Complications of endovascular polymers associated with vascular introducer sheaths and metallic coils in 3 patients, with literature review

Michael E. Fealey, William D. Edwards, Caterina Giannini, David G. Piepgras, Harry Cloft, Charanjit S. Rihal

Research output: Contribution to journalArticlepeer-review

54 Scopus citations

Abstract

BACKGROUND: Hydrophilic coatings on arterial introducer sheaths reduce the frequency of spasm during cardiac catheterization. Moreover, during treatment of cerebral artery aneurysms, hydrophilic coatings on microcatheters ease their maneuverability, and polymers within embolization coils enhance neointimal fibrosis at the aneurysm neck. Foreign-body giant cell reactions can result when a polymer is stripped from its source. CASES 1 AND 2: A 51-year-old man and a 66-year-old woman underwent coronary angiography via the radial artery, using hydrophilic-coated sheaths. They both developed tender lesions at the access site 2 weeks later. Microscopy of resected lesions showed pieces of hydrophilic coating and granulomatous inflammation, as has been reported by others. CASE 3: A 58-year-old woman underwent embolization of a ruptured distal right internal carotid artery aneurysm using polymer-containing coils. Nine months later, she began developing multiple right-sided cerebral ring-enhancing lesions. Biopsy revealed granulomas and microabscesses, in which polymer filaments were later identified. To our knowledge, this complication has not been described previously. CONCLUSIONS: Hydrophilic coating may dislodge and induce a prominent foreign-body granulomatous response or microabscesses. Although the culprit radial artery sheath is now rarely used, embolization coils containing polymers are commonly deployed in clinical practice and may be a source of recurrent inflammatory lesions.

Original languageEnglish (US)
Pages (from-to)1310-1316
Number of pages7
JournalAmerican Journal of Surgical Pathology
Volume32
Issue number9
DOIs
StatePublished - Sep 2008

Keywords

  • Foreign-body granulomas
  • Hydrophilic coat
  • Intravascular interventional devices

ASJC Scopus subject areas

  • Anatomy
  • Surgery
  • Pathology and Forensic Medicine

Fingerprint

Dive into the research topics of 'Complications of endovascular polymers associated with vascular introducer sheaths and metallic coils in 3 patients, with literature review'. Together they form a unique fingerprint.

Cite this