Biomechanical evaluation of posterior screw fixation in cadaveric cervical spines

Panayiotis J. Papagelopoulos, Bradford L. Currier, Patricia G. Neale, Yukitaka Hokari, Lawrence J. Berglund, Dirk R. Larson, Dean R. Fisher, Kai N. An

Research output: Contribution to journalArticle

27 Citations (Scopus)

Abstract

Sixteen fresh-frozen spines from cadavers (C4-T1) were randomized on the basis of dual energy xray absorptiometry analysis of bone mineral density. The specimens were subjected to physiologic loads (≤1.5 N-m) in a device that applied pure unconstrained, flexion and extension, lateral bending, and torsional moments. After a major injury, including a wide C6 laminectomy, bilateral capsulectomies, and 65% facetectomy at C6-C7, four constructs were applied to each specimen in a balanced sequence, and the stability tests were repeated. All the constructs were identical posterior cervical rods secured to the spine with lateral mass screws in C5 and pedicle screws in C7. The constructs differed only in the presence or absence of a transverse connector between the rods, presence or absence of lateral mass screws in C6, and unicortical or bicortical lateral mass screws. Insertional screw torque was higher in specimens with greater bone mineral density. Pedicle screws had greater torque than lateral mass screws. Posterior cervical spine rod fixation provided an equivalent stability by use of either unicortical or bicortical lateral mass screws. A transverse stabilizer between the rods reduces the number of lateral mass screws needed in posterior cervical spinal rod systems. Repeated stability tests, even when done with physiologic loads, may compromise the screw-bone interface. This effect is most pronounced in specimens that have low bone mineral density.

Original languageEnglish (US)
Pages (from-to)13-24
Number of pages12
JournalClinical Orthopaedics and Related Research
Issue number411
StatePublished - Jun 1 2003

Fingerprint

Bone Density
Spine
Torque
Bone Screws
Laminectomy
Cadaver
X-Rays
Equipment and Supplies
Wounds and Injuries
Pedicle Screws

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine
  • Surgery

Cite this

Papagelopoulos, P. J., Currier, B. L., Neale, P. G., Hokari, Y., Berglund, L. J., Larson, D. R., ... An, K. N. (2003). Biomechanical evaluation of posterior screw fixation in cadaveric cervical spines. Clinical Orthopaedics and Related Research, (411), 13-24.

Biomechanical evaluation of posterior screw fixation in cadaveric cervical spines. / Papagelopoulos, Panayiotis J.; Currier, Bradford L.; Neale, Patricia G.; Hokari, Yukitaka; Berglund, Lawrence J.; Larson, Dirk R.; Fisher, Dean R.; An, Kai N.

In: Clinical Orthopaedics and Related Research, No. 411, 01.06.2003, p. 13-24.

Research output: Contribution to journalArticle

Papagelopoulos, PJ, Currier, BL, Neale, PG, Hokari, Y, Berglund, LJ, Larson, DR, Fisher, DR & An, KN 2003, 'Biomechanical evaluation of posterior screw fixation in cadaveric cervical spines', Clinical Orthopaedics and Related Research, no. 411, pp. 13-24.
Papagelopoulos PJ, Currier BL, Neale PG, Hokari Y, Berglund LJ, Larson DR et al. Biomechanical evaluation of posterior screw fixation in cadaveric cervical spines. Clinical Orthopaedics and Related Research. 2003 Jun 1;(411):13-24.
Papagelopoulos, Panayiotis J. ; Currier, Bradford L. ; Neale, Patricia G. ; Hokari, Yukitaka ; Berglund, Lawrence J. ; Larson, Dirk R. ; Fisher, Dean R. ; An, Kai N. / Biomechanical evaluation of posterior screw fixation in cadaveric cervical spines. In: Clinical Orthopaedics and Related Research. 2003 ; No. 411. pp. 13-24.
@article{24e8849c51644fef885c63487ffac7f3,
title = "Biomechanical evaluation of posterior screw fixation in cadaveric cervical spines",
abstract = "Sixteen fresh-frozen spines from cadavers (C4-T1) were randomized on the basis of dual energy xray absorptiometry analysis of bone mineral density. The specimens were subjected to physiologic loads (≤1.5 N-m) in a device that applied pure unconstrained, flexion and extension, lateral bending, and torsional moments. After a major injury, including a wide C6 laminectomy, bilateral capsulectomies, and 65{\%} facetectomy at C6-C7, four constructs were applied to each specimen in a balanced sequence, and the stability tests were repeated. All the constructs were identical posterior cervical rods secured to the spine with lateral mass screws in C5 and pedicle screws in C7. The constructs differed only in the presence or absence of a transverse connector between the rods, presence or absence of lateral mass screws in C6, and unicortical or bicortical lateral mass screws. Insertional screw torque was higher in specimens with greater bone mineral density. Pedicle screws had greater torque than lateral mass screws. Posterior cervical spine rod fixation provided an equivalent stability by use of either unicortical or bicortical lateral mass screws. A transverse stabilizer between the rods reduces the number of lateral mass screws needed in posterior cervical spinal rod systems. Repeated stability tests, even when done with physiologic loads, may compromise the screw-bone interface. This effect is most pronounced in specimens that have low bone mineral density.",
author = "Papagelopoulos, {Panayiotis J.} and Currier, {Bradford L.} and Neale, {Patricia G.} and Yukitaka Hokari and Berglund, {Lawrence J.} and Larson, {Dirk R.} and Fisher, {Dean R.} and An, {Kai N.}",
year = "2003",
month = "6",
day = "1",
language = "English (US)",
pages = "13--24",
journal = "Clinical Orthopaedics and Related Research",
issn = "0009-921X",
publisher = "Springer New York",
number = "411",

