Analysis of localized quadratic encoding and reconstruction

Research output: Contribution to journalArticle

32 Citations (Scopus)

Abstract

Localized quadratic (LQ) encoding is a method of collecting multislice MRI data sets with greatly overlapping excited slices in such a way that the slice overlap can be deconvolved. This results in reconstructed data with resolution equal to the center-to-center slice spacing, regardless of the amount of excited slice overlap. LQ encoding is analyzed using the modulation transfer function of the encoding and reconstruction process. This allows analysis of many aspects of the technique in a well established theoretical framework. Many different characteristics of the method are explored, including excited slice profiles, required RF magnitudes and specific absorption rate, signal to noise ratio, signal dynamic range, reconstruction artifacts, sensitivity to motion, saturation and inflow effects, modulation transfer function shifting, and off-resonance artifacts. It is suggested that this technique is best suited for applications currently using multiple thin- slab three-dimensional encoding.

Original languageEnglish (US)
Pages (from-to)137-146
Number of pages10
JournalMagnetic Resonance in Medicine
Volume36
Issue number1
DOIs
StatePublished - Jul 1 1996
Externally publishedYes

Fingerprint

Artifacts
Signal-To-Noise Ratio
Datasets

Keywords

  • angiography
  • chirp pulses
  • frequency modulated
  • magnetic resonance imaging

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Analysis of localized quadratic encoding and reconstruction. / Pipe, James.

In: Magnetic Resonance in Medicine, Vol. 36, No. 1, 01.07.1996, p. 137-146.

Research output: Contribution to journalArticle

@article{31a91d1376f0440194fcd504e000a965,
title = "Analysis of localized quadratic encoding and reconstruction",
abstract = "Localized quadratic (LQ) encoding is a method of collecting multislice MRI data sets with greatly overlapping excited slices in such a way that the slice overlap can be deconvolved. This results in reconstructed data with resolution equal to the center-to-center slice spacing, regardless of the amount of excited slice overlap. LQ encoding is analyzed using the modulation transfer function of the encoding and reconstruction process. This allows analysis of many aspects of the technique in a well established theoretical framework. Many different characteristics of the method are explored, including excited slice profiles, required RF magnitudes and specific absorption rate, signal to noise ratio, signal dynamic range, reconstruction artifacts, sensitivity to motion, saturation and inflow effects, modulation transfer function shifting, and off-resonance artifacts. It is suggested that this technique is best suited for applications currently using multiple thin- slab three-dimensional encoding.",
keywords = "angiography, chirp pulses, frequency modulated, magnetic resonance imaging",
author = "James Pipe",
year = "1996",
month = "7",
day = "1",
doi = "10.1002/mrm.1910360122",
language = "English (US)",
volume = "36",
pages = "137--146",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",
number = "1",

}

TY - JOUR

T1 - Analysis of localized quadratic encoding and reconstruction

AU - Pipe, James

PY - 1996/7/1

Y1 - 1996/7/1

N2 - Localized quadratic (LQ) encoding is a method of collecting multislice MRI data sets with greatly overlapping excited slices in such a way that the slice overlap can be deconvolved. This results in reconstructed data with resolution equal to the center-to-center slice spacing, regardless of the amount of excited slice overlap. LQ encoding is analyzed using the modulation transfer function of the encoding and reconstruction process. This allows analysis of many aspects of the technique in a well established theoretical framework. Many different characteristics of the method are explored, including excited slice profiles, required RF magnitudes and specific absorption rate, signal to noise ratio, signal dynamic range, reconstruction artifacts, sensitivity to motion, saturation and inflow effects, modulation transfer function shifting, and off-resonance artifacts. It is suggested that this technique is best suited for applications currently using multiple thin- slab three-dimensional encoding.

AB - Localized quadratic (LQ) encoding is a method of collecting multislice MRI data sets with greatly overlapping excited slices in such a way that the slice overlap can be deconvolved. This results in reconstructed data with resolution equal to the center-to-center slice spacing, regardless of the amount of excited slice overlap. LQ encoding is analyzed using the modulation transfer function of the encoding and reconstruction process. This allows analysis of many aspects of the technique in a well established theoretical framework. Many different characteristics of the method are explored, including excited slice profiles, required RF magnitudes and specific absorption rate, signal to noise ratio, signal dynamic range, reconstruction artifacts, sensitivity to motion, saturation and inflow effects, modulation transfer function shifting, and off-resonance artifacts. It is suggested that this technique is best suited for applications currently using multiple thin- slab three-dimensional encoding.

KW - angiography

KW - chirp pulses

KW - frequency modulated

KW - magnetic resonance imaging

UR - http://www.scopus.com/inward/record.url?scp=0029884564&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0029884564&partnerID=8YFLogxK

U2 - 10.1002/mrm.1910360122

DO - 10.1002/mrm.1910360122

M3 - Article

VL - 36

SP - 137

EP - 146

JO - Magnetic Resonance in Medicine

JF - Magnetic Resonance in Medicine

SN - 0740-3194

IS - 1