Balloon-assisted coiling through a 5-French system

Bradley J. White, Kennith F. Layton, David F Kallmes, Harry J. Cloft

Research output: Contribution to journalArticle

7 Citations (Scopus)

Abstract

Introduction: We present a catheter technique that utilizes a 5F system for the purpose of balloon-assisted coiling in the setting of intracranial aneurysms. Methods: A standard 5F short sheath is placed in the common femoral artery, and a 5F diagnostic catheter is placed through the sheath and used for selective vessel angiography. When endovascular intervention is pursued, the diagnostic catheter is placed in the appropriate vessel and systemic heparinization is ensured. Over an exchange length wire, the 5F vertebral catheter and 5F short sheath are exchanged for a 5F Shuttle (Cook) sheath. We then routinely place a 10, 14 or 18 microcatheter over an appropriately gauged microguidewire into the aneurysm. As needed, balloon catheters are then placed across the neck of the aneurysm for remodeling purposes. During the course of the procedure, control angiography is performed through the Shuttle sheath. Following the placement of coils, the microcatheter and balloon catheter are removed and a final biplane image is obtained via the 5F Shuttle sheath. Results: This technique has been employed in 15 patients who required balloon-assisted coiling of an intracranial aneurysm. There were no technical difficulties or arterial access site complications from the procedures. Catheter mobility and torque were not affected, nor was the quality of our imaging. Conclusion: We conclude that this small-diameter system provides ample "room" for catheter placement and interventional treatment while reducing the known risks of postprocedural complications. Angiographic images remain excellent and are comparable to those obtained by larger catheters.

Original languageEnglish (US)
Pages (from-to)157-159
Number of pages3
JournalNeuroradiology
Volume49
Issue number2
DOIs
StatePublished - Feb 2007

Fingerprint

Catheters
Intracranial Aneurysm
Aneurysm
Angiography
Torque
Femoral Artery

Keywords

  • Aneurysm
  • Balloon-assisted coiling
  • Catheter

ASJC Scopus subject areas

  • Clinical Neurology
  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Balloon-assisted coiling through a 5-French system. / White, Bradley J.; Layton, Kennith F.; Kallmes, David F; Cloft, Harry J.

In: Neuroradiology, Vol. 49, No. 2, 02.2007, p. 157-159.

Research output: Contribution to journalArticle

White, Bradley J. ; Layton, Kennith F. ; Kallmes, David F ; Cloft, Harry J. / Balloon-assisted coiling through a 5-French system. In: Neuroradiology. 2007 ; Vol. 49, No. 2. pp. 157-159.
@article{3b45fd91eb7142509c809ee810388b75,
title = "Balloon-assisted coiling through a 5-French system",
abstract = "Introduction: We present a catheter technique that utilizes a 5F system for the purpose of balloon-assisted coiling in the setting of intracranial aneurysms. Methods: A standard 5F short sheath is placed in the common femoral artery, and a 5F diagnostic catheter is placed through the sheath and used for selective vessel angiography. When endovascular intervention is pursued, the diagnostic catheter is placed in the appropriate vessel and systemic heparinization is ensured. Over an exchange length wire, the 5F vertebral catheter and 5F short sheath are exchanged for a 5F Shuttle (Cook) sheath. We then routinely place a 10, 14 or 18 microcatheter over an appropriately gauged microguidewire into the aneurysm. As needed, balloon catheters are then placed across the neck of the aneurysm for remodeling purposes. During the course of the procedure, control angiography is performed through the Shuttle sheath. Following the placement of coils, the microcatheter and balloon catheter are removed and a final biplane image is obtained via the 5F Shuttle sheath. Results: This technique has been employed in 15 patients who required balloon-assisted coiling of an intracranial aneurysm. There were no technical difficulties or arterial access site complications from the procedures. Catheter mobility and torque were not affected, nor was the quality of our imaging. Conclusion: We conclude that this small-diameter system provides ample {"}room{"} for catheter placement and interventional treatment while reducing the known risks of postprocedural complications. Angiographic images remain excellent and are comparable to those obtained by larger catheters.",
keywords = "Aneurysm, Balloon-assisted coiling, Catheter",
author = "White, {Bradley J.} and Layton, {Kennith F.} and Kallmes, {David F} and Cloft, {Harry J.}",
year = "2007",
month = "2",
doi = "10.1007/s00234-006-0163-3",
language = "English (US)",
volume = "49",
pages = "157--159",
journal = "Neuroradiology",
issn = "0028-3940",
publisher = "Springer Verlag",
number = "2",