}

TY - JOUR

T1 - Biomechanical evaluation of posterior screw fixation in cadaveric cervical spines

AU - Papagelopoulos, Panayiotis J.

AU - Currier, Bradford L.

AU - Neale, Patricia G.

AU - Hokari, Yukitaka

AU - Berglund, Lawrence J.

AU - Larson, Dirk R.

AU - Fisher, Dean R.

AU - An, Kai N.

PY - 2003/6/1

Y1 - 2003/6/1

N2 - Sixteen fresh-frozen spines from cadavers (C4-T1) were randomized on the basis of dual energy xray absorptiometry analysis of bone mineral density. The specimens were subjected to physiologic loads (≤1.5 N-m) in a device that applied pure unconstrained, flexion and extension, lateral bending, and torsional moments. After a major injury, including a wide C6 laminectomy, bilateral capsulectomies, and 65% facetectomy at C6-C7, four constructs were applied to each specimen in a balanced sequence, and the stability tests were repeated. All the constructs were identical posterior cervical rods secured to the spine with lateral mass screws in C5 and pedicle screws in C7. The constructs differed only in the presence or absence of a transverse connector between the rods, presence or absence of lateral mass screws in C6, and unicortical or bicortical lateral mass screws. Insertional screw torque was higher in specimens with greater bone mineral density. Pedicle screws had greater torque than lateral mass screws. Posterior cervical spine rod fixation provided an equivalent stability by use of either unicortical or bicortical lateral mass screws. A transverse stabilizer between the rods reduces the number of lateral mass screws needed in posterior cervical spinal rod systems. Repeated stability tests, even when done with physiologic loads, may compromise the screw-bone interface. This effect is most pronounced in specimens that have low bone mineral density.

AB - Sixteen fresh-frozen spines from cadavers (C4-T1) were randomized on the basis of dual energy xray absorptiometry analysis of bone mineral density. The specimens were subjected to physiologic loads (≤1.5 N-m) in a device that applied pure unconstrained, flexion and extension, lateral bending, and torsional moments. After a major injury, including a wide C6 laminectomy, bilateral capsulectomies, and 65% facetectomy at C6-C7, four constructs were applied to each specimen in a balanced sequence, and the stability tests were repeated. All the constructs were identical posterior cervical rods secured to the spine with lateral mass screws in C5 and pedicle screws in C7. The constructs differed only in the presence or absence of a transverse connector between the rods, presence or absence of lateral mass screws in C6, and unicortical or bicortical lateral mass screws. Insertional screw torque was higher in specimens with greater bone mineral density. Pedicle screws had greater torque than lateral mass screws. Posterior cervical spine rod fixation provided an equivalent stability by use of either unicortical or bicortical lateral mass screws. A transverse stabilizer between the rods reduces the number of lateral mass screws needed in posterior cervical spinal rod systems. Repeated stability tests, even when done with physiologic loads, may compromise the screw-bone interface. This effect is most pronounced in specimens that have low bone mineral density.

UR - http://www.scopus.com/inward/record.url?scp=0037904404&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0037904404&partnerID=8YFLogxK

M3 - Article

C2 - 12782855

AN - SCOPUS:0037904404

SP - 13

EP - 24

JO - Clinical Orthopaedics and Related Research

JF - Clinical Orthopaedics and Related Research

SN - 0009-921X

IS - 411

ER -