An effective noise reduction method for multi-energy CT images that exploit spatio-spectral features

Research output: Contribution to journalArticle

10 Citations (Scopus)

Abstract

PURPOSE: To develop and evaluate an image-domain noise reduction method for multi-energy CT (MECT) data.

METHODS: Multi-Energy Non-Local Means (MENLM) is a technique that uses the redundant information in MECT images to achieve noise reduction. In this method, spatio-spectral features are used to determine the similarity between pixels, making the similarity evaluation more robust to image noise. The performance of this MENLM filter was tested on images acquired on a whole-body research photon counting CT system. The impact of filtering on image quality was quantitatively evaluated in phantom studies in terms of image noise level (standard deviation of pixel values), noise power spectrum (NPS), in-plane and cross-plane spatial resolution, CT number accuracy, material decomposition performance, and subjective low-contrast spatial resolution using the American College of Radiology (ACR) CT accreditation phantom. Clinical feasibility was assessed by performing MENLM on contrast-enhanced swine images and unenhanced cadaver head images using clinically relevant doses and dose rates.

RESULTS: The phantom studies demonstrated that the MENLM filter reduced noise substantially and still preserved the shape and peak frequency of the NPS. With 80% noise reduction, MENLM filtering caused no degradation of high-contrast spatial resolution, as illustrated by the modulation transfer function (MTF) and slice sensitivity profile (SSP). CT number accuracy was also maintained for all energy channels, demonstrating that energy resolution was not affected by filtering. Material decomposition performance was improved with MENLM filtering. The subjective evaluation using the ACR phantom demonstrated an improvement in low-contrast performance. MENLM achieved effective noise reduction in both contrast-enhanced swine images and unenhanced cadaver head images, resulting in improved detection of subtle vascular structures and the differentiation of white/gray matter.

CONCLUSION: In MECT, MENLM achieved around 80% noise reduction and greatly improved material decomposition performance and the detection of subtle anatomical/low-contrast features while maintaining spatial and energy resolution. MENLM filtering may improve diagnostic or functional analysis accuracy and facilitate radiation dose and contrast media reduction for MECT.

Original languageEnglish (US)
Pages (from-to)1610-1623
Number of pages14
JournalMedical Physics
Volume44
Issue number5
DOIs
StatePublished - May 1 2017

Fingerprint

Noise
Cadaver
Radiology
Swine
Head
Accreditation
Photons
Contrast Media
Blood Vessels
Radiation
Research

Keywords

  • CT dose reduction
  • image denoising
  • multi-energy CT
  • non-local means filtering
  • photon counting CT

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

@article{80f0ac06a0604d3eb18400d27fcec4f9,
title = "An effective noise reduction method for multi-energy CT images that exploit spatio-spectral features",
abstract = "PURPOSE: To develop and evaluate an image-domain noise reduction method for multi-energy CT (MECT) data.METHODS: Multi-Energy Non-Local Means (MENLM) is a technique that uses the redundant information in MECT images to achieve noise reduction. In this method, spatio-spectral features are used to determine the similarity between pixels, making the similarity evaluation more robust to image noise. The performance of this MENLM filter was tested on images acquired on a whole-body research photon counting CT system. The impact of filtering on image quality was quantitatively evaluated in phantom studies in terms of image noise level (standard deviation of pixel values), noise power spectrum (NPS), in-plane and cross-plane spatial resolution, CT number accuracy, material decomposition performance, and subjective low-contrast spatial resolution using the American College of Radiology (ACR) CT accreditation phantom. Clinical feasibility was assessed by performing MENLM on contrast-enhanced swine images and unenhanced cadaver head images using clinically relevant doses and dose rates.RESULTS: The phantom studies demonstrated that the MENLM filter reduced noise substantially and still preserved the shape and peak frequency of the NPS. With 80{\%} noise reduction, MENLM filtering caused no degradation of high-contrast spatial resolution, as illustrated by the modulation transfer function (MTF) and slice sensitivity profile (SSP). CT number accuracy was also maintained for all energy channels, demonstrating that energy resolution was not affected by filtering. Material decomposition performance was improved with MENLM filtering. The subjective evaluation using the ACR phantom demonstrated an improvement in low-contrast performance. MENLM achieved effective noise reduction in both contrast-enhanced swine images and unenhanced cadaver head images, resulting in improved detection of subtle vascular structures and the differentiation of white/gray matter.CONCLUSION: In MECT, MENLM achieved around 80{\%} noise reduction and greatly improved material decomposition performance and the detection of subtle anatomical/low-contrast features while maintaining spatial and energy resolution. MENLM filtering may improve diagnostic or functional analysis accuracy and facilitate radiation dose and contrast media reduction for MECT.",
keywords = "CT dose reduction, image denoising, multi-energy CT, non-local means filtering, photon counting CT",
author = "Zhoubo Li and Shuai Leng and Lifeng Yu and Armando Manduca and McCollough, {Cynthia H}",
year = "2017",
month = "5",
day = "1",
doi = "10.1002/mp.12174",
language = "English (US)",
volume = "44",
pages = "1610--1623",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "5",

