Cementation of a dual-mobility acetabular component into a well-fixed metal shell during revision total hip arthroplasty

A biomechanical validation

Julien Wegrzyn, Andrew R. Thoreson, Olivier Guyen, David G. Lewallen, Kai Nan An

Research output: Contribution to journalArticle

14 Citations (Scopus)

Abstract

Cementation of polyethylene (PE) liners into well-fixed metal shells has become a popular option during revision total hip arthroplasty (THA) particularly for older and frail patients. Although dramatic results were reported with dual-mobility acetabular components to manage hip instability during revision THA, no study evaluated the fixation strength of the cementation of dual-mobility components into well-fixed metal shells. Eight dual-mobility and eight all-PE components were cemented into a metal shell with a uniform 2- to 3-mm cement mantle. The cemented fixation strength was evaluated using lever-out and torsion testing. The interface at which failure occurred was determined. Lever-out testing showed that dual-mobility components failed at significantly higher maximum moment than the all-PE components. No direct comparison could be performed with torsion testing due to early failure of the all-PE component itself before failure of the cement fixation. However, the maximum moments measured were dramatically higher than the in vivo frictional moments classically reported in THA. In addition, failure was always observed at the metal shell/cement interface whenever it did occur. In conclusion, a dual-mobility acetabular component cemented into a well-fixed metal shell could constitute a biomechanically acceptable alternative to acetabular shell removal or PE liner cementation while simultaneously preventing instability of the THA revision. Clinical studies are warranted.

Original languageEnglish (US)
Pages (from-to)991-997
Number of pages7
JournalJournal of Orthopaedic Research
Volume31
Issue number6
DOIs
StatePublished - Jun 2013

Fingerprint

Cementation
Polyethylene
Arthroplasty
Hip
Metals

Keywords

  • cement
  • dual-mobility acetabular component
  • fixation strength
  • instability
  • revision total hip arthroplasty

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine

Cite this

Cementation of a dual-mobility acetabular component into a well-fixed metal shell during revision total hip arthroplasty : A biomechanical validation. / Wegrzyn, Julien; Thoreson, Andrew R.; Guyen, Olivier; Lewallen, David G.; An, Kai Nan.

In: Journal of Orthopaedic Research, Vol. 31, No. 6, 06.2013, p. 991-997.

Research output: Contribution to journalArticle

Wegrzyn, Julien ; Thoreson, Andrew R. ; Guyen, Olivier ; Lewallen, David G. ; An, Kai Nan. / Cementation of a dual-mobility acetabular component into a well-fixed metal shell during revision total hip arthroplasty : A biomechanical validation. In: Journal of Orthopaedic Research. 2013 ; Vol. 31, No. 6. pp. 991-997.
@article{2dedbdd287aa4a72bededf3698aa490f,
title = "Cementation of a dual-mobility acetabular component into a well-fixed metal shell during revision total hip arthroplasty: A biomechanical validation",
abstract = "Cementation of polyethylene (PE) liners into well-fixed metal shells has become a popular option during revision total hip arthroplasty (THA) particularly for older and frail patients. Although dramatic results were reported with dual-mobility acetabular components to manage hip instability during revision THA, no study evaluated the fixation strength of the cementation of dual-mobility components into well-fixed metal shells. Eight dual-mobility and eight all-PE components were cemented into a metal shell with a uniform 2- to 3-mm cement mantle. The cemented fixation strength was evaluated using lever-out and torsion testing. The interface at which failure occurred was determined. Lever-out testing showed that dual-mobility components failed at significantly higher maximum moment than the all-PE components. No direct comparison could be performed with torsion testing due to early failure of the all-PE component itself before failure of the cement fixation. However, the maximum moments measured were dramatically higher than the in vivo frictional moments classically reported in THA. In addition, failure was always observed at the metal shell/cement interface whenever it did occur. In conclusion, a dual-mobility acetabular component cemented into a well-fixed metal shell could constitute a biomechanically acceptable alternative to acetabular shell removal or PE liner cementation while simultaneously preventing instability of the THA revision. Clinical studies are warranted.",
keywords = "cement, dual-mobility acetabular component, fixation strength, instability, revision total hip arthroplasty",
author = "Julien Wegrzyn and Thoreson, {Andrew R.} and Olivier Guyen and Lewallen, {David G.} and An, {Kai Nan}",
year = "2013",
month = "6",
doi = "10.1002/jor.22314",
language = "English (US)",
volume = "31",
pages = "991--997",
journal = "Journal of Orthopaedic Research",
issn = "0736-0266",
publisher = "John Wiley and Sons Inc.",
number = "6",

