A monitoring, feedback, and triggering system for reproducible breath‐hold MR imaging

Yu L. Liu, Stephen J. Riederer, Phillip J. Rossman, Roger C. Grim, Josef P. Debbins, Richard L. Ehman

Research output: Contribution to journalArticlepeer-review

151 Scopus citations

Abstract

A technique is described that provides improved reproducibility of breath‐holding for MR image acquisition by monitoring the superior‐inferior (S/I) position of the diaphragm. The method incorporates detection of the level of inspiration using an MR signal, rapid display to the patient of diaphragm position to enable breath‐hold adjustment, and triggering of image data acquisition once appropriate position is attained. The response time of the system is short, approximately 10 ms. Studies in six volunteers using this method demonstrate a considerable decrease in the S/I range of diaphragm position over 10 consecutive periods of suspended respiration. The mean range is 1.3 mm with the system, while it is 8.3 mm without using it is expected that this method will be of assistance in many abdominal and cardiothoracic studies that use breath‐hold techniques.

Original languageEnglish (US)
Pages (from-to)507-511
Number of pages5
JournalMagnetic Resonance in Medicine
Volume30
Issue number4
DOIs
StatePublished - Oct 1993

Keywords

  • abdominal imaging
  • breath‐hold imaging
  • cardiac imaging