Carotid intima-media thickness measurements are not affected by the ultrasound frequency

Kapil M. Bhagirath, Steven Jay Lester, Julie Humphries, Joseph G. Hentz, R. Todd Hurst

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

Objective: Carotid intima-media thickness (IMT) is a B-mode ultrasound measure of subclinical atherosclerosis predictive of future cardiovascular risk. Carotid IMT measurements were historically obtained at an ultrasound frequency of 8 MHz or lower, but it is unknown whether measurements obtained at higher frequencies using newer, more advanced ultrasound technology allow for valid comparison to the older general population databases that are commonly used for the interpretation of carotid IMT results. Methods: Carotid IMT studies were conducted in 35 consecutive patients at standard (8 MHz) and high (14 MHz) frequencies and measurements were performed by two independent expert readers. Systematic bias was assessed by using the paired t-test and agreement was analyzed with the Bland-Altman method. Results: The sample mean carotid IMT obtained at 14 MHz was 0.006 mm lower than that obtained at 8 MHz. The 95% confidence interval (CI) for the mean difference between frequencies indicated that the population mean for 14 MHz is unlikely to be more than 0.02 mm lower than for 8 MHz (95% CI -0.017 to 0.004). The 95% reference range for the difference between the two transducer frequencies indicated that the thickness obtained at 14 MHz was within 0.05 mm of that obtained at 8 MHz for 95% of subjects. Conclusions: Carotid IMT measurements obtained at higher transducer frequencies are similar to those obtained at standard frequency. This finding has important clinical implications because it validates comparison of carotid IMT measurements obtained with newer, more advanced ultrasound technology with the landmark reference carotid IMT studies commonly used for interpretation of carotid IMT results.

Original languageEnglish (US)
Pages (from-to)354-357
Number of pages4
JournalEchocardiography
Volume29
Issue number3
DOIs
StatePublished - Mar 2012

Fingerprint

Carotid Intima-Media Thickness
Transducers
Confidence Intervals
Technology
Population
Atherosclerosis
Reference Values
Databases

Keywords

  • atherosclerosis
  • cardiovascular risk
  • carotid intima-media thickness
  • ultrasound frequency

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Radiology Nuclear Medicine and imaging

Cite this

Carotid intima-media thickness measurements are not affected by the ultrasound frequency. / Bhagirath, Kapil M.; Lester, Steven Jay; Humphries, Julie; Hentz, Joseph G.; Hurst, R. Todd.

In: Echocardiography, Vol. 29, No. 3, 03.2012, p. 354-357.

Research output: Contribution to journalArticle

Bhagirath, Kapil M. ; Lester, Steven Jay ; Humphries, Julie ; Hentz, Joseph G. ; Hurst, R. Todd. / Carotid intima-media thickness measurements are not affected by the ultrasound frequency. In: Echocardiography. 2012 ; Vol. 29, No. 3. pp. 354-357.
@article{f9c281bd24544197b4f001cb7afa2a13,
title = "Carotid intima-media thickness measurements are not affected by the ultrasound frequency",
abstract = "Objective: Carotid intima-media thickness (IMT) is a B-mode ultrasound measure of subclinical atherosclerosis predictive of future cardiovascular risk. Carotid IMT measurements were historically obtained at an ultrasound frequency of 8 MHz or lower, but it is unknown whether measurements obtained at higher frequencies using newer, more advanced ultrasound technology allow for valid comparison to the older general population databases that are commonly used for the interpretation of carotid IMT results. Methods: Carotid IMT studies were conducted in 35 consecutive patients at standard (8 MHz) and high (14 MHz) frequencies and measurements were performed by two independent expert readers. Systematic bias was assessed by using the paired t-test and agreement was analyzed with the Bland-Altman method. Results: The sample mean carotid IMT obtained at 14 MHz was 0.006 mm lower than that obtained at 8 MHz. The 95{\%} confidence interval (CI) for the mean difference between frequencies indicated that the population mean for 14 MHz is unlikely to be more than 0.02 mm lower than for 8 MHz (95{\%} CI -0.017 to 0.004). The 95{\%} reference range for the difference between the two transducer frequencies indicated that the thickness obtained at 14 MHz was within 0.05 mm of that obtained at 8 MHz for 95{\%} of subjects. Conclusions: Carotid IMT measurements obtained at higher transducer frequencies are similar to those obtained at standard frequency. This finding has important clinical implications because it validates comparison of carotid IMT measurements obtained with newer, more advanced ultrasound technology with the landmark reference carotid IMT studies commonly used for interpretation of carotid IMT results.",
keywords = "atherosclerosis, cardiovascular risk, carotid intima-media thickness, ultrasound frequency",
author = "Bhagirath, {Kapil M.} and Lester, {Steven Jay} and Julie Humphries and Hentz, {Joseph G.} and Hurst, {R. Todd}",
year = "2012",
month = "3",
doi = "10.1111/j.1540-8175.2011.01557.x",
language = "English (US)",
volume = "29",
pages = "354--357",
journal = "Echocardiography",
issn = "0742-2822",
publisher = "Wiley-Blackwell",
number = "3",

