Expanding Transcatheter Aortic Valve With Pericardial Wrap. JACC: Cardiovascular Interventions, 11(2), 181-191. https://doi.org/10.1016/j.jcin.2017.10.014

Early Outcomes With the Evolut PRO Repositionable Self-Expanding Transcatheter Aortic Valve With Pericardial Wrap. / Forrest, John K.; Mangi, Abeel A.; Popma, Jeffrey J.; Khabbaz, Kamal; Reardon, Michael J.; Kleiman, Neal S.; Yakubov, Steven J.; Watson, Daniel; Kodali, Susheel; George, Isaac; Tadros, Peter; Zorn, George L.; Brown, John; Kipperman, Robert; Saul, Sara; Qiao, Hongyan; Oh, Jae Kuen; Williams, Mathew R.

In: JACC: Cardiovascular Interventions, Vol. 11, No. 2, 22.01.2018, p. 181-191.

Research output: Contribution to journalArticle

Forrest, JK, Mangi, AA, Popma, JJ, Khabbaz, K, Reardon, MJ, Kleiman, NS, Yakubov, SJ, Watson, D, Kodali, S, George, I, Tadros, P, Zorn, GL, Brown, J, Kipperman, R, Saul, S, Qiao, H, Oh, JK & Williams, MR 2018, 'Early Outcomes With the Evolut PRO Repositionable Self-Expanding Transcatheter Aortic Valve With Pericardial Wrap', JACC: Cardiovascular Interventions, vol. 11, no. 2, pp. 181-191. https://doi.org/10.1016/j.jcin.2017.10.014
Forrest, John K. ; Mangi, Abeel A. ; Popma, Jeffrey J. ; Khabbaz, Kamal ; Reardon, Michael J. ; Kleiman, Neal S. ; Yakubov, Steven J. ; Watson, Daniel ; Kodali, Susheel ; George, Isaac ; Tadros, Peter ; Zorn, George L. ; Brown, John ; Kipperman, Robert ; Saul, Sara ; Qiao, Hongyan ; Oh, Jae Kuen ; Williams, Mathew R. / Early Outcomes With the Evolut PRO Repositionable Self-Expanding Transcatheter Aortic Valve With Pericardial Wrap. In: JACC: Cardiovascular Interventions. 2018 ; Vol. 11, No. 2. pp. 181-191.
@article{0cf734cda4c945738c5b8c96fec82f84,
title = "Early Outcomes With the Evolut PRO Repositionable Self-Expanding Transcatheter Aortic Valve With Pericardial Wrap",
abstract = "Objectives This study sought to evaluate the Medtronic Evolut PRO Transcatheter Aortic Valve System in patients with severe symptomatic aortic stenosis. Background A next-generation self-expanding transcatheter aortic valve was designed with an external pericardial wrap with the intent to reduce paravalvular leak while maintaining the benefits of a low-profile, self-expanding, and repositionable supra-annular valve. Methods The Medtronic Evolut PRO Clinical Study included 60 patients undergoing transcatheter aortic valve replacement with the Evolut PRO valve at 8 investigational sites in the United States. Clinical outcomes at 30 days were evaluated using Valve Academic Research Consortium-2 criteria. The 2 primary safety endpoints were the incidence of all-cause mortality at 30 days and the incidence of disabling stroke at 30 days. The primary efficacy endpoint was the proportion of patients with no or trace prosthetic valve regurgitation at 30 days. An independent echocardiographic core laboratory (Mayo Clinic, Rochester, Minnesota) was used to adjudicate all echocardiographic assessments. Results All 60 patients received the Evolut PRO valve. At 30 days, 1 patient (1.7{\%}) died and 1 patient (1.7{\%}) experienced a nonfatal disabling stroke. Paravalvular regurgitation at 30 days was absent or trace in 72.4{\%} of patients and was mild in the remainder of patients, with no patients having worse than mild paravavlular leak. The mean atrioventricular gradient was 6.4 ± 2.1 mm Hg and effective orifice area was 2.0 ± 0.5 cm2 at 30 days. Conclusions The safety and efficacy results of this study support the use of the Evolut PRO System for the treatment of severe symptomatic aortic stenosis in patients who are at increased surgical risk, resulting in excellent hemodynamics and minimal paravalvular leak (The Medtronic TAVR 2.0 US Clinical Study; NCT02738853)",
keywords = "aortic stenosis, paravalvular leak, transcatheter aortic valve replacement",
author = "Forrest, {John K.} and Mangi, {Abeel A.} and Popma, {Jeffrey J.} and Kamal Khabbaz and Reardon, {Michael J.} and Kleiman, {Neal S.} and Yakubov, {Steven J.} and Daniel Watson and Susheel Kodali and Isaac George and Peter Tadros and Zorn, {George L.} and John Brown and Robert Kipperman and Sara Saul and Hongyan Qiao and Oh, {Jae Kuen} and Williams, {Mathew R.}",
year = "2018",
month = "1",
day = "22",
doi = "10.1016/j.jcin.2017.10.014",
language = "English (US)",
volume = "11",
pages = "181--191",
journal = "JACC: Cardiovascular Interventions",
issn = "1936-8798",
publisher = "Elsevier Inc.",
number = "2",

}

TY - JOUR

T1 - Early Outcomes With the Evolut PRO Repositionable Self-Expanding Transcatheter Aortic Valve With Pericardial Wrap

AU - Forrest, John K.