}

TY - JOUR

T1 - An effective noise reduction method for multi-energy CT images that exploit spatio-spectral features

AU - Li, Zhoubo

AU - Leng, Shuai

AU - Yu, Lifeng

AU - Manduca, Armando

AU - McCollough, Cynthia H

PY - 2017/5/1

Y1 - 2017/5/1

N2 - PURPOSE: To develop and evaluate an image-domain noise reduction method for multi-energy CT (MECT) data.METHODS: Multi-Energy Non-Local Means (MENLM) is a technique that uses the redundant information in MECT images to achieve noise reduction. In this method, spatio-spectral features are used to determine the similarity between pixels, making the similarity evaluation more robust to image noise. The performance of this MENLM filter was tested on images acquired on a whole-body research photon counting CT system. The impact of filtering on image quality was quantitatively evaluated in phantom studies in terms of image noise level (standard deviation of pixel values), noise power spectrum (NPS), in-plane and cross-plane spatial resolution, CT number accuracy, material decomposition performance, and subjective low-contrast spatial resolution using the American College of Radiology (ACR) CT accreditation phantom. Clinical feasibility was assessed by performing MENLM on contrast-enhanced swine images and unenhanced cadaver head images using clinically relevant doses and dose rates.RESULTS: The phantom studies demonstrated that the MENLM filter reduced noise substantially and still preserved the shape and peak frequency of the NPS. With 80% noise reduction, MENLM filtering caused no degradation of high-contrast spatial resolution, as illustrated by the modulation transfer function (MTF) and slice sensitivity profile (SSP). CT number accuracy was also maintained for all energy channels, demonstrating that energy resolution was not affected by filtering. Material decomposition performance was improved with MENLM filtering. The subjective evaluation using the ACR phantom demonstrated an improvement in low-contrast performance. MENLM achieved effective noise reduction in both contrast-enhanced swine images and unenhanced cadaver head images, resulting in improved detection of subtle vascular structures and the differentiation of white/gray matter.CONCLUSION: In MECT, MENLM achieved around 80% noise reduction and greatly improved material decomposition performance and the detection of subtle anatomical/low-contrast features while maintaining spatial and energy resolution. MENLM filtering may improve diagnostic or functional analysis accuracy and facilitate radiation dose and contrast media reduction for MECT.

AB - PURPOSE: To develop and evaluate an image-domain noise reduction method for multi-energy CT (MECT) data.METHODS: Multi-Energy Non-Local Means (MENLM) is a technique that uses the redundant information in MECT images to achieve noise reduction. In this method, spatio-spectral features are used to determine the similarity between pixels, making the similarity evaluation more robust to image noise. The performance of this MENLM filter was tested on images acquired on a whole-body research photon counting CT system. The impact of filtering on image quality was quantitatively evaluated in phantom studies in terms of image noise level (standard deviation of pixel values), noise power spectrum (NPS), in-plane and cross-plane spatial resolution, CT number accuracy, material decomposition performance, and subjective low-contrast spatial resolution using the American College of Radiology (ACR) CT accreditation phantom. Clinical feasibility was assessed by performing MENLM on contrast-enhanced swine images and unenhanced cadaver head images using clinically relevant doses and dose rates.RESULTS: The phantom studies demonstrated that the MENLM filter reduced noise substantially and still preserved the shape and peak frequency of the NPS. With 80% noise reduction, MENLM filtering caused no degradation of high-contrast spatial resolution, as illustrated by the modulation transfer function (MTF) and slice sensitivity profile (SSP). CT number accuracy was also maintained for all energy channels, demonstrating that energy resolution was not affected by filtering. Material decomposition performance was improved with MENLM filtering. The subjective evaluation using the ACR phantom demonstrated an improvement in low-contrast performance. MENLM achieved effective noise reduction in both contrast-enhanced swine images and unenhanced cadaver head images, resulting in improved detection of subtle vascular structures and the differentiation of white/gray matter.CONCLUSION: In MECT, MENLM achieved around 80% noise reduction and greatly improved material decomposition performance and the detection of subtle anatomical/low-contrast features while maintaining spatial and energy resolution. MENLM filtering may improve diagnostic or functional analysis accuracy and facilitate radiation dose and contrast media reduction for MECT.

KW - CT dose reduction

KW - image denoising

KW - multi-energy CT

KW - non-local means filtering

KW - photon counting CT

UR - http://www.scopus.com/inward/record.url?scp=85045368443&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85045368443&partnerID=8YFLogxK

U2 - 10.1002/mp.12174

DO - 10.1002/mp.12174

M3 - Article

C2 - 28236645

AN - SCOPUS:85045368443

VL - 44

SP - 1610

EP - 1623

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 5

ER -