Case study a novel approach to scavenging anesthetic gases in rodent surgery

James Couch, Jeffrey C. Nesbitt, Dale A. Krageschmidt, Michael C Blanco

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

Laboratory animal procedures using gas anesthetics may amass elevated waste gas concentrations in operating rooms if controls are not implemented for capturing and removing the vapors. Area sampling using an infrared analyzer indicated isoflurane concentrations likely to exceed occupational exposure guidelines. Our study showed environmental concentrations of oxygen as high as 40% and isoflurane concentrations > 100 ppm when no controls or merely passive controls were utilized. These extraneous isoflurane emissions were determined to be originating from the pre-procedural induction process as well as the gas delivery nose cone. A novel waste gas collection cylinder was designed to enclose the gas delivery nose cone and animal head during the administration of anesthetic gases. The vented cylinder utilized a house vacuum to remove the waste anesthetic gases from the surgical field. A commercially available induction chamber designed to be actively and externally exhausted was used to lower concentrations during the induction process. With implementation of local exhaust ventilation controls, waste anesthetic gas concentrations decreased to below recommended occupational exposure levels. In vitro (sham) testing compared favorably to in vivo measurements validating the reduction capability of active ventilation during rodent anesthetic administration. In vivo isoflurane reductions for the induction chamber emissions, the operating room, and the surgeon's breathing zone were 95%, 60%, and 53%, respectively. The same measurements for an in vitro procedure were 98%, 84%, and 87%, respectively.

Original languageEnglish (US)
JournalJournal of Occupational and Environmental Hygiene
Volume10
Issue number9
DOIs
StatePublished - 2013

Fingerprint

Inhalation Anesthetics
Isoflurane
Rodentia
Oil and Gas Fields
Gases
Operating Rooms
Occupational Exposure
Nose
Ventilation
Laboratory Animals
Vacuum
Anesthetics
Respiration
Head
Guidelines
Oxygen
In Vitro Techniques

ASJC Scopus subject areas

  • Public Health, Environmental and Occupational Health
  • Medicine(all)

Cite this

Case study a novel approach to scavenging anesthetic gases in rodent surgery. / Couch, James; Nesbitt, Jeffrey C.; Krageschmidt, Dale A.; Blanco, Michael C.

In: Journal of Occupational and Environmental Hygiene, Vol. 10, No. 9, 2013.

Research output: Contribution to journalArticle

Couch, James ; Nesbitt, Jeffrey C. ; Krageschmidt, Dale A. ; Blanco, Michael C. / Case study a novel approach to scavenging anesthetic gases in rodent surgery. In: Journal of Occupational and Environmental Hygiene. 2013 ; Vol. 10, No. 9.
@article{dcb4c124ccb643808af5ddcb4a3edf11,
title = "Case study a novel approach to scavenging anesthetic gases in rodent surgery",
abstract = "Laboratory animal procedures using gas anesthetics may amass elevated waste gas concentrations in operating rooms if controls are not implemented for capturing and removing the vapors. Area sampling using an infrared analyzer indicated isoflurane concentrations likely to exceed occupational exposure guidelines. Our study showed environmental concentrations of oxygen as high as 40{\%} and isoflurane concentrations > 100 ppm when no controls or merely passive controls were utilized. These extraneous isoflurane emissions were determined to be originating from the pre-procedural induction process as well as the gas delivery nose cone. A novel waste gas collection cylinder was designed to enclose the gas delivery nose cone and animal head during the administration of anesthetic gases. The vented cylinder utilized a house vacuum to remove the waste anesthetic gases from the surgical field. A commercially available induction chamber designed to be actively and externally exhausted was used to lower concentrations during the induction process. With implementation of local exhaust ventilation controls, waste anesthetic gas concentrations decreased to below recommended occupational exposure levels. In vitro (sham) testing compared favorably to in vivo measurements validating the reduction capability of active ventilation during rodent anesthetic administration. In vivo isoflurane reductions for the induction chamber emissions, the operating room, and the surgeon's breathing zone were 95{\%}, 60{\%}, and 53{\%}, respectively. The same measurements for an in vitro procedure were 98{\%}, 84{\%}, and 87{\%}, respectively.",
author = "James Couch and Nesbitt, {Jeffrey C.} and Krageschmidt, {Dale A.} and Blanco, {Michael C}",
year = "2013",
doi = "10.1080/15459624.2013.818242",
language = "English (US)",
volume = "10",
journal = "Journal of Occupational and Environmental Hygiene",
issn = "1545-9624",
publisher = "Taylor and Francis Ltd.",
number = "9",

