Comparative evaluation of the VERSANT HCV RNA 3.0, QUANTIPLEX HCV RNA 2.0, and COBAS AMPLICOR HCV MONITOR version 2.0 assays for quantification of hepatitis C virus RNA in serum

Jeffrey J. Germer, Paul J. Heimgartner, Duane M. Ilstrup, W. Scott Harmsen, Greg D. Jenkins, Robin Patel

Research output: Contribution to journalArticle

33 Scopus citations

Abstract

A comparison of quantitative results expressed in hepatitis C virus (HCV) international units per milliliter, obtained from the VERSANT HCV RNA 3.0 (bDNA-3.0) assay, the QUANTIPLEX HCV RNA 2.0 (bDNA-2.0) assay, and the COBAS AMPLICOR HCV MONITOR version 2.0 (HCM-2.0) test was performed. A total of 168 patient specimens submitted to the Mayo Clinic Molecular Microbiology Laboratory for HCV quantification or HCV genotyping were studied. Of the specimens tested, 97, 88, and 79% yielded quantitative results within the dynamic range of the bDNA-3.0, bDNA-2.0, and HCM-2.0 assays, respectively. Overall, there was substantial agreement between the results generated by all three assays. A total of 15 out of 29 (52%) of the specimens determined to contain viral loads of <31,746 IU/ml by the bDNA-3.0 assay were categorized as containing viral loads within the range of 31,746 to 500,000 IU/ml by the bDNA-2.0 assay. Although substantial agreement was noted between the results generated by the bDNA-2.0 and bDNA-3.0 assays, a bias toward higher viral titer by the bDNA-2.0 assay was noted (P = 0.001). Likewise, although substantial agreement was noted between the results generated by the HCM-2.0 and bDNA-3.0 assays, a bias toward higher viral titer by the bDNA-3.0 assay was noted (P ≤ 0.001). The discrepancy between the HCM-2.0 and bDNA-3.0 results was more pronounced when viral loads were >500,000 IU/ml and resulted in statistically significant differences (P ≤ 0.001) in determining whether viral loads were above or below 800,000 IU/ml of HCV RNA, the proposed threshold value for tailoring the duration of combination therapy. The expression of quantitative values in HCV international units per milliliter was a strength of both the bDNA-3.0 and HCM-2.0 assays.

Original languageEnglish (US)
Pages (from-to)495-500
Number of pages6
JournalJournal of clinical microbiology
Volume40
Issue number2
DOIs
StatePublished - Feb 18 2002

    Fingerprint

ASJC Scopus subject areas

  • Microbiology (medical)

Cite this