TY - JOUR
T1 - A modified saturation‐recovery approximation for multiple spin‐echo pulse sequences
AU - Lee, James N.
AU - Riederer, Stephen J.
PY - 1986/2
Y1 - 1986/2
N2 - In equations describing multiple spin‐echo pulse sequences, the T1 dependence is often approximated by a saturation‐recovery expression, which is valid only when the repetition time TR is much greater than the time required for the echoes. A slight modification of this approximation greatly increases its accuracy for small TR values. © 1986 Academic Press, Inc.
AB - In equations describing multiple spin‐echo pulse sequences, the T1 dependence is often approximated by a saturation‐recovery expression, which is valid only when the repetition time TR is much greater than the time required for the echoes. A slight modification of this approximation greatly increases its accuracy for small TR values. © 1986 Academic Press, Inc.
UR - http://www.scopus.com/inward/record.url?scp=0022644517&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0022644517&partnerID=8YFLogxK
U2 - 10.1002/mrm.1910030119
DO - 10.1002/mrm.1910030119
M3 - Comment/debate
C2 - 3959877
AN - SCOPUS:0022644517
SN - 0740-3194
VL - 3
SP - 132
EP - 134
JO - Magnetic Resonance in Medicine
JF - Magnetic Resonance in Medicine
IS - 1
ER -