}

TY - JOUR

T1 - Balloon-assisted coiling through a 5-French system

AU - White, Bradley J.

AU - Layton, Kennith F.

AU - Kallmes, David F

AU - Cloft, Harry J.

PY - 2007/2

Y1 - 2007/2

N2 - Introduction: We present a catheter technique that utilizes a 5F system for the purpose of balloon-assisted coiling in the setting of intracranial aneurysms. Methods: A standard 5F short sheath is placed in the common femoral artery, and a 5F diagnostic catheter is placed through the sheath and used for selective vessel angiography. When endovascular intervention is pursued, the diagnostic catheter is placed in the appropriate vessel and systemic heparinization is ensured. Over an exchange length wire, the 5F vertebral catheter and 5F short sheath are exchanged for a 5F Shuttle (Cook) sheath. We then routinely place a 10, 14 or 18 microcatheter over an appropriately gauged microguidewire into the aneurysm. As needed, balloon catheters are then placed across the neck of the aneurysm for remodeling purposes. During the course of the procedure, control angiography is performed through the Shuttle sheath. Following the placement of coils, the microcatheter and balloon catheter are removed and a final biplane image is obtained via the 5F Shuttle sheath. Results: This technique has been employed in 15 patients who required balloon-assisted coiling of an intracranial aneurysm. There were no technical difficulties or arterial access site complications from the procedures. Catheter mobility and torque were not affected, nor was the quality of our imaging. Conclusion: We conclude that this small-diameter system provides ample "room" for catheter placement and interventional treatment while reducing the known risks of postprocedural complications. Angiographic images remain excellent and are comparable to those obtained by larger catheters.

AB - Introduction: We present a catheter technique that utilizes a 5F system for the purpose of balloon-assisted coiling in the setting of intracranial aneurysms. Methods: A standard 5F short sheath is placed in the common femoral artery, and a 5F diagnostic catheter is placed through the sheath and used for selective vessel angiography. When endovascular intervention is pursued, the diagnostic catheter is placed in the appropriate vessel and systemic heparinization is ensured. Over an exchange length wire, the 5F vertebral catheter and 5F short sheath are exchanged for a 5F Shuttle (Cook) sheath. We then routinely place a 10, 14 or 18 microcatheter over an appropriately gauged microguidewire into the aneurysm. As needed, balloon catheters are then placed across the neck of the aneurysm for remodeling purposes. During the course of the procedure, control angiography is performed through the Shuttle sheath. Following the placement of coils, the microcatheter and balloon catheter are removed and a final biplane image is obtained via the 5F Shuttle sheath. Results: This technique has been employed in 15 patients who required balloon-assisted coiling of an intracranial aneurysm. There were no technical difficulties or arterial access site complications from the procedures. Catheter mobility and torque were not affected, nor was the quality of our imaging. Conclusion: We conclude that this small-diameter system provides ample "room" for catheter placement and interventional treatment while reducing the known risks of postprocedural complications. Angiographic images remain excellent and are comparable to those obtained by larger catheters.

KW - Aneurysm

KW - Balloon-assisted coiling

KW - Catheter

UR - http://www.scopus.com/inward/record.url?scp=33846857322&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33846857322&partnerID=8YFLogxK

U2 - 10.1007/s00234-006-0163-3

DO - 10.1007/s00234-006-0163-3

M3 - Article

C2 - 17146636

AN - SCOPUS:33846857322

VL - 49

SP - 157

EP - 159

JO - Neuroradiology

JF - Neuroradiology

SN - 0028-3940

IS - 2

ER -