}

TY - JOUR

T1 - Carotid intima-media thickness measurements are not affected by the ultrasound frequency

AU - Bhagirath, Kapil M.

AU - Lester, Steven Jay

AU - Humphries, Julie

AU - Hentz, Joseph G.

AU - Hurst, R. Todd

PY - 2012/3

Y1 - 2012/3

N2 - Objective: Carotid intima-media thickness (IMT) is a B-mode ultrasound measure of subclinical atherosclerosis predictive of future cardiovascular risk. Carotid IMT measurements were historically obtained at an ultrasound frequency of 8 MHz or lower, but it is unknown whether measurements obtained at higher frequencies using newer, more advanced ultrasound technology allow for valid comparison to the older general population databases that are commonly used for the interpretation of carotid IMT results. Methods: Carotid IMT studies were conducted in 35 consecutive patients at standard (8 MHz) and high (14 MHz) frequencies and measurements were performed by two independent expert readers. Systematic bias was assessed by using the paired t-test and agreement was analyzed with the Bland-Altman method. Results: The sample mean carotid IMT obtained at 14 MHz was 0.006 mm lower than that obtained at 8 MHz. The 95% confidence interval (CI) for the mean difference between frequencies indicated that the population mean for 14 MHz is unlikely to be more than 0.02 mm lower than for 8 MHz (95% CI -0.017 to 0.004). The 95% reference range for the difference between the two transducer frequencies indicated that the thickness obtained at 14 MHz was within 0.05 mm of that obtained at 8 MHz for 95% of subjects. Conclusions: Carotid IMT measurements obtained at higher transducer frequencies are similar to those obtained at standard frequency. This finding has important clinical implications because it validates comparison of carotid IMT measurements obtained with newer, more advanced ultrasound technology with the landmark reference carotid IMT studies commonly used for interpretation of carotid IMT results.

AB - Objective: Carotid intima-media thickness (IMT) is a B-mode ultrasound measure of subclinical atherosclerosis predictive of future cardiovascular risk. Carotid IMT measurements were historically obtained at an ultrasound frequency of 8 MHz or lower, but it is unknown whether measurements obtained at higher frequencies using newer, more advanced ultrasound technology allow for valid comparison to the older general population databases that are commonly used for the interpretation of carotid IMT results. Methods: Carotid IMT studies were conducted in 35 consecutive patients at standard (8 MHz) and high (14 MHz) frequencies and measurements were performed by two independent expert readers. Systematic bias was assessed by using the paired t-test and agreement was analyzed with the Bland-Altman method. Results: The sample mean carotid IMT obtained at 14 MHz was 0.006 mm lower than that obtained at 8 MHz. The 95% confidence interval (CI) for the mean difference between frequencies indicated that the population mean for 14 MHz is unlikely to be more than 0.02 mm lower than for 8 MHz (95% CI -0.017 to 0.004). The 95% reference range for the difference between the two transducer frequencies indicated that the thickness obtained at 14 MHz was within 0.05 mm of that obtained at 8 MHz for 95% of subjects. Conclusions: Carotid IMT measurements obtained at higher transducer frequencies are similar to those obtained at standard frequency. This finding has important clinical implications because it validates comparison of carotid IMT measurements obtained with newer, more advanced ultrasound technology with the landmark reference carotid IMT studies commonly used for interpretation of carotid IMT results.

KW - atherosclerosis

KW - cardiovascular risk

KW - carotid intima-media thickness

KW - ultrasound frequency

UR - http://www.scopus.com/inward/record.url?scp=84858793065&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84858793065&partnerID=8YFLogxK

U2 - 10.1111/j.1540-8175.2011.01557.x

DO - 10.1111/j.1540-8175.2011.01557.x

M3 - Article

C2 - 22066737

AN - SCOPUS:84858793065

VL - 29

SP - 354

EP - 357

JO - Echocardiography

JF - Echocardiography

SN - 0742-2822

IS - 3

ER -