AU - Mangi, Abeel A.

AU - Popma, Jeffrey J.

AU - Khabbaz, Kamal

AU - Reardon, Michael J.

AU - Kleiman, Neal S.

AU - Yakubov, Steven J.

AU - Watson, Daniel

AU - Kodali, Susheel

AU - George, Isaac

AU - Tadros, Peter

AU - Zorn, George L.

AU - Brown, John

AU - Kipperman, Robert

AU - Saul, Sara

AU - Qiao, Hongyan

AU - Oh, Jae Kuen

AU - Williams, Mathew R.

PY - 2018/1/22

Y1 - 2018/1/22

N2 - Objectives This study sought to evaluate the Medtronic Evolut PRO Transcatheter Aortic Valve System in patients with severe symptomatic aortic stenosis. Background A next-generation self-expanding transcatheter aortic valve was designed with an external pericardial wrap with the intent to reduce paravalvular leak while maintaining the benefits of a low-profile, self-expanding, and repositionable supra-annular valve. Methods The Medtronic Evolut PRO Clinical Study included 60 patients undergoing transcatheter aortic valve replacement with the Evolut PRO valve at 8 investigational sites in the United States. Clinical outcomes at 30 days were evaluated using Valve Academic Research Consortium-2 criteria. The 2 primary safety endpoints were the incidence of all-cause mortality at 30 days and the incidence of disabling stroke at 30 days. The primary efficacy endpoint was the proportion of patients with no or trace prosthetic valve regurgitation at 30 days. An independent echocardiographic core laboratory (Mayo Clinic, Rochester, Minnesota) was used to adjudicate all echocardiographic assessments. Results All 60 patients received the Evolut PRO valve. At 30 days, 1 patient (1.7%) died and 1 patient (1.7%) experienced a nonfatal disabling stroke. Paravalvular regurgitation at 30 days was absent or trace in 72.4% of patients and was mild in the remainder of patients, with no patients having worse than mild paravavlular leak. The mean atrioventricular gradient was 6.4 ± 2.1 mm Hg and effective orifice area was 2.0 ± 0.5 cm2 at 30 days. Conclusions The safety and efficacy results of this study support the use of the Evolut PRO System for the treatment of severe symptomatic aortic stenosis in patients who are at increased surgical risk, resulting in excellent hemodynamics and minimal paravalvular leak (The Medtronic TAVR 2.0 US Clinical Study; NCT02738853)

AB - Objectives This study sought to evaluate the Medtronic Evolut PRO Transcatheter Aortic Valve System in patients with severe symptomatic aortic stenosis. Background A next-generation self-expanding transcatheter aortic valve was designed with an external pericardial wrap with the intent to reduce paravalvular leak while maintaining the benefits of a low-profile, self-expanding, and repositionable supra-annular valve. Methods The Medtronic Evolut PRO Clinical Study included 60 patients undergoing transcatheter aortic valve replacement with the Evolut PRO valve at 8 investigational sites in the United States. Clinical outcomes at 30 days were evaluated using Valve Academic Research Consortium-2 criteria. The 2 primary safety endpoints were the incidence of all-cause mortality at 30 days and the incidence of disabling stroke at 30 days. The primary efficacy endpoint was the proportion of patients with no or trace prosthetic valve regurgitation at 30 days. An independent echocardiographic core laboratory (Mayo Clinic, Rochester, Minnesota) was used to adjudicate all echocardiographic assessments. Results All 60 patients received the Evolut PRO valve. At 30 days, 1 patient (1.7%) died and 1 patient (1.7%) experienced a nonfatal disabling stroke. Paravalvular regurgitation at 30 days was absent or trace in 72.4% of patients and was mild in the remainder of patients, with no patients having worse than mild paravavlular leak. The mean atrioventricular gradient was 6.4 ± 2.1 mm Hg and effective orifice area was 2.0 ± 0.5 cm2 at 30 days. Conclusions The safety and efficacy results of this study support the use of the Evolut PRO System for the treatment of severe symptomatic aortic stenosis in patients who are at increased surgical risk, resulting in excellent hemodynamics and minimal paravalvular leak (The Medtronic TAVR 2.0 US Clinical Study; NCT02738853)

KW - aortic stenosis

KW - paravalvular leak

KW - transcatheter aortic valve replacement

UR - http://www.scopus.com/inward/record.url?scp=85040543556&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85040543556&partnerID=8YFLogxK

U2 - 10.1016/j.jcin.2017.10.014

DO - 10.1016/j.jcin.2017.10.014

M3 - Article

C2 - 29348010

AN - SCOPUS:85040543556

VL - 11

SP - 181

EP - 191

JO - JACC: Cardiovascular Interventions

JF - JACC: Cardiovascular Interventions

SN - 1936-8798

IS - 2

ER -