A method is presented in which an extended longitudinal field of view (FOV), as required for whole-body MRI or MRA peripheral runoff studies, is acquired in one seamless image. Previous methods typically either acquired 3D data at multiple static "stations" which covered the extended FOV or as a series of 2D axial sections. The method presented here maintains the benefits of 3D acquisition while removing the discrete nature of the multistation method by continuous acquisition of MR data as the patient table moves through the desired FOV. Although the technique acquires data only from a homogeneous central volume of the magnet at any point in time, by spatially registering all data it is possible to extend the FOV well beyond this volume. The method is demonstrated experimentally with phantoms, in vivo angiographic animal studies, and in vivo human studies.

Original languageEnglish (US)
Pages (from-to)224-231
Number of pages8
JournalMagnetic Resonance in Medicine
Volume47
Issue number2
DOIs
StatePublished - 2002

Fingerprint

Magnets

Keywords

  • Extended FOV
  • Moving table MRI
  • Peripheral MR angiography
  • Whole body MRI

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Continuously moving table data acquisition method for long FOV contrast-enhanced MRA and whole-body MRI. / Kruger, David G.; Riederer, Stephen J; Grimm, Roger C.; Rossman, Phillip J.; Minnesota.

In: Magnetic Resonance in Medicine, Vol. 47, No. 2, 2002, p. 224-231.

Research output: Contribution to journalArticle

Kruger, David G. ; Riederer, Stephen J ; Grimm, Roger C. ; Rossman, Phillip J. ; Minnesota. / Continuously moving table data acquisition method for long FOV contrast-enhanced MRA and whole-body MRI. In: Magnetic Resonance in Medicine. 2002 ; Vol. 47, No. 2. pp. 224-231.
@article{4e8fa3af850841fe895467c8c24ced8b,
title = "Continuously moving table data acquisition method for long FOV contrast-enhanced MRA and whole-body MRI",
abstract = "A method is presented in which an extended longitudinal field of view (FOV), as required for whole-body MRI or MRA peripheral runoff studies, is acquired in one seamless image. Previous methods typically either acquired 3D data at multiple static {"}stations{"} which covered the extended FOV or as a series of 2D axial sections. The method presented here maintains the benefits of 3D acquisition while removing the discrete nature of the multistation method by continuous acquisition of MR data as the patient table moves through the desired FOV. Although the technique acquires data only from a homogeneous central volume of the magnet at any point in time, by spatially registering all data it is possible to extend the FOV well beyond this volume. The method is demonstrated experimentally with phantoms, in vivo angiographic animal studies, and in vivo human studies.",
keywords = "Extended FOV, Moving table MRI, Peripheral MR angiography, Whole body MRI",
author = "Kruger, {David G.} and Riederer, {Stephen J} and Grimm, {Roger C.} and Rossman, {Phillip J.} and Minnesota",
year = "2002",
doi = "10.1002/mrm.10061",
language = "English (US)",
volume = "47",
pages = "224--231",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",
number = "2",

}

TY - JOUR

T1 - Continuously moving table data acquisition method for long FOV contrast-enhanced MRA and whole-body MRI

AU - Kruger, David G.

AU - Riederer, Stephen J

AU - Grimm, Roger C.

AU - Rossman, Phillip J.

AU - Minnesota,

PY - 2002

Y1 - 2002

N2 - A method is presented in which an extended longitudinal field of view (FOV), as required for whole-body MRI or MRA peripheral runoff studies, is acquired in one seamless image. Previous methods typically either acquired 3D data at multiple static "stations" which covered the extended FOV or as a series of 2D axial sections. The method presented here maintains the benefits of 3D acquisition while removing the discrete nature of the multistation method by continuous acquisition of MR data as the patient table moves through the desired FOV. Although the technique acquires data only from a homogeneous central volume of the magnet at any point in time, by spatially registering all data it is possible to extend the FOV well beyond this volume. The method is demonstrated experimentally with phantoms, in vivo angiographic animal studies, and in vivo human studies.

AB - A method is presented in which an extended longitudinal field of view (FOV), as required for whole-body MRI or MRA peripheral runoff studies, is acquired in one seamless image. Previous methods typically either acquired 3D data at multiple static "stations" which covered the extended FOV or as a series of 2D axial sections. The method presented here maintains the benefits of 3D acquisition while removing the discrete nature of the multistation method by continuous acquisition of MR data as the patient table moves through the desired FOV. Although the technique acquires data only from a homogeneous central volume of the magnet at any point in time, by spatially registering all data it is possible to extend the FOV well beyond this volume. The method is demonstrated experimentally with phantoms, in vivo angiographic animal studies, and in vivo human studies.

KW - Extended FOV

KW - Moving table MRI

KW - Peripheral MR angiography

KW - Whole body MRI

UR - http://www.scopus.com/inward/record.url?scp=0036158390&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0036158390&partnerID=8YFLogxK

U2 - 10.1002/mrm.10061

DO - 10.1002/mrm.10061

M3 - Article

C2 - 11810664

AN - SCOPUS:0036158390

VL - 47

SP - 224

EP - 231

JO - Magnetic Resonance in Medicine

JF - Magnetic Resonance in Medicine

SN - 0740-3194

IS - 2

ER -