Catheter tip orientation affects radiofrequency ablation lesion size in the canine left ventricle

Sumeet S. Chugh, Rodrigo C. Chan, Susan B. Johnson, Douglas L Packer

Research output: Contribution to journalArticle

31 Citations (Scopus)

Abstract

While some factors influencing size of RF lesions in ventricular tissue have been characterized, the effects of catheter electrode-endocardial surface orientation on lesion generation have not been investigated. Therefore, the effects of parallel versus perpendicular catheter electrode- endocardial surface orientation on dimensions of RF lesion produced with 4-, 6-, 8-, 10-, and 12-mm distal electrode lengths were studied in 20 closed- chested dogs. Orientation was established by biplane fluoroscopy and confirmed by intracardiac echocardiography for the majority of energy deliveries (71%). RF voltage was titrated to maintain constant catheter electrode temperature of 75°C for 60 seconds. In the perpendicular orientation, lesion size did not change significantly with increasing electrode lengths. There was a statistically significant interaction between electrode orientation and maximum lesion length (analysis of variance [ANOVA] P = 0.04), lesion width (ANOVA P = 0.01), lesion area (ANOVA P = 0.02), and estimated lesion volume (ANOVA P < 0.005) over all electrode lengths. With parallel tip-tissue orientation, lesion size was a function of increasing electrode length. For 4-, 6-, 8-, 10-, and 12-mm electrodes, maximum lesion surface areas were 95 ± 38, 97 ± 38, 119 ± 29, 147 ± 52, and 147 ± 67 mm 2, respectively. For electrode lengths 8, 10, and 12 mm, estimated lesion volumes were significantly greater with parallel orientation (P < 0.05 for all). Thus, ventricular lesion size is dependent on catheter electrode length, but only when the catheter is oriented parallel to the endocardial surface. This information may be helpful in increasing lesion dimensions for RF ablation of ventricular tachycardias.

Original languageEnglish (US)
Pages (from-to)413-420
Number of pages8
JournalPACE - Pacing and Clinical Electrophysiology
Volume22
Issue number3
StatePublished - 1999

Fingerprint

Heart Ventricles
Canidae
Electrodes
Catheters
Analysis of Variance
Fluoroscopy
Ventricular Tachycardia
Echocardiography
Thorax
Dogs
Temperature

Keywords

  • Intracardiac ultrasound
  • Orientation
  • Radiofrequency ablation
  • Ventricle
  • Ventricular tachycardia

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this

Catheter tip orientation affects radiofrequency ablation lesion size in the canine left ventricle. / Chugh, Sumeet S.; Chan, Rodrigo C.; Johnson, Susan B.; Packer, Douglas L.

In: PACE - Pacing and Clinical Electrophysiology, Vol. 22, No. 3, 1999, p. 413-420.

Research output: Contribution to journalArticle

@article{670b3b5cd4f44a0f9b5ccea58117a606,
title = "Catheter tip orientation affects radiofrequency ablation lesion size in the canine left ventricle",
abstract = "While some factors influencing size of RF lesions in ventricular tissue have been characterized, the effects of catheter electrode-endocardial surface orientation on lesion generation have not been investigated. Therefore, the effects of parallel versus perpendicular catheter electrode- endocardial surface orientation on dimensions of RF lesion produced with 4-, 6-, 8-, 10-, and 12-mm distal electrode lengths were studied in 20 closed- chested dogs. Orientation was established by biplane fluoroscopy and confirmed by intracardiac echocardiography for the majority of energy deliveries (71{\%}). RF voltage was titrated to maintain constant catheter electrode temperature of 75°C for 60 seconds. In the perpendicular orientation, lesion size did not change significantly with increasing electrode lengths. There was a statistically significant interaction between electrode orientation and maximum lesion length (analysis of variance [ANOVA] P = 0.04), lesion width (ANOVA P = 0.01), lesion area (ANOVA P = 0.02), and estimated lesion volume (ANOVA P < 0.005) over all electrode lengths. With parallel tip-tissue orientation, lesion size was a function of increasing electrode length. For 4-, 6-, 8-, 10-, and 12-mm electrodes, maximum lesion surface areas were 95 ± 38, 97 ± 38, 119 ± 29, 147 ± 52, and 147 ± 67 mm 2, respectively. For electrode lengths 8, 10, and 12 mm, estimated lesion volumes were significantly greater with parallel orientation (P < 0.05 for all). Thus, ventricular lesion size is dependent on catheter electrode length, but only when the catheter is oriented parallel to the endocardial surface. This information may be helpful in increasing lesion dimensions for RF ablation of ventricular tachycardias.",
keywords = "Intracardiac ultrasound, Orientation, Radiofrequency ablation, Ventricle, Ventricular tachycardia",
author = "Chugh, {Sumeet S.} and Chan, {Rodrigo C.} and Johnson, {Susan B.} and Packer, {Douglas L}",
year = "1999",
language = "English (US)",
volume = "22",
pages = "413--420",
journal = "PACE - Pacing and Clinical Electrophysiology",
issn = "0147-8389",
publisher = "Wiley-Blackwell",
number = "3",

}

TY - JOUR

T1 - Catheter tip orientation affects radiofrequency ablation lesion size in the canine left ventricle

AU - Chugh, Sumeet S.

AU - Chan, Rodrigo C.

AU - Johnson, Susan B.

AU - Packer, Douglas L

PY - 1999

Y1 - 1999

N2 - While some factors influencing size of RF lesions in ventricular tissue have been characterized, the effects of catheter electrode-endocardial surface orientation on lesion generation have not been investigated. Therefore, the effects of parallel versus perpendicular catheter electrode- endocardial surface orientation on dimensions of RF lesion produced with 4-, 6-, 8-, 10-, and 12-mm distal electrode lengths were studied in 20 closed- chested dogs. Orientation was established by biplane fluoroscopy and confirmed by intracardiac echocardiography for the majority of energy deliveries (71%). RF voltage was titrated to maintain constant catheter electrode temperature of 75°C for 60 seconds. In the perpendicular orientation, lesion size did not change significantly with increasing electrode lengths. There was a statistically significant interaction between electrode orientation and maximum lesion length (analysis of variance [ANOVA] P = 0.04), lesion width (ANOVA P = 0.01), lesion area (ANOVA P = 0.02), and estimated lesion volume (ANOVA P < 0.005) over all electrode lengths. With parallel tip-tissue orientation, lesion size was a function of increasing electrode length. For 4-, 6-, 8-, 10-, and 12-mm electrodes, maximum lesion surface areas were 95 ± 38, 97 ± 38, 119 ± 29, 147 ± 52, and 147 ± 67 mm 2, respectively. For electrode lengths 8, 10, and 12 mm, estimated lesion volumes were significantly greater with parallel orientation (P < 0.05 for all). Thus, ventricular lesion size is dependent on catheter electrode length, but only when the catheter is oriented parallel to the endocardial surface. This information may be helpful in increasing lesion dimensions for RF ablation of ventricular tachycardias.

AB - While some factors influencing size of RF lesions in ventricular tissue have been characterized, the effects of catheter electrode-endocardial surface orientation on lesion generation have not been investigated. Therefore, the effects of parallel versus perpendicular catheter electrode- endocardial surface orientation on dimensions of RF lesion produced with 4-, 6-, 8-, 10-, and 12-mm distal electrode lengths were studied in 20 closed- chested dogs. Orientation was established by biplane fluoroscopy and confirmed by intracardiac echocardiography for the majority of energy deliveries (71%). RF voltage was titrated to maintain constant catheter electrode temperature of 75°C for 60 seconds. In the perpendicular orientation, lesion size did not change significantly with increasing electrode lengths. There was a statistically significant interaction between electrode orientation and maximum lesion length (analysis of variance [ANOVA] P = 0.04), lesion width (ANOVA P = 0.01), lesion area (ANOVA P = 0.02), and estimated lesion volume (ANOVA P < 0.005) over all electrode lengths. With parallel tip-tissue orientation, lesion size was a function of increasing electrode length. For 4-, 6-, 8-, 10-, and 12-mm electrodes, maximum lesion surface areas were 95 ± 38, 97 ± 38, 119 ± 29, 147 ± 52, and 147 ± 67 mm 2, respectively. For electrode lengths 8, 10, and 12 mm, estimated lesion volumes were significantly greater with parallel orientation (P < 0.05 for all). Thus, ventricular lesion size is dependent on catheter electrode length, but only when the catheter is oriented parallel to the endocardial surface. This information may be helpful in increasing lesion dimensions for RF ablation of ventricular tachycardias.

KW - Intracardiac ultrasound

KW - Orientation

KW - Radiofrequency ablation

KW - Ventricle

KW - Ventricular tachycardia

UR - http://www.scopus.com/inward/record.url?scp=0033063965&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0033063965&partnerID=8YFLogxK

M3 - Article

C2 - 10192850

AN - SCOPUS:0033063965

VL - 22

SP - 413

EP - 420

JO - PACE - Pacing and Clinical Electrophysiology

JF - PACE - Pacing and Clinical Electrophysiology

SN - 0147-8389

IS - 3

ER -