}

TY - JOUR

T1 - Cementation of a dual-mobility acetabular component into a well-fixed metal shell during revision total hip arthroplasty

T2 - A biomechanical validation

AU - Wegrzyn, Julien

AU - Thoreson, Andrew R.

AU - Guyen, Olivier

AU - Lewallen, David G.

AU - An, Kai Nan

PY - 2013/6

Y1 - 2013/6

N2 - Cementation of polyethylene (PE) liners into well-fixed metal shells has become a popular option during revision total hip arthroplasty (THA) particularly for older and frail patients. Although dramatic results were reported with dual-mobility acetabular components to manage hip instability during revision THA, no study evaluated the fixation strength of the cementation of dual-mobility components into well-fixed metal shells. Eight dual-mobility and eight all-PE components were cemented into a metal shell with a uniform 2- to 3-mm cement mantle. The cemented fixation strength was evaluated using lever-out and torsion testing. The interface at which failure occurred was determined. Lever-out testing showed that dual-mobility components failed at significantly higher maximum moment than the all-PE components. No direct comparison could be performed with torsion testing due to early failure of the all-PE component itself before failure of the cement fixation. However, the maximum moments measured were dramatically higher than the in vivo frictional moments classically reported in THA. In addition, failure was always observed at the metal shell/cement interface whenever it did occur. In conclusion, a dual-mobility acetabular component cemented into a well-fixed metal shell could constitute a biomechanically acceptable alternative to acetabular shell removal or PE liner cementation while simultaneously preventing instability of the THA revision. Clinical studies are warranted.

AB - Cementation of polyethylene (PE) liners into well-fixed metal shells has become a popular option during revision total hip arthroplasty (THA) particularly for older and frail patients. Although dramatic results were reported with dual-mobility acetabular components to manage hip instability during revision THA, no study evaluated the fixation strength of the cementation of dual-mobility components into well-fixed metal shells. Eight dual-mobility and eight all-PE components were cemented into a metal shell with a uniform 2- to 3-mm cement mantle. The cemented fixation strength was evaluated using lever-out and torsion testing. The interface at which failure occurred was determined. Lever-out testing showed that dual-mobility components failed at significantly higher maximum moment than the all-PE components. No direct comparison could be performed with torsion testing due to early failure of the all-PE component itself before failure of the cement fixation. However, the maximum moments measured were dramatically higher than the in vivo frictional moments classically reported in THA. In addition, failure was always observed at the metal shell/cement interface whenever it did occur. In conclusion, a dual-mobility acetabular component cemented into a well-fixed metal shell could constitute a biomechanically acceptable alternative to acetabular shell removal or PE liner cementation while simultaneously preventing instability of the THA revision. Clinical studies are warranted.

KW - cement

KW - dual-mobility acetabular component

KW - fixation strength

KW - instability

KW - revision total hip arthroplasty

UR - http://www.scopus.com/inward/record.url?scp=84876289014&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84876289014&partnerID=8YFLogxK

U2 - 10.1002/jor.22314

DO - 10.1002/jor.22314

M3 - Article

VL - 31

SP - 991

EP - 997

JO - Journal of Orthopaedic Research

JF - Journal of Orthopaedic Research

SN - 0736-0266

IS - 6

ER -