CT dosimetry: Comparison of measurement techniques and devices

John A. Bauhs, Thomas J. Vrieze, Andrew N. Primak, Michael R. Bruesewitz, Cynthia H McCollough

Research output: Contribution to journalArticle

127 Citations (Scopus)

Abstract

In x-ray computed tomography (CT), the most common parameter used to estimate and minimize patient dose is the CT dose index (CTDI). The CTDI is a volume-averaged measure that is used in situations where the table is incremented in conjunction with the tube rotation. Variants of the CTDI correct for averaging across the field of view and for adjacent beam overlaps or gaps. CTDI is usually measured with a pencil-shaped ionization chamber, although methods have been developed that use alternative detectors, including an optically stimulated luminescence probe and a solid-state real-time dosimeter. Because the CTDI represents an averaged dose to a homogeneous cylindrical phantom, the measurements are only an approximation of the patient dose. Furthermore, dose from interventional or perfusion CT, in which the table remains stationary between multiple scans, is best evaluated with point dose measurements made with small detectors. CTDI and point dose values are nearly the same for measurement of surface dose from spiral CT. However, for measurement of surface dose from perfusion CT, the dose is overestimated by a factor of two or more with CTDI values in comparison with point dose values. Both CTDI and point dose measurement are valuable for evaluating CT scanner output and estimating patient dose.

Original languageEnglish (US)
Pages (from-to)245-253
Number of pages9
JournalRadiographics
Volume28
Issue number1
DOIs
StatePublished - Jan 2008

Fingerprint

Tomography
Equipment and Supplies
Perfusion
X-Ray Computed Tomography Scanners
Spiral Computed Tomography
Luminescence
X-Rays

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Bauhs, J. A., Vrieze, T. J., Primak, A. N., Bruesewitz, M. R., & McCollough, C. H. (2008). CT dosimetry: Comparison of measurement techniques and devices. Radiographics, 28(1), 245-253. https://doi.org/10.1148/rg.281075024

CT dosimetry : Comparison of measurement techniques and devices. / Bauhs, John A.; Vrieze, Thomas J.; Primak, Andrew N.; Bruesewitz, Michael R.; McCollough, Cynthia H.

In: Radiographics, Vol. 28, No. 1, 01.2008, p. 245-253.

Research output: Contribution to journalArticle

Bauhs, JA, Vrieze, TJ, Primak, AN, Bruesewitz, MR & McCollough, CH 2008, 'CT dosimetry: Comparison of measurement techniques and devices', Radiographics, vol. 28, no. 1, pp. 245-253. https://doi.org/10.1148/rg.281075024
Bauhs, John A. ; Vrieze, Thomas J. ; Primak, Andrew N. ; Bruesewitz, Michael R. ; McCollough, Cynthia H. / CT dosimetry : Comparison of measurement techniques and devices. In: Radiographics. 2008 ; Vol. 28, No. 1. pp. 245-253.
@article{7c09e489011049bb98e883c09d54a81d,
title = "CT dosimetry: Comparison of measurement techniques and devices",
abstract = "In x-ray computed tomography (CT), the most common parameter used to estimate and minimize patient dose is the CT dose index (CTDI). The CTDI is a volume-averaged measure that is used in situations where the table is incremented in conjunction with the tube rotation. Variants of the CTDI correct for averaging across the field of view and for adjacent beam overlaps or gaps. CTDI is usually measured with a pencil-shaped ionization chamber, although methods have been developed that use alternative detectors, including an optically stimulated luminescence probe and a solid-state real-time dosimeter. Because the CTDI represents an averaged dose to a homogeneous cylindrical phantom, the measurements are only an approximation of the patient dose. Furthermore, dose from interventional or perfusion CT, in which the table remains stationary between multiple scans, is best evaluated with point dose measurements made with small detectors. CTDI and point dose values are nearly the same for measurement of surface dose from spiral CT. However, for measurement of surface dose from perfusion CT, the dose is overestimated by a factor of two or more with CTDI values in comparison with point dose values. Both CTDI and point dose measurement are valuable for evaluating CT scanner output and estimating patient dose.",
author = "Bauhs, {John A.} and Vrieze, {Thomas J.} and Primak, {Andrew N.} and Bruesewitz, {Michael R.} and McCollough, {Cynthia H}",
year = "2008",
month = "1",
doi = "10.1148/rg.281075024",
language = "English (US)",
volume = "28",
pages = "245--253",
journal = "Radiographics",
issn = "0271-5333",
publisher = "Radiological Society of North America Inc.",
number = "1",

}

TY - JOUR

T1 - CT dosimetry

T2 - Comparison of measurement techniques and devices

AU - Bauhs, John A.

AU - Vrieze, Thomas J.

AU - Primak, Andrew N.

AU - Bruesewitz, Michael R.

AU - McCollough, Cynthia H

PY - 2008/1

Y1 - 2008/1

N2 - In x-ray computed tomography (CT), the most common parameter used to estimate and minimize patient dose is the CT dose index (CTDI). The CTDI is a volume-averaged measure that is used in situations where the table is incremented in conjunction with the tube rotation. Variants of the CTDI correct for averaging across the field of view and for adjacent beam overlaps or gaps. CTDI is usually measured with a pencil-shaped ionization chamber, although methods have been developed that use alternative detectors, including an optically stimulated luminescence probe and a solid-state real-time dosimeter. Because the CTDI represents an averaged dose to a homogeneous cylindrical phantom, the measurements are only an approximation of the patient dose. Furthermore, dose from interventional or perfusion CT, in which the table remains stationary between multiple scans, is best evaluated with point dose measurements made with small detectors. CTDI and point dose values are nearly the same for measurement of surface dose from spiral CT. However, for measurement of surface dose from perfusion CT, the dose is overestimated by a factor of two or more with CTDI values in comparison with point dose values. Both CTDI and point dose measurement are valuable for evaluating CT scanner output and estimating patient dose.

AB - In x-ray computed tomography (CT), the most common parameter used to estimate and minimize patient dose is the CT dose index (CTDI). The CTDI is a volume-averaged measure that is used in situations where the table is incremented in conjunction with the tube rotation. Variants of the CTDI correct for averaging across the field of view and for adjacent beam overlaps or gaps. CTDI is usually measured with a pencil-shaped ionization chamber, although methods have been developed that use alternative detectors, including an optically stimulated luminescence probe and a solid-state real-time dosimeter. Because the CTDI represents an averaged dose to a homogeneous cylindrical phantom, the measurements are only an approximation of the patient dose. Furthermore, dose from interventional or perfusion CT, in which the table remains stationary between multiple scans, is best evaluated with point dose measurements made with small detectors. CTDI and point dose values are nearly the same for measurement of surface dose from spiral CT. However, for measurement of surface dose from perfusion CT, the dose is overestimated by a factor of two or more with CTDI values in comparison with point dose values. Both CTDI and point dose measurement are valuable for evaluating CT scanner output and estimating patient dose.

UR - http://www.scopus.com/inward/record.url?scp=38349099222&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=38349099222&partnerID=8YFLogxK

U2 - 10.1148/rg.281075024

DO - 10.1148/rg.281075024

M3 - Article

C2 - 18203941

AN - SCOPUS:38349099222

VL - 28

SP - 245

EP - 253

JO - Radiographics

JF - Radiographics

SN - 0271-5333

IS - 1

ER -