}

TY - JOUR

T1 - Case study a novel approach to scavenging anesthetic gases in rodent surgery

AU - Couch, James

AU - Nesbitt, Jeffrey C.

AU - Krageschmidt, Dale A.

AU - Blanco, Michael C

PY - 2013

Y1 - 2013

N2 - Laboratory animal procedures using gas anesthetics may amass elevated waste gas concentrations in operating rooms if controls are not implemented for capturing and removing the vapors. Area sampling using an infrared analyzer indicated isoflurane concentrations likely to exceed occupational exposure guidelines. Our study showed environmental concentrations of oxygen as high as 40% and isoflurane concentrations > 100 ppm when no controls or merely passive controls were utilized. These extraneous isoflurane emissions were determined to be originating from the pre-procedural induction process as well as the gas delivery nose cone. A novel waste gas collection cylinder was designed to enclose the gas delivery nose cone and animal head during the administration of anesthetic gases. The vented cylinder utilized a house vacuum to remove the waste anesthetic gases from the surgical field. A commercially available induction chamber designed to be actively and externally exhausted was used to lower concentrations during the induction process. With implementation of local exhaust ventilation controls, waste anesthetic gas concentrations decreased to below recommended occupational exposure levels. In vitro (sham) testing compared favorably to in vivo measurements validating the reduction capability of active ventilation during rodent anesthetic administration. In vivo isoflurane reductions for the induction chamber emissions, the operating room, and the surgeon's breathing zone were 95%, 60%, and 53%, respectively. The same measurements for an in vitro procedure were 98%, 84%, and 87%, respectively.

AB - Laboratory animal procedures using gas anesthetics may amass elevated waste gas concentrations in operating rooms if controls are not implemented for capturing and removing the vapors. Area sampling using an infrared analyzer indicated isoflurane concentrations likely to exceed occupational exposure guidelines. Our study showed environmental concentrations of oxygen as high as 40% and isoflurane concentrations > 100 ppm when no controls or merely passive controls were utilized. These extraneous isoflurane emissions were determined to be originating from the pre-procedural induction process as well as the gas delivery nose cone. A novel waste gas collection cylinder was designed to enclose the gas delivery nose cone and animal head during the administration of anesthetic gases. The vented cylinder utilized a house vacuum to remove the waste anesthetic gases from the surgical field. A commercially available induction chamber designed to be actively and externally exhausted was used to lower concentrations during the induction process. With implementation of local exhaust ventilation controls, waste anesthetic gas concentrations decreased to below recommended occupational exposure levels. In vitro (sham) testing compared favorably to in vivo measurements validating the reduction capability of active ventilation during rodent anesthetic administration. In vivo isoflurane reductions for the induction chamber emissions, the operating room, and the surgeon's breathing zone were 95%, 60%, and 53%, respectively. The same measurements for an in vitro procedure were 98%, 84%, and 87%, respectively.

UR - http://www.scopus.com/inward/record.url?scp=84884494005&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84884494005&partnerID=8YFLogxK

U2 - 10.1080/15459624.2013.818242

DO - 10.1080/15459624.2013.818242

M3 - Article

C2 - 23915295

AN - SCOPUS:84884494005

VL - 10

JO - Journal of Occupational and Environmental Hygiene

JF - Journal of Occupational and Environmental Hygiene

SN - 1545-9624

IS - 